3D-printed boluses for radiotherapy: influence of geometrical and printing parameters on dosimetric characterization and air gap evaluation
Radiological Physics and Technology · 2024
DOI: 10.1007/s12194-024-00782-1 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiological Physics and Technology, 2024. Verfasst von Simone Giovanni Gugliandolo und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Politecnico di Milano, Via La Masa, 1, 20156, Milano, Italy. Hauptthemen laut MEDLINE: Printing, Three-Dimensional, Radiometry, Phantoms, Imaging. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Politecnico di Milano.
Abstract (Originalsprache)
The work investigates the implementation of personalized radiotherapy boluses by means of additive manufacturing technologies. Boluses materials that are currently used need an excessive amount of human intervention which leads to reduced repeatability in terms of dosimetry. Additive manufacturing can solve this problem by eliminating the human factor in the process of fabrication. Planar boluses with fixed geometry and personalized boluses printed starting from a computed tomography scan of a radiotherapy phantom were produced. First, a dosimetric characterization study on planar bolus designs to quantify the effects of print parameters such as infill density and geometry on the radiation beam was made. Secondly, a volumetric quantification of air gap between the bolus and the skin of the patient as well as dosimetric analyses were performed. The optimization process according to the obtained dosimetric and airgap results allowed us to find a combination of parameters to have the 3D-printed bolus performing similarly to that in conventional use. These preliminary results confirm those in the relevant literature, with 3D-printed boluses showing a dosimetric performance similar to conventional boluses with the additional advantage of being perfectly conformed to the patient geometry.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechanical Engineering, Politecnico di Milano, Via La Masa, 1, 20156, Milano, Italy.
- Department of Chemistry, Materials and Chemical Engineering "Giulio Natta", Politecnico di Milano, Piazza Leonardo da Vinci, 32, 20133, Milano, Italy.
- Division of Radiation Oncology, IEO European Institute of Oncology, IRCCS, Milano, Italy.
- Clinical Department, Bioengineering Unit, National Center for Oncological Hadrontherapy (CNAO), Pavia, Italy.
- Unit of Medical Physics, IEO European Institute of Oncology, IRCCS, Milano, Italy.
- Department of Oncology and Hemato-Oncology, University of Milan, Milano, Italy.
Förderung
- Politecnico di Milano
Zitieren
APA
Gugliandolo, S. G., Pillai, S. P., Rajendran, S., Vincini, M. G., Pepa, M., Pansini, F., Zaffaroni, M., Marvaso, G., Alterio, D., Vavassori, A., Durante, S., Volpe, S., Cattani, F., Jereczek-Fossa, B. A., Moscatelli, D., & Colosimo, B. M. (2024). 3D-printed boluses for radiotherapy: influence of geometrical and printing parameters on dosimetric characterization and air gap evaluation. Radiological Physics and Technology. https://doi.org/10.1007/s12194-024-00782-1
Vancouver
Gugliandolo SG, Pillai SP, Rajendran S, Vincini MG, Pepa M, Pansini F, et al. 3D-printed boluses for radiotherapy: influence of geometrical and printing parameters on dosimetric characterization and air gap evaluation. Radiological Physics and Technology. 2024. doi:10.1007/s12194-024-00782-1
BibTeX
@article{Gugliandolo2024,
author = {Gugliandolo, Simone Giovanni and Pillai, Shabarish Purushothaman and Rajendran, Shankar and Vincini, Maria Giulia and Pepa, Matteo and Pansini, Floriana and Zaffaroni, Mattia and Marvaso, Giulia and Alterio, Daniela and Vavassori, Andrea and Durante, Stefano and Volpe, Stefania and Cattani, Federica and Jereczek-Fossa, Barbara Alicja and Moscatelli, Davide and Colosimo, Bianca Maria},
title = {3D-printed boluses for radiotherapy: influence of geometrical and printing parameters on dosimetric characterization and air gap evaluation},
journal = {Radiological Physics and Technology},
year = {2024},
doi = {10.1007/s12194-024-00782-1}
}
EUVIMED