Heart rate sensitivity of virtual non-contrast calcium scores derived from photon counting detector CT data: a phantom study
La radiologia medica · 06.02.2024 · 129 (3) · 401-410
DOI: 10.1007/s11547-024-01773-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift La radiologia medica, online erschienen am 06.02.2024. Verfasst von Franka Risch und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology, University Hospital Augsburg, Stenglinstr. 2, 86156, Augsburg, Germany. Hauptthemen laut MEDLINE: Calcium, Tomography, X-Ray Computed. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universität Augsburg.
Abstract (Originalsprache)
Purpose
To assess the reliability of virtual non-contrast (VNC) derived coronary artery calcium quantities in relation to heart rate and the VNC algorithm used compared to reference true non-contrast (TNC), considering several clinically established acquisition modes. Material and methods An ad hoc built coronary phantom containing four calcified lesions and an iodinated lumen was scanned using three cardiac acquisition modes three times within an anthropomorphic cardiac motion phantom simulating different heart rates (0, 60, 80, 100 bpm) and reconstructed with a conventional (VNC conv ) and a calcium-sensitive (VNC pc ) VNC algorithm. TNC reference was scanned at 0 bpm with non-iodinated lumen. Calcium scores were assessed in terms of number of lesions detected, Agatston and volume scores and global noise was measured. Paired t-test and Wilcoxon test were performed to test measurements for significant difference.
Results
For both VNC algorithms used, calcium levels or noise were not significantly affected by heart rate. Measurements on VNC pc reconstructions best reproduced TNC results, but with increased variability (Agatston scores at 0 bpm for TNC, VNC conv , and VNC pc were 47.1 ± 1.1, 6.7 ± 2.8 ( p < 0.001), and 45.3 ± 7.6 ( p > 0.05), respectively). VNC reconstructions showed lower noise levels compared to TNC, especially for VNC pc (noise heart on TNC, VNC conv and VNC pc at 0 bpm was 5.0 ± 0.4, 4.5 ± 0.2, 4.2 ± 0.2).
Conclusion
No significant heart rate dependence of VNC-based calcium scores was observed in an intra-reconstruction comparison. VNC pc reproduces TNC scores better than VNC conv without significant differences and decreased noise, however, with an increasing average deviation with rising heart rates. VNC-based CACS should be used with caution as the measures show higher variability compared to reference TNC and therefore hold the potential of incorrect risk categorization.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Diagnostic and Interventional Radiology, University Hospital Augsburg, Stenglinstr. 2, 86156, Augsburg, Germany.
- Medical Faculty, Ludwig Maximilian University Munich, Munich, Germany.
- Clinic for Diagnostic and Interventional Radiology, Donau-Isar-Klinikum, Deggendorf, Germany.
- Department of Diagnostic and Interventional Radiology, University Hospital Augsburg, Stenglinstr. 2, 86156, Augsburg, Germany. Thomas.Kroencke@uk-augsburg.de.
- Centre for Advanced Analytics and Predictive Sciences (CAAPS), University Augsburg, Augsburg, Germany. Thomas.Kroencke@uk-augsburg.de.
Förderung
- Universität Augsburg
Zitieren
APA
Risch, F., Schwarz, F., Kroencke, T., & Decker, J. A. (2024). Heart rate sensitivity of virtual non-contrast calcium scores derived from photon counting detector CT data: a phantom study. La radiologia medica, 129(3), 401-410. https://doi.org/10.1007/s11547-024-01773-3
Vancouver
Risch F, Schwarz F, Kroencke T, Decker JA. Heart rate sensitivity of virtual non-contrast calcium scores derived from photon counting detector CT data: a phantom study. La radiologia medica. 2024;129(3):401-410. doi:10.1007/s11547-024-01773-3
BibTeX
@article{Risch2024,
author = {Risch, Franka and Schwarz, Florian and Kroencke, Thomas and Decker, Josua A},
title = {Heart rate sensitivity of virtual non-contrast calcium scores derived from photon counting detector CT data: a phantom study},
journal = {La radiologia medica},
year = {2024},
volume = {129},
number = {3},
pages = {401--410},
doi = {10.1007/s11547-024-01773-3}
}
EUVIMED