Automatic Image Registration Provides Superior Accuracy Compared with Surface Matching in Cranial Navigation
Sensors · 2024
DOI: 10.3390/s24227341 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Henrik Frisk und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Neuroscience, Karolinska Institutet, SE 17177 Stockholm, Sweden. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Neuronavigation, Imaging, Three-Dimensional, Magnetic Resonance Imaging. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Brainlab.
Abstract (Originalsprache)
Objective
The precision of neuronavigation systems relies on the correct registration of the patient’s position in space and aligning it with radiological 3D imaging data. Registration is usually performed by the acquisition of anatomical landmarks or surface matching based on facial features. Another possibility is automatic image registration using intraoperative imaging. This could provide better accuracy, especially in rotated or prone positions where the other methods may be difficult to perform. The aim of this study was to validate automatic image registration (AIR) using intraoperative cone-beam computed tomography (CBCT) for cranial neurosurgical procedures and compare the registration accuracy to the traditional surface matching (SM) registration method based on preoperative MRI. The preservation of navigation accuracy throughout the surgery was also investigated.
Methods
Adult patients undergoing intracranial tumor surgery were enrolled after consent. A standard SM registration was performed, and reference points were acquired. An AIR was then performed, and the same reference points were acquired again. Accuracy was calculated based on the referenced and acquired coordinates of the points for each registration method. The reference points were acquired before and after draping and at the end of the procedure to assess the persistency of accuracy.
Results
In total, 22 patients were included. The mean accuracy was 6.6 ± 3.1 mm for SM registration and 1.0 ± 0.3 mm for AIR. The AIR was superior to the SM registration (p < 0.0001), with a mean improvement in accuracy of 5.58 mm (3.71–7.44 mm 99% CI). The mean accuracy for the AIR registration pre-drape was 1.0 ± 0.3 mm. The corresponding accuracies post-drape and post-resection were 2.9 ± 4.6 mm and 4.1 ± 4.9 mm, respectively. Although a loss of accuracy was identified between the preoperative and end-of-procedure measurements, there was no statistically significant decline during surgery.
Conclusions
AIR for cranial neuronavigation consistently delivered greater accuracy than SM and should be considered the new gold standard for patient registration in cranial neuronavigation. If intraoperative imaging is a limited resource, AIR should be prioritized in rotated or prone position procedures, where the benefits are the greatest.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Clinical Neuroscience, Karolinska Institutet, SE 17177 Stockholm, Sweden.
- Department of Neurosurgery, Karolinska University Hospital, SE 17176 Stockholm, Sweden.
- Clinical Affairs, Brainlab AG, 81829 Munich, Germany.
- Capio Spine Center Stockholm, Löwenströmska Hospital, SE 19489 Upplands-Väsby, Sweden.
- Department of Surgical Sciences, Uppsala University, SE 75236 Uppsala, Sweden.
ORCID-Profile: Henrik Frisk, Victor Gabriel El-Hajj, Erik Edström, Adrian Elmi-Terander
Förderung
- Brainlab
Zitieren
APA
Frisk, H., Jensdottir, M., Coronado, L., Conrad, M., Hager, S., Arvidsson, L., Bartek, J., Burström, G., El-Hajj, V. G., Edström, E., Elmi-Terander, A., & Persson, O. (2024). Automatic Image Registration Provides Superior Accuracy Compared with Surface Matching in Cranial Navigation. Sensors. https://doi.org/10.3390/s24227341
Vancouver
Frisk H, Jensdottir M, Coronado L, Conrad M, Hager S, Arvidsson L, et al. Automatic Image Registration Provides Superior Accuracy Compared with Surface Matching in Cranial Navigation. Sensors. 2024. doi:10.3390/s24227341
BibTeX
@article{Frisk2024,
author = {Frisk, Henrik and Jensdottir, Margret and Coronado, Luisa and Conrad, Markus and Hager, Susanne and Arvidsson, Lisa and Bartek, Jiri and Burström, Gustav and El-Hajj, Victor Gabriel and Edström, Erik and Elmi-Terander, Adrian and Persson, Oscar},
title = {Automatic Image Registration Provides Superior Accuracy Compared with Surface Matching in Cranial Navigation},
journal = {Sensors},
year = {2024},
doi = {10.3390/s24227341}
}
EUVIMED