Dosimetric effects of small field size, dose grid size, and variable split-arc methods on gamma pass rates in radiation therapy

Tsunekazu Kuwae, Takuro Ariga, Takeaki Kusada, Akihiro Nishie

Radiological Physics and Technology · 2024

DOI: 10.1007/s12194-024-00809-7 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiological Physics and Technology, 2024. Verfasst von Tsunekazu Kuwae und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Yuuai Medical Center, Tomigusuku, Okinawa, Japan. cyunekazu@gmail.com. Hauptthemen laut MEDLINE: Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted, Radiometry, Gamma Rays. Die Arbeit nennt 8 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study investigates the influence of calculation accuracy in peripheral low-dose regions on the gamma pass rate (GPR), utilizing the Acuros XB (AXB) algorithm and ArcCHECK™ measurement. The effects of varying small field sizes, dose grid sizes, and split-arc techniques on GPR were analyzed. Various small field sizes were employed. Thirty-two single-arc plans with dose grid sizes of 2 mm and 1 mm and prescribed doses of 2, 5, 10, and 20 Gy were calculated using the AXB algorithm. In total, 128 GPR plans were examined. These plans were categorized into three sub-fields (3SF), four sub-fields (4SF), and six sub-fields (6SF). The GPR results deteriorated with smaller target sizes and a 2 mm dose grid size in a single arc. A similar degradation in GPR was observed with smaller target sizes and a 1 mm dose grid size. However, the 1 mm dose grid size generally resulted in better GPR compared with the 2 mm dose grid size for the same target sizes. The GPR improved with finer split angles and a 2 mm dose grid size in the split-arc method. However, no statistically significant improvement was observed with finer split angles and a 1 mm dose grid size. This study demonstrates that coarser dose grid sizes result in lower GPRs in peripheral low-dose regions as calculated by AXB with ArcCHECK™ measurement. To enhance GPR, employing split-arc methods and finer dose grid sizes could be beneficial.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy DosageRadiotherapy Planning, Computer-AssistedRadiometryGamma RaysAlgorithmsHumansRadiation DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

Discretization errorDose grid sizeGamma pass rateSmall field sizeSplit-arc method

Autor:innen und Institutionen

  • Department of Radiology, Yuuai Medical Center, Tomigusuku, Okinawa, Japan. cyunekazu@gmail.com.
  • Health Information Management Center, University of the Ryukyus Hospital, Nishihara, Okinawa, Japan.
  • Department of Radiology, Yuuai Medical Center, Tomigusuku, Okinawa, Japan.
  • Department of Radiology, Graduate School of Medicine, University of the Ryukyus, Nishihara, Okinawa, Japan.

Zitieren

APA

Kuwae, T., Ariga, T., Kusada, T., & Nishie, A. (2024). Dosimetric effects of small field size, dose grid size, and variable split-arc methods on gamma pass rates in radiation therapy. Radiological Physics and Technology. https://doi.org/10.1007/s12194-024-00809-7

Vancouver

Kuwae T, Ariga T, Kusada T, Nishie A. Dosimetric effects of small field size, dose grid size, and variable split-arc methods on gamma pass rates in radiation therapy. Radiological Physics and Technology. 2024. doi:10.1007/s12194-024-00809-7

BibTeX
@article{Kuwae2024,
  author = {Kuwae, Tsunekazu and Ariga, Takuro and Kusada, Takeaki and Nishie, Akihiro},
  title = {Dosimetric effects of small field size, dose grid size, and variable split-arc methods on gamma pass rates in radiation therapy},
  journal = {Radiological Physics and Technology},
  year = {2024},
  doi = {10.1007/s12194-024-00809-7}
}
Kostenlos starten →