Intra-Cardiac Flow from Geometry Prescribed Computational Fluid Dynamics: Comparison with Ultrasound Vector Flow Imaging
Cardiovascular Engineering and Technology · 2023-08 · 14 (4) · 489-504
DOI: 10.1007/s13239-023-00666-2 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, online erschienen am 15.06.2023. Verfasst von Rasmus Hvid und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Health Technology, Technical University of Denmark, 2800, Kongens Lyngby, Denmark. Hauptthemen laut MEDLINE: Hydrodynamics, Models, Cardiovascular. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Innovationsfonden, Technical University of Denmark.
Abstract (Originalsprache)
Purpose
This paper investigates the accuracy of blood flow velocities simulated from a geometry prescribed computational fluid dynamics (CFD) pipeline by applying it to a dynamic heart phantom. The CFD flow patterns are compared to a direct flow measurement by ultrasound vector flow imaging (VFI). The hypothesis is that the simulated velocity magnitudes are within one standard deviation of the measured velocities.
Methods
The CFD pipeline uses computed tomography angiography (CTA) images with 20 volumes per cardiac cycle as geometry input. Fluid domain movement is prescribed from volumetric image registration using the CTA image data. Inlet and outlet conditions are defined by the experimental setup. VFI is systematically measured in parallel planes, and compared to the corresponding planes in the simulated time dependent three dimensional fluid velocity field.
Results
The measured VFI and simulated CFD have similar flow patterns when compared qualitatively. A quantitative comparison of the velocity magnitude is also performed at specific regions of interest. These are evaluated at 11 non-overlapping time bins and compared by linear regression giving R 2 = 0.809, SD = 0.060 m/s, intercept = − 0.039 m/s, and slope = 1.09. Excluding an outlier at the inlet, the correspondence between CFD and VFI improves to: R 2 = 0.823, SD = 0.048 m/s, intercept = -0.030 m/s, and slope = 1.01.
Conclusion
The direct comparison of flow patterns shows that the proposed CFD pipeline provide realistic flow patterns in a well-controlled experimental setup. The demanded accuracy is obtained close to the inlet and outlet, but not in locations far from these.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Health Technology, Technical University of Denmark, 2800, Kongens Lyngby, Denmark.
- Department of Health Technology, Technical University of Denmark, 2800, Kongens Lyngby, Denmark. msene@dtu.dk.
ORCID-Profile: Rasmus Hvid, Matthias Bo Stuart, Jørgen Arendt Jensen, Marie Sand Traberg
Förderung
- Innovationsfonden
- Technical University of Denmark
Zitieren
APA
Hvid, R., Stuart, M. B., Jensen, J. A., & Traberg, M. S. (2023). Intra-Cardiac Flow from Geometry Prescribed Computational Fluid Dynamics: Comparison with Ultrasound Vector Flow Imaging. Cardiovascular Engineering and Technology, 14(4), 489-504. https://doi.org/10.1007/s13239-023-00666-2
Vancouver
Hvid R, Stuart MB, Jensen JA, Traberg MS. Intra-Cardiac Flow from Geometry Prescribed Computational Fluid Dynamics: Comparison with Ultrasound Vector Flow Imaging. Cardiovascular Engineering and Technology. 2023;14(4):489-504. doi:10.1007/s13239-023-00666-2
BibTeX
@article{Hvid2023,
author = {Hvid, Rasmus and Stuart, Matthias Bo and Jensen, Jørgen Arendt and Traberg, Marie Sand},
title = {Intra-Cardiac Flow from Geometry Prescribed Computational Fluid Dynamics: Comparison with Ultrasound Vector Flow Imaging},
journal = {Cardiovascular Engineering and Technology},
year = {2023},
volume = {14},
number = {4},
pages = {489--504},
doi = {10.1007/s13239-023-00666-2}
}
EUVIMED