High Spatial Resolution 23Na-MRI for Ischemic Brain Injury Detection
Translational Stroke Research · 2026-12 · 17 (6)
DOI: 10.1007/s12975-026-01503-z ↗Kurz gefasst
Fachartikel in der Zeitschrift Translational Stroke Research, online erschienen am 30.10.2026. Verfasst von Yuanyuan Jiang und 9 weiteren Autor:innen. Beteiligte Einrichtung: Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain Ischemia, Brain, Brain Injuries. Die Arbeit nennt 73 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NINDS NIH HHS, NCATS NIH HHS, NIH HHS.
Abstract (Originalsprache)
High spatial resolution sodium (^23Na) imaging of brain lesions remains challenging due to the intrinsically low signal-to-noise ratio (SNR) of ^23Na-MRI compared with conventional proton (^1H) MRI. In this preclinical mouse study, we developed a high-resolution ^23Na-MRI platform based on a 14 T preclinical scanner using a dual-tuned head-implanted RF coil. This implanted coil configuration enabled longitudinal multimodal imaging in the same animal, including 1^1H-based T_2-weighted anatomical and diffusion-weighted imaging (DWI), and high-resolution ^23Na-MRI. To evaluate the capability of this approach for detecting acute ischemic injury, ischemic stroke was induced in mice using the middle cerebral artery occlusion (MCAO) model. Brain-wide ^23Na-MRI was acquired within the first 4 hours following MCAO at an in-plane resolution of 300 × 300 µm². Hyperintense ^23Na-MRI signals were observed in the ischemic brain regions, presenting a contrast-to-noise ratio (CNR) that exceeded that of T_2-weighted images. Infarct regions can be well identified by elevated ^23Na-MRI signal, which is spatially consistent with regions of reduced apparent diffusion coefficient (ADC) values, indicating water restriction in the ischemic core. This work established implanted dual-tuned ^1H/^23Na RF coils as a powerful approach for high-resolution ^23Na-MRI readout, facilitating the early detection of ischemic brain injury.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, USA.
- Department of Neurology, Massachusetts General Hospital, Harvard Medical School, Charlestown, MA, USA.
- Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, USA. xyu9@mgh.harvard.edu.
Förderung
- NINDS NIH HHS
- NCATS NIH HHS
- NIH HHS
- American Heart Association and American Stroke Association
- National Science Foundation
- NIMH NIH HHS
- Andrew David Heitman Foundation
- National Institutes of Health
Zitieren
APA
Jiang, Y., Zhou, X. A., Imai, T., Hawley, L., Hike, D., Mohammed, S., Yu, G., Ayata, C., Chung, D. Y., & Yu, X. (2026). High Spatial Resolution 23Na-MRI for Ischemic Brain Injury Detection. Translational Stroke Research, 17(6). https://doi.org/10.1007/s12975-026-01503-z
Vancouver
Jiang Y, Zhou XA, Imai T, Hawley L, Hike D, Mohammed S, et al. High Spatial Resolution 23Na-MRI for Ischemic Brain Injury Detection. Translational Stroke Research. 2026;17(6). doi:10.1007/s12975-026-01503-z
BibTeX
@article{Jiang2026,
author = {Jiang, Yuanyuan and Zhou, Xiaoqing Alice and Imai, Takahiko and Hawley, Lydia and Hike, David and Mohammed, Sohail and Yu, Grace and Ayata, Cenk and Chung, David Y and Yu, Xin},
title = {High Spatial Resolution 23Na-MRI for Ischemic Brain Injury Detection},
journal = {Translational Stroke Research},
year = {2026},
volume = {17},
number = {6},
doi = {10.1007/s12975-026-01503-z}
}