Image-quality optimization for late iodine enhancement with photon-counting CT: impact of spectral analysis and reconstruction parameters
La radiologia medica · 2026
DOI: 10.1007/s11547-026-02260-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift La radiologia medica, 2026. Verfasst von Elisa Bruno und 5 weiteren Autor:innen. Beteiligte Einrichtung: Advanced Imaging for Personalized Medicine Unit, Experimental Imaging Center, IRCCS San Raffaele Hospital, Via Olgettina 60, 20132, Milan, Italy. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Contrast Media, Cicatrix, Radiographic Image Interpretation, Computer-Assisted. Die Arbeit nennt 32 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: Clinical application of late iodine enhancement (LIE) for myocardial scar identification is limited by low contrast-to-noise ratio (CNR). Photon-counting detector-CT (PCD-CT) can improve image quality providing spectral information and low image noise. The aim of the study was to assess the impact of spectral analysis and reconstruction parameters on LIE image quality using PCD-CT. MATERIAL AND METHODS: This single-center retrospective study included 74 patients undergoing PCD-CT between May 2024 and July 2025 for myocardial scar evaluation. Images from patients with visible LIE were reconstructed using four parameter combinations (Qr40 and Qr36 kernels; 0.4-mm and 2-mm slice thickness) and analyzed at monoenergetic levels (40-120 keV, 10 keV increment) using a dedicated workstation. CNR and signal-to-noise ratio (SNR) were calculated for each reconstruction. RESULTS: Among 74 patients, 41 (55%) had myocardial scar (M:F 32:9; 62 years [IQR,56-71]) and 24/41 (59%) had an implantable cardioverter defibrillator (ICD). Scars were predominantly transmural (21/41, 51%). A 40-keV VMI achieved the highest CNR and SNR, regardless of reconstruction parameters (p 0.05). In the exploratory subgroup of 16 patients undergoing LGE-MRI, CNR is not statistically different between LIE-CT at 40 keV with 2-mm slice thickness from LGE-MRI (5.36 [IQR,4.07-6.88] vs. 8.04 [IQR,6.67-9.89]; p = 0.178). CONCLUSION: PCD-CT LIE demonstrated the highest CNR at 40-keV VMI with 2-mm slice thickness, regardless of scar pattern, BMI, or ICD presence.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Advanced Imaging for Personalized Medicine Unit, Experimental Imaging Center, IRCCS San Raffaele Hospital, Via Olgettina 60, 20132, Milan, Italy.
- School of Medicine, Vita-Salute San Raffaele University, Via Olgettina 58, 20132, Milan, Italy.
- Advanced Imaging for Personalized Medicine Unit, Experimental Imaging Center, IRCCS San Raffaele Hospital, Via Olgettina 60, 20132, Milan, Italy. palmisano.anna@hsr.it.
- School of Medicine, Vita-Salute San Raffaele University, Via Olgettina 58, 20132, Milan, Italy. palmisano.anna@hsr.it.
Zitieren
APA
Bruno, E., Palmisano, A., Pisu, F., Bettinelli, A., Vignale, D., & Esposito, A. (2026). Image-quality optimization for late iodine enhancement with photon-counting CT: impact of spectral analysis and reconstruction parameters. La radiologia medica. https://doi.org/10.1007/s11547-026-02260-7
Vancouver
Bruno E, Palmisano A, Pisu F, Bettinelli A, Vignale D, Esposito A. Image-quality optimization for late iodine enhancement with photon-counting CT: impact of spectral analysis and reconstruction parameters. La radiologia medica. 2026. doi:10.1007/s11547-026-02260-7
BibTeX
@article{Bruno2026,
author = {Bruno, Elisa and Palmisano, Anna and Pisu, Francesco and Bettinelli, Andrea and Vignale, Davide and Esposito, Antonio},
title = {Image-quality optimization for late iodine enhancement with photon-counting CT: impact of spectral analysis and reconstruction parameters},
journal = {La radiologia medica},
year = {2026},
doi = {10.1007/s11547-026-02260-7}
}
EUVIMED