Comparative evaluation of quantitative analysis using two phantom-based software programs for 18F-FDG-PET/CT

Kosuke Yamashita, Kazuki Motegi, Noriaki Miyaji, Shohei Fukai, Yasuhiro Fujiwara, Takashi Terauchi

Radiological Physics and Technology · 2025

DOI: 10.1007/s12194-025-00974-3 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Radiological Physics and Technology, 2025. Verfasst von Kosuke Yamashita und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Imaging Technology, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan. kyamashita@kumamoto-u.ac.jp. Hauptthemen laut MEDLINE: Fluorodeoxyglucose F18, Phantoms, Imaging, Positron Emission Tomography Computed Tomography, Software. Die Arbeit nennt 23 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Variability in image quality and quantitative accuracy of Fluorine-18-fluorodeoxyglucose positron emission tomography/computed tomography (18F-FDG-PET/CT) has been reported across institutions and devices. The Japanese Society of Nuclear Medicine (JSNM) introduced two guidelines to facilitate image quality assurance. Recently, a fully automated software, Arimaru (PDRadiopharma Inc.), has been developed to streamline this process. However, its performance relative to conventional software has not yet been fully validated. This study aimed to compare the physical parameters calculated using Arimaru and conventional software (PETquact IE, Nihon Medi-Physics Co., Ltd.) from PET images acquired using two PET/CT systems: Discovery MI and Discovery IQ (GE Healthcare). Images of the NEMA IEC Body phantom were acquired in list mode for 1800 s and reconstructed at multiple time points (30-1800 s) to simulate different noise levels. Five physical parameters (QH,10, N10, QH,10/N10, CVBG, and SUVmax) were calculated using both the methods. The results showed that the automated method accurately positioned the region of interests (ROIs) and had a strong correlation with the conventional method across all parameters (r > 0.85, p < 0.05). However, some physical parameter values from the automated method were significantly different from those obtained using conventional software program. In conclusion, the automated software showed strong concordance with the conventional method and met JSNM guidelines. Nevertheless, systematic differences in the calculated values highlight the need to understand software-specific characteristics. The adoption of such tools may promote a broader and more consistent implementation of standardized PET imaging practices.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Fluorodeoxyglucose F18Phantoms, ImagingPositron Emission Tomography Computed TomographySoftwareImage Processing, Computer-AssistedHumans

Stichwörter der Autor:innen

18F-FDGJapanese guideline for the oncology FDG-PET/CT data acquisition protocolPhysical parametersPositron emission tomography (PET)The standard PET imaging protocols and phantom test procedures and criteria

Substanzen

Fluorodeoxyglucose F18

Autor:innen und Institutionen

  • Department of Medical Imaging Technology, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan. kyamashita@kumamoto-u.ac.jp.
  • Department of Nuclear Medicine, The Cancer Institute Hospital of Japanese Foundation for Cancer Research, Tokyo, Japan.
  • Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima, Japan.
  • Department of Medical Imaging Technology, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan.

Zitieren

APA

Yamashita, K., Motegi, K., Miyaji, N., Fukai, S., Fujiwara, Y., & Terauchi, T. (2025). Comparative evaluation of quantitative analysis using two phantom-based software programs for 18F-FDG-PET/CT. Radiological Physics and Technology. https://doi.org/10.1007/s12194-025-00974-3

Vancouver

Yamashita K, Motegi K, Miyaji N, Fukai S, Fujiwara Y, Terauchi T. Comparative evaluation of quantitative analysis using two phantom-based software programs for 18F-FDG-PET/CT. Radiological Physics and Technology. 2025. doi:10.1007/s12194-025-00974-3

BibTeX
@article{Yamashita2025,
  author = {Yamashita, Kosuke and Motegi, Kazuki and Miyaji, Noriaki and Fukai, Shohei and Fujiwara, Yasuhiro and Terauchi, Takashi},
  title = {Comparative evaluation of quantitative analysis using two phantom-based software programs for 18F-FDG-PET/CT},
  journal = {Radiological Physics and Technology},
  year = {2025},
  doi = {10.1007/s12194-025-00974-3}
}
Kostenlos starten →