EUVIMEDEUVIMED entdecken
Crossref

Basic characteristics of do-it-yourself dosimeter in general radiography system

Nao Ichikawa, Atsushi Fukuda, Takuma Hayashi, Taku Kuramoto, Kosuke Matsubara

Radiological Physics and Technology · 2026-06 · 19 (2) · 863-870

DOI: 10.1007/s12194-026-01034-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiological Physics and Technology, online erschienen am 09.03.2026. Verfasst von Nao Ichikawa und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Faculty of Health Sciences, Kobe Tokiwa University, 2-6-2 Otani-cho, Nagata-ku, Kobe, Hyogo, 653-0838, Japan. …. Hauptthemen laut MEDLINE: Radiation Dosimeters, Radiography. Die Arbeit nennt 23 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study evaluated the basic characteristics of a do-it-yourself dosimeter (DIYD) and its applicability in general radiography systems. An ionization chamber (IC), solid-state detector (SSD), and DIYD were placed 100 cm from the X-ray focus. Incident air kerma ([Formula: see text], [Formula: see text], [Formula: see text]) and the rate ([Formula: see text], [Formula: see text], [Formula: see text]) were measured as functions of exposure time, tube current, tube voltage, and dosimeter angles to assess [Formula: see text] and [Formula: see text] linearity, exposure time accuracy, energy dependence, and angular dependence. The [Formula: see text] and [Formula: see text] were 1.00–1.05 and 0.95–1.04. DIYD showed energy dependence peaking at 36 keV (ratio: 1.01–1.11). Exposure time by DIYD agreed with IC and SSD for ≥ 40 ms. DIYD showed superior angular dependence compared with SSD. DIYD showed characteristics comparable to IC and SSD, indicating its potential for periodic quality control and optimization of patient exposure in general radiography systems.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation DosimetersRadiography

Stichwörter der Autor:innen

Basic characteristicsDo-it-yourself dosimeterGeneral radiography systemMeasurement accuracyQuality control

Autor:innen und Institutionen

  • Department of Radiological Technology, Faculty of Health Sciences, Kobe Tokiwa University, 2-6-2 Otani-cho, Nagata-ku, Kobe, Hyogo, 653-0838, Japan. n_a_o523@yahoo.co.jp.
  • Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima, Japan.
  • Department of Radiology, Shiga Psychiatric Medical Center, Shiga, Japan.
  • Department of Radiological Technology, Faculty of Health Sciences, Kobe Tokiwa University, 2-6-2 Otani-cho, Nagata-ku, Kobe, Hyogo, 653-0838, Japan.
  • Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Ishikawa, Japan.

Zitieren

APA

Ichikawa, N., Fukuda, A., Hayashi, T., Kuramoto, T., & Matsubara, K. (2026). Basic characteristics of do-it-yourself dosimeter in general radiography system. Radiological Physics and Technology, 19(2), 863-870. https://doi.org/10.1007/s12194-026-01034-0

Vancouver

Ichikawa N, Fukuda A, Hayashi T, Kuramoto T, Matsubara K. Basic characteristics of do-it-yourself dosimeter in general radiography system. Radiological Physics and Technology. 2026;19(2):863-870. doi:10.1007/s12194-026-01034-0

BibTeX
@article{Ichikawa2026,
  author = {Ichikawa, Nao and Fukuda, Atsushi and Hayashi, Takuma and Kuramoto, Taku and Matsubara, Kosuke},
  title = {Basic characteristics of do-it-yourself dosimeter in general radiography system},
  journal = {Radiological Physics and Technology},
  year = {2026},
  volume = {19},
  number = {2},
  pages = {863--870},
  doi = {10.1007/s12194-026-01034-0}
}
Kostenlos starten →