Comparative evaluation of CT-to-Fluoroscopy and Scan-and-Plan registration techniques in robot-assisted pedicle screw placement: a preliminary cadaveric faesibility study

Jacob Oh, Monchai Ruangchainikom, Satish Kripalani, S Vidyadhara, Saw Lim Beng, Wayne Ming Quan Yap, Jitesh Manghwani, Mirant Dave, H S Chhabra

Journal of Robotic Surgery · 13.08.2026 · 20 (1)

DOI: 10.1007/s11701-026-03751-5 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Robotic Surgery, online erschienen am 13.08.2026. Verfasst von Jacob Oh und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Tan Tock Seng Hospital, Singapore, Singapore. Hauptthemen laut MEDLINE: Pedicle Screws, Robotic Surgical Procedures, Tomography, X-Ray Computed. Die Arbeit nennt 17 Literaturangaben. Gefördert von: Medtronic.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

STUDY DESIGN: Preliminary Cadaveric Faesibility study. PURPOSE: To evaluate the preliminary faesibility, technical accuracy, and procedural metrics of CT-to-Fluoroscopy and Scan-and-Plan registration techniques in robot-assisted pedicle screw placement using the Mazor X robotic system. BACKGROUND: Robot-assisted pedicle screw placement has been associated with improved accuracy and safety compared with conventional freehand techniques. However, direct comparisons between the two commonly used robotic registration workflows-CT-to-Fluoroscopy and Scan-and-Plan-remain limited. Evaluating their relative performance is important for optimizing surgical workflow and radiation utilization. METHODOLOGY: Four fresh-frozen male cadavers were allocated to CT-to-Fluoroscopy (n = 2) or Scan-and-Plan (n = 2) registration groups. Eight experienced spine surgeons inserted a total of 32 pedicle screws bilaterally from L2 to L5. Screw accuracy was assessed using the Gertzbein-Robbins classification and trajectory deviation analysis in axial, sagittal, and coronal planes. Procedural time, surgeon-reported ease of use, and radiation exposure were also evaluated. RESULTS: All 32 screws were clinically acceptable (Grade A + B), resulting in 100% overall accuracy in both groups. CT-to-Fluoroscopy demonstrated a higher proportion of Grade A screws than Scan-and-Plan (87.5% vs. 50.0%; p = 0.054). Scan-and-Plan showed significantly lower sagittal angular deviation (0.55 ± 0.71° vs. 1.66 ± 1.14°; p = 0.002), superior clearance (0.20 ± 0.16 mm vs. 0.45 ± 0.36 mm; p = 0.038), and lower coronal angular deviation (0.59 ± 0.76° vs. 1.99 ± 1.74°; p = 0.002). Planning, execution, and overall procedure times were comparable between groups, although Scan-and-Plan required shorter registration time. Surgeon satisfaction was identical (4.75 ± 0.46). Radiation exposure to surgeons was negligible in both groups, while cadaver radiation exposure was higher with CT-to-Fluoroscopy. CONCLUSION: Both registration workflows achieved excellent accuracy. Scan-and-Plan demonstrated advantages in registration efficiency and patient-equivalent radiation exposure while maintaining comparable procedural performance, supporting its use when workflow optimization and radiation reduction are priorities.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pedicle ScrewsRobotic Surgical ProceduresTomography, X-Ray ComputedHumansMaleCadaverFluoroscopyLumbar Vertebrae

Stichwörter der Autor:innen

Robot-assisted pedicle screw placement; CT-to-Fluoroscopy registration; Scan-an…

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, Tan Tock Seng Hospital, Singapore, Singapore.
  • Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
  • Spine Services, Sri Balaji Action Medical Institute, New Delhi, India.
  • Manipal Hospitals, Bengaluru, India.
  • Department of Orthopaedic Surgery, Sunway Medical Centre, Chennai, India.
  • Orthopaedic Spine Surgery, Max Shalimar Bag, New Delhi, India.

Förderung

  • Medtronic

Zitieren

APA

Oh, J., Ruangchainikom, M., Kripalani, S., Vidyadhara, S., Beng, S. L., Yap, W. M. Q., Manghwani, J., Dave, M., & Chhabra, H. S. (2026). Comparative evaluation of CT-to-Fluoroscopy and Scan-and-Plan registration techniques in robot-assisted pedicle screw placement: a preliminary cadaveric faesibility study. Journal of Robotic Surgery, 20(1). https://doi.org/10.1007/s11701-026-03751-5

Vancouver

Oh J, Ruangchainikom M, Kripalani S, Vidyadhara S, Beng SL, Yap WMQ, et al. Comparative evaluation of CT-to-Fluoroscopy and Scan-and-Plan registration techniques in robot-assisted pedicle screw placement: a preliminary cadaveric faesibility study. Journal of Robotic Surgery. 2026;20(1). doi:10.1007/s11701-026-03751-5

BibTeX
@article{Oh2026,
  author = {Oh, Jacob and Ruangchainikom, Monchai and Kripalani, Satish and Vidyadhara, S and Beng, Saw Lim and Yap, Wayne Ming Quan and Manghwani, Jitesh and Dave, Mirant and Chhabra, H S},
  title = {Comparative evaluation of CT-to-Fluoroscopy and Scan-and-Plan registration techniques in robot-assisted pedicle screw placement: a preliminary cadaveric faesibility study},
  journal = {Journal of Robotic Surgery},
  year = {2026},
  volume = {20},
  number = {1},
  doi = {10.1007/s11701-026-03751-5}
}
Kostenlos starten →