Design consideration on integration of mechanical intravascular ultrasound and electromagnetic tracking sensor for intravascular reconstruction

Wenran Cai, Kazuaki Hara, Naoki Tomii, Etsuko Kobayashi, Takashi Ohya, Ichiro Sakuma

International Journal of Computer Assisted Radiology and Surgery · 2024

DOI: 10.1007/s11548-024-03059-5 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, 2024. Verfasst von Wenran Cai und 5 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Engineering, The University of Tokyo, Tokyo, Japan. cai@bmpe.t.u-tokyo.ac.jp. Hauptthemen laut MEDLINE: Ultrasonography, Interventional, Phantoms, Imaging, Equipment Design, Electromagnetic Phenomena. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Grant-in-Aid for Scientific Research (B), The University of Tokyo.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Considering vessel deformation, endovascular navigation requires intraoperative geometric information. Mechanical intravascular ultrasound (IVUS) with an electromagnetic (EM) sensor can be used to reconstruct blood vessels with thin diameter. However, the integration design should be evaluated based on the factors affecting the reconstruction error.

Methods

The interference between the mechanical IVUS and EM sensor was measured in different relative positions. Two designs of the integrated catheter were evaluated by measuring the reconstruction errors using a rigid vascular phantom.

Results

When the distance from the EM sensor to the field generator was 75 mm, the interference from mechanical IVUS to an EM sensor was negligible, with position and rotation errors less than 0.1 mm and 0.6°, respectively. The reconstructed vessel model for proximal IVUS transducer had a smooth surface but an inaccurate shape at large curvature of the vascular phantom. When the distance to the field generator was 175 mm, the error increased significantly.

Conclusion

Placing the IVUS transducer on the proximal side of the EM sensor is superior in terms of interference reduction but inferior in terms of mechanical stability compared to a distal transducer. The distal side is preferred due to better mechanical stability during catheter manipulation at larger curvature. With this configuration, surface reconstruction errors less than 1.7 mm (with RMS 0.57 mm) were achieved when the distance to the field generator was less than 175 mm.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Ultrasonography, InterventionalPhantoms, ImagingEquipment DesignElectromagnetic PhenomenaHumansTransducers

Stichwörter der Autor:innen

3D intravascular reconstructionElectromagnetic (EM) sensorEndovascular navigationIntravascular ultrasound (IVUS)

Autor:innen und Institutionen

  • Graduate School of Engineering, The University of Tokyo, Tokyo, Japan. cai@bmpe.t.u-tokyo.ac.jp.
  • Graduate School of Engineering, The University of Tokyo, Tokyo, Japan.
  • Department of Oral and Maxillofacial Surgery, Yokohama City University Graduate School of Medicine, Yokohama, Japan.

Förderung

  • Grant-in-Aid for Scientific Research (B)
  • The University of Tokyo

Zitieren

APA

Cai, W., Hara, K., Tomii, N., Kobayashi, E., Ohya, T., & Sakuma, I. (2024). Design consideration on integration of mechanical intravascular ultrasound and electromagnetic tracking sensor for intravascular reconstruction. International Journal of Computer Assisted Radiology and Surgery. https://doi.org/10.1007/s11548-024-03059-5

Vancouver

Cai W, Hara K, Tomii N, Kobayashi E, Ohya T, Sakuma I. Design consideration on integration of mechanical intravascular ultrasound and electromagnetic tracking sensor for intravascular reconstruction. International Journal of Computer Assisted Radiology and Surgery. 2024. doi:10.1007/s11548-024-03059-5

BibTeX
@article{Cai2024,
  author = {Cai, Wenran and Hara, Kazuaki and Tomii, Naoki and Kobayashi, Etsuko and Ohya, Takashi and Sakuma, Ichiro},
  title = {Design consideration on integration of mechanical intravascular ultrasound and electromagnetic tracking sensor for intravascular reconstruction},
  journal = {International Journal of Computer Assisted Radiology and Surgery},
  year = {2024},
  doi = {10.1007/s11548-024-03059-5}
}
Kostenlos starten →