Irradiation safety, anesthesia time, surgical complications, and patient-reported outcomes in the robotic Mazor X versus fluoroscopy guided minimally invasive transforaminal lumbar interbody fusion surgery: a comparative cohort study
Neurosurgical Focus · 2024
DOI: 10.3171/2024.9.focus24489 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Neurosurgical Focus, 2024. Verfasst von Daniel W Griepp und 16 weiteren Autor:innen. Beteiligte Einrichtung: 1Division of Neurosurgery, Henry Ford Providence Hospital, Southfield, Michigan; and. Hauptthemen laut MEDLINE: Spinal Fusion, Lumbar Vertebrae, Robotic Surgical Procedures, Patient Reported Outcome Measures. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 68 Literaturangaben.
Abstract (Originalsprache)
OBJECTIVE Robot-assisted (RA) technology is becoming more widely integrated and accepted in spine surgery. The authors sought to evaluate operative and patient-reported outcomes (PROs) in RA versus fluoroscopy-assisted (FA) pedicle screw placement during minimally invasive surgery (MIS) transforaminal lumbar interbody fusion (TLIF). METHODS The authors retrospectively studied elective patients who underwent single- or multilevel MIS TLIF for degenerative indication using FA versus RA pedicle screw placement. Patients were selected from September 2021 to May 2023 at a single institution with multiple surgeons whose practice consists of primarily MIS. Outcomes included fluoroscopy dosage per screw, operative time per screw, anesthesia time per screw, estimated blood loss (EBL), screw revision rate, inpatient surgical complications, and minimal clinically important difference (MCID) of Oswestry Disability Index (ODI) and numeric rating scale (NRS) scores at the 6- and 12-month follow-ups. Comparability of groups was analyzed by univariate analysis. Multivariable analysis modeling fluoroscopy time per screw was performed, adjusting for confounders. RESULTS One hundred eighty-three patients (n = 133 in the FA group vs 50 in the RA group) were included. Patients in the RA cohort were significantly younger than those in the FA group (mean age 63.8 ± 11.9 vs 59.8 ± 11.0 years, p = 0.037). A total of 932 pedicle screws were placed (mean 5.1, range 4–8 per patient). The RA cohort demonstrated significantly lower intraoperative fluoroscopy dosage per screw (4.9 ± 7.6 mGy per screw vs 20.3 ± 14.0 mGy per screw, p < 0.001), significantly longer anesthesia time per screw (49.1 ± 12.6 vs 43.6 ± 9.2, p = 0.009), and similar operative time per screw (33.3 vs 30.7 minutes, p = 0.125). The screw revision rate for symptomatic radiculopathy was zero in both groups. Revision surgery requiring screw removal or reposition was performed in 4 total cases (RA group: 1/50 for infection; FA group: 2/133 for infection, 1/133 for foraminotomy). Both groups demonstrated significant improvement in PROs at 6 and 12 months compared with preoperatively. Moreover, both groups achieved MCID at similar rates. CONCLUSIONS When implementing RA technology, one can expect similar perioperative outcomes as FA techniques in addition to significantly lower radiation exposure. Moreover, there is no statistically significant difference in postoperative PROs between RA and FA. Longer anesthesia times may also be encountered, as in this study, which is likely a result of more complex robotic setup and workflow.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- 1Division of Neurosurgery, Henry Ford Providence Hospital, Southfield, Michigan; and.
- 2Department of Medical Education, Lake Erie College of Medicine, Erie, Pennsylvania.
Zitieren
APA
Griepp, D. W., Caskey, J., Bunjaj, A., Turnbull, J., Alsalahi, A., Alexander, H., Dragonette, J., Sarcar, B., Desai, S., Tong, D., Soo, T. M., Bono, P., Kelkar, P., Houseman, C., Claus, C. F., Richards, B. F., & Carr, D. A. (2024). Irradiation safety, anesthesia time, surgical complications, and patient-reported outcomes in the robotic Mazor X versus fluoroscopy guided minimally invasive transforaminal lumbar interbody fusion surgery: a comparative cohort study. Neurosurgical Focus. https://doi.org/10.3171/2024.9.focus24489
Vancouver
Griepp DW, Caskey J, Bunjaj A, Turnbull J, Alsalahi A, Alexander H, et al. Irradiation safety, anesthesia time, surgical complications, and patient-reported outcomes in the robotic Mazor X versus fluoroscopy guided minimally invasive transforaminal lumbar interbody fusion surgery: a comparative cohort study. Neurosurgical Focus. 2024. doi:10.3171/2024.9.focus24489
BibTeX
@article{Griepp2024,
author = {Griepp, Daniel W and Caskey, Joshua and Bunjaj, Armando and Turnbull, Jeffrey and Alsalahi, Ammar and Alexander, Hepzibha and Dragonette, James and Sarcar, Bryce and Desai, Shivum and Tong, Doris and Soo, Teck M and Bono, Peter and Kelkar, Prashant and Houseman, Clifford and Claus, Chad F and Richards, Boyd F and Carr, Daniel A},
title = {Irradiation safety, anesthesia time, surgical complications, and patient-reported outcomes in the robotic Mazor X versus fluoroscopy guided minimally invasive transforaminal lumbar interbody fusion surgery: a comparative cohort study},
journal = {Neurosurgical Focus},
year = {2024},
doi = {10.3171/2024.9.focus24489}
}
EUVIMED