Near-real-time Mueller polarimetric image processing for neurosurgical intervention
International Journal of Computer Assisted Radiology and Surgery · 19.03.2024 · 19 (6) · 1033-1043
DOI: 10.1007/s11548-024-03090-6 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, online erschienen am 19.03.2024. Verfasst von Stefano Moriconi und 9 weiteren Autor:innen. Beteiligte Einrichtung: Support Center for Advanced Neuroimaging (SCAN), University Institute of Diagnostic and Interventional Neuroradiology, Inselspital, Bern University H…. Hauptthemen laut MEDLINE: Neurosurgical Procedures, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung.
Abstract (Originalsprache)
Purpose
Wide-field imaging Mueller polarimetry is a revolutionary, label-free, and non-invasive modality for computer-aided intervention; in neurosurgery, it aims to provide visual feedback of white matter fibre bundle orientation from derived parameters. Conventionally, robust polarimetric parameters are estimated after averaging multiple measurements of intensity for each pair of probing and detected polarised light. Long multi-shot averaging, however, is not compatible with real-time in vivo imaging, and the current performance of polarimetric data processing hinders the translation to clinical practice.
Methods
A learning-based denoising framework is tailored for fast, single-shot, noisy acquisitions of polarimetric intensities. Also, performance-optimised image processing tools are devised for the derivation of clinically relevant parameters. The combination recovers accurate polarimetric parameters from fast acquisitions with near-real-time performance, under the assumption of pseudo-Gaussian polarimetric acquisition noise.
Results
The denoising framework is trained, validated, and tested on experimental data comprising tumour-free and diseased human brain samples in different conditions. Accuracy and image quality indices showed significant ( $$p < 0.05$$ p < 0.05 ) improvements on testing data for a fast single-pass denoising versus the state-of-the-art and high polarimetric image quality standards. The computational time is reported for the end-to-end processing.
Conclusion
The end-to-end image processing achieved real-time performance for a localised field of view ( $$\approx 6.5\ \text {mm}^2$$ ≈ 6.5 mm 2 ). The denoised polarimetric intensities produced visibly clear directional patterns of neuronal fibre tracts in line with reference polarimetric image quality standards; directional disruption was kept in case of neoplastic lesions. The presented advances pave the way towards feasible oncological neurosurgical translations of novel, label-free, interventional feedback.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Support Center for Advanced Neuroimaging (SCAN), University Institute of Diagnostic and Interventional Neuroradiology, Inselspital, Bern University Hospital, University of Bern, 3010, Bern, Switzerland. stefano.moriconi…
- Department of Neurosurgery, Inselspital, Bern University Hospital, University of Bern, 3010, Bern, Switzerland.
- Institute of Tissue Medicine and Pathology, University of Bern, 3008, Bern, Switzerland.
- Institute of Pathology, Lausanne University Hospital, 1011, Lausanne, Switzerland.
- LPICM, CNRS, Ecole Polytechnique, IP Paris, 91120, Palaiseau, France.
- Support Center for Advanced Neuroimaging (SCAN), University Institute of Diagnostic and Interventional Neuroradiology, Inselspital, Bern University Hospital, University of Bern, 3010, Bern, Switzerland.
Förderung
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung
Zitieren
APA
Moriconi, S., Rodríguez-Núñez, O., Gros, É., Felger, L. A., Maragkou, T., Hewer, E., Pierangelo, A., Novikova, T., Schucht, P., & McKinley, R. (2024). Near-real-time Mueller polarimetric image processing for neurosurgical intervention. International Journal of Computer Assisted Radiology and Surgery, 19(6), 1033-1043. https://doi.org/10.1007/s11548-024-03090-6
Vancouver
Moriconi S, Rodríguez-Núñez O, Gros É, Felger LA, Maragkou T, Hewer E, et al. Near-real-time Mueller polarimetric image processing for neurosurgical intervention. International Journal of Computer Assisted Radiology and Surgery. 2024;19(6):1033-1043. doi:10.1007/s11548-024-03090-6
BibTeX
@article{Moriconi2024,
author = {Moriconi, Stefano and Rodríguez-Núñez, Omar and Gros, Éléa and Felger, Leonard A and Maragkou, Theoni and Hewer, Ekkehard and Pierangelo, Angelo and Novikova, Tatiana and Schucht, Philippe and McKinley, Richard},
title = {Near-real-time Mueller polarimetric image processing for neurosurgical intervention},
journal = {International Journal of Computer Assisted Radiology and Surgery},
year = {2024},
volume = {19},
number = {6},
pages = {1033--1043},
doi = {10.1007/s11548-024-03090-6}
}
EUVIMED