EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

The contribution of E3D imaging integrated with robotic navigation: analysis of the first 80 consecutive posterior spinal fusion cases

Jeffrey J Stewart, Paul Asdourian, Bradley Moatz, Mosope Soda, Mesfin Lemma, Bryan W Cunningham, Daina M Brooks, Paul C McAfee

Journal of Robotic Surgery · 2024

DOI: 10.1007/s11701-024-02014-5 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Robotic Surgery, 2024. Verfasst von Jeffrey J Stewart und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurosurgery, Georgetown School of Medicine, Washington, DC, USA. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Spinal Fusion, Operative Time, Imaging, Three-Dimensional. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 3 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Eighty consecutive complex spinal robotic cases utilizing intraoperative 3D CT imaging (E3D, Group 2) were compared to 80 age-matched controls using the Excelsius robot alone with C-arm Fluoroscopic registration (Robot Only, Group 1). The demographics between the two groups were similar—severity of deformity, ASA Score for general anesthesia, patient age, gender, number of spinal levels instrumented, number of patients with prior spinal surgery, and amount of neurologic compression. The intraoperative CT scanning added several objective factors improving patient safety. There were significantly fewer complications in the E3D group with only 3 of 80 (4%) patients requiring a return to the operating room compared to 11 of 80 (14%) patients in the Robot Only Group requiring repeat surgery for implant related problems (Chi squared analysis = 5.00, p = 0.025). There was a significant reduction the amount of fluoroscopy time in the E3D Group (36 s, range 4–102 s) compared to Robot only group (51 s, range 15–160 s) ( p = 0.0001). There was also shorter mean operative time in the E3D group (257 ± 59.5 min) compared to the robot only group (306 ± 73.8 min) due to much faster registration time (45 s). A longer registration time was required in the Robot only group to register each vertebral level with AP and Lateral fluoroscopy shots. The estimated blood loss was also significantly lower in Group 2 (mean 345 ± 225 ml) vs Group 1 (474 ± 397 ml) ( p = 0.012). The mean hospital length of stay was also significantly shorter for Group 2 (3.77 ± 1.86 days) compared to Group 1 (5.16 ± 3.40) ( p = 0.022). There was no significant difference in the number of interbody implants nor corrective osteotomies in both groups—Robot only 52 cases vs. 42 cases in E3D group. Level of evidence : IV, Retrospective review.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Robotic Surgical ProceduresSpinal FusionOperative TimeImaging, Three-DimensionalTomography, X-Ray ComputedHumansFemaleMaleMiddle AgedAdultAgedFluoroscopySurgery, Computer-AssistedYoung AdultAged, 80 and overRetrospective StudiesPostoperative Complications

Stichwörter der Autor:innen

Intraoperative CT registrationPosterior spinal fusionRobotic navigationSpinal deformity

Autor:innen und Institutionen

  • Department of Neurosurgery, Georgetown School of Medicine, Washington, DC, USA.
  • Department of Orthopaedic Surgery, MedStar Union Memorial Hospital, 3333 North Calvert Street, Suite 400, Baltimore, MD, 21218, USA.
  • Department of Orthopaedic Surgery, MedStar Union Memorial Hospital, 3333 North Calvert Street, Suite 400, Baltimore, MD, 21218, USA. Mack8132@gmail.com.

Zitieren

APA

Stewart, J. J., Asdourian, P., Moatz, B., Soda, M., Lemma, M., Cunningham, B. W., Brooks, D. M., & McAfee, P. C. (2024). The contribution of E3D imaging integrated with robotic navigation: analysis of the first 80 consecutive posterior spinal fusion cases. Journal of Robotic Surgery. https://doi.org/10.1007/s11701-024-02014-5

Vancouver

Stewart JJ, Asdourian P, Moatz B, Soda M, Lemma M, Cunningham BW, et al. The contribution of E3D imaging integrated with robotic navigation: analysis of the first 80 consecutive posterior spinal fusion cases. Journal of Robotic Surgery. 2024. doi:10.1007/s11701-024-02014-5

BibTeX
@article{Stewart2024,
  author = {Stewart, Jeffrey J and Asdourian, Paul and Moatz, Bradley and Soda, Mosope and Lemma, Mesfin and Cunningham, Bryan W and Brooks, Daina M and McAfee, Paul C},
  title = {The contribution of E3D imaging integrated with robotic navigation: analysis of the first 80 consecutive posterior spinal fusion cases},
  journal = {Journal of Robotic Surgery},
  year = {2024},
  doi = {10.1007/s11701-024-02014-5}
}
Kostenlos starten →