Initial experience with a novel stent-based mechanical thrombectomy device for management of acute myocardial infarction cases with large thrombus burden
Cardiovascular Intervention and Therapeutics · 2024
DOI: 10.1007/s12928-024-00998-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cardiovascular Intervention and Therapeutics, 2024. Verfasst von Yuan Zhi und 9 weiteren Autor:innen. Beteiligte Einrichtung: Cardiology Division, Heart Center, Luzerner Kantonsspital, 6000, Lucerne 16, Switzerland. Hauptthemen laut MEDLINE: Thrombectomy, Myocardial Infarction, Stents. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Luzern.
Abstract (Originalsprache)
Background
Patients with acute myocardial infarction (AMI) and large thrombus burden (LTB) still represent a challenge. Afflicted patients have a high morbidity and mortality. Aspiration thrombectomy is often ineffective in those cases. Mechanical thrombectomy devices (MTDs), which are effective for management of ischemic strokes, were recently CE-approved for treatment of thrombotic coronary lesions. Real-world data about their performance in AMI cases with LTB are scarce. This study sought to summarize our early experience with a novel MTD device in this context.
Methods
We analyzed consecutive patients from the prospective OPTIMISER registry (NCT04988672), who have been managed with the NeVa™ MTD (Vesalio, USA) for AMI with LTB at a tertiary cardiology facility. Outcomes of interest included, among others, periprocedural complications, target lesion failure (TLF), target lesion revascularization (TLR) and target vessel myocardial infarction (TV-MI).
Results
Overall, 15 patients underwent thrombectomy with the NeVa™ device. Thrombectomy was successfully performed in 14 (93%) patients. Final TIMI 3 flow was achieved in 13 (87%) patients, while 2 (13%) patients had TIMI 2 flow. We encountered no relevant periprocedural complications, especially no stroke, stent thrombosis or vessel closure. After a mean follow-up time of 26 ± 2.9 months, 1 (7%) patient presented with TLR due to stent thrombosis (10 months after treatment with the MTD and stenting).
Conclusions
In AMI patients with LTB, the deployment of the novel NeVa™ MTD seems efficient and safe. Further randomized trials are warranted to assess whether the use of the NeVa™ device in cases with LTB improves procedural and clinical outcomes. Graphical abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiology Division, Heart Center, Luzerner Kantonsspital, 6000, Lucerne 16, Switzerland.
- Faculty of Medicine, University of Basel, Basel, Switzerland.
- Faculty of Health Sciences and Medicine, University of Lucerne, Lucerne, Switzerland.
- Università Statale Di Milano, Milan, Italy.
- Cardiology Division, Heart Center, Luzerner Kantonsspital, 6000, Lucerne 16, Switzerland. matthias.bossard@luks.ch.
- Faculty of Health Sciences and Medicine, University of Lucerne, Lucerne, Switzerland. matthias.bossard@luks.ch.
Förderung
- University of Luzern
Zitieren
APA
Zhi, Y., Madanchi, M., Cioffi, G. M., Brunner, J., Stutz, L., Gnan, E., Gjergjizi, V., Attinger-Toller, A., Cuculi, F., & Bossard, M. (2024). Initial experience with a novel stent-based mechanical thrombectomy device for management of acute myocardial infarction cases with large thrombus burden. Cardiovascular Intervention and Therapeutics. https://doi.org/10.1007/s12928-024-00998-3
Vancouver
Zhi Y, Madanchi M, Cioffi GM, Brunner J, Stutz L, Gnan E, et al. Initial experience with a novel stent-based mechanical thrombectomy device for management of acute myocardial infarction cases with large thrombus burden. Cardiovascular Intervention and Therapeutics. 2024. doi:10.1007/s12928-024-00998-3
BibTeX
@article{Zhi2024,
author = {Zhi, Yuan and Madanchi, Mehdi and Cioffi, Giacomo Maria and Brunner, Julian and Stutz, Leah and Gnan, Eleonora and Gjergjizi, Varis and Attinger-Toller, Adrian and Cuculi, Florim and Bossard, Matthias},
title = {Initial experience with a novel stent-based mechanical thrombectomy device for management of acute myocardial infarction cases with large thrombus burden},
journal = {Cardiovascular Intervention and Therapeutics},
year = {2024},
doi = {10.1007/s12928-024-00998-3}
}
EUVIMED