EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

Performance of a Deep Learning Reconstruction Method on Clinical Chest–Abdomen–Pelvis Scans from a Dual-Layer Detector CT System

Christopher Schuppert iD, Stefanie Rahn iD, Nikolas D Schnellbächer iD, Frank Bergner, Michael Grass, Hans-Ulrich Kauczor, Stephan Skornitzke iD, Tim F Weber, Thuy D Do iD

Tomography · 2025

DOI: 10.3390/tomography11090094 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Tomography, 2025. Verfasst von Christopher Schuppert und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology, Medical Center, Faculty of Medicine, University of Freiburg, 79106 Freiburg, Germany. Hauptthemen laut MEDLINE: Deep Learning, Tomography, X-Ray Computed, Pelvis, Radiographic Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Philips Healthcare.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The objective of this study was to compare the performance and robustness of a deep learning reconstruction method against established alternatives for soft tissue CT image reconstruction. Materials and Methods: Images were generated from portal venous phase chest–abdomen–pelvis CT scans (n = 99) acquired on a dual-layer spectral detector CT using filtered back projection, iterative model reconstruction (IMR), and deep learning reconstruction (DLR) with three parameter settings, namely ‘standard’, ‘sharper’, and ‘smoother’. Experienced raters performed a quantitative assessment by considering attenuation stability and image noise levels in ten representative structures across all reconstruction methods, as well as a qualitative assessment using a four-point Likert scale (1 = poor, 2 = fair, 3 = good, 4 = excellent) for their overall perception of ‘smoother’ DLR and IMR images. One scan was excluded due to cachexia, which limited the quantitative measurements.

Results

The inter-rater reliability for quantitative measurements ranged from moderate to excellent (r = 0.63–0.96). Attenuation values did not differ significantly between reconstruction methods except for DLR against IMR in the psoas muscle (mean + 3.0 HU, p < 0.001). Image noise levels differed significantly between reconstruction methods for all structures (all p < 0.001) and were lower than FBP with any DLR parameter setting. Image noise levels with ‘smoother’ DLR were predominantly lower than or equal to IMR, while they were higher with ‘standard’ DLR and ‘sharper’ DLR. The ‘smoother’ DLR images received a higher mean rating for overall image quality than the IMR images (3.7 vs. 2.3, p < 0.001). Conclusions: ‘Smoother’ DLR images were perceived by experienced readers as having improved quality compared to FBP and IMR while also exhibiting objectively lower or equivalent noise levels.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Deep LearningTomography, X-Ray ComputedPelvisRadiographic Image Interpretation, Computer-AssistedRadiography, ThoracicRadiography, AbdominalImage Processing, Computer-AssistedHumansMaleReproducibility of ResultsFemaleMiddle AgedAgedAdultAged, 80 and overAbdomenThorax

Stichwörter der Autor:innen

computed tomographydeep learning reconstructiondenoisingimage noiseimage quality

Autor:innen und Institutionen

  • Department of Diagnostic and Interventional Radiology, Medical Center, Faculty of Medicine, University of Freiburg, 79106 Freiburg, Germany.
  • Clinic of Diagnostic and Interventional Radiology, Heidelberg University Hospital, 69120 Heidelberg, Germany.
  • Translational Lung Research Center Heidelberg (TLRC), German Center for Lung Research (DZL), Heidelberg University Hospital, 69120 Heidelberg, Germany.
  • Philips GmbH Innovative Technologies, 22335 Hamburg, Germany.

ORCID-Profile: Christopher Schuppert, Stefanie Rahn, Nikolas D Schnellbächer, Stephan Skornitzke, Thuy D Do

Förderung

  • Philips Healthcare

Zitieren

APA

Schuppert, C., Rahn, S., Schnellbächer, N. D., Bergner, F., Grass, M., Kauczor, H. U., Skornitzke, S., Weber, T. F., & Do, T. D. (2025). Performance of a Deep Learning Reconstruction Method on Clinical Chest–Abdomen–Pelvis Scans from a Dual-Layer Detector CT System. Tomography. https://doi.org/10.3390/tomography11090094

Vancouver

Schuppert C, Rahn S, Schnellbächer ND, Bergner F, Grass M, Kauczor HU, et al. Performance of a Deep Learning Reconstruction Method on Clinical Chest–Abdomen–Pelvis Scans from a Dual-Layer Detector CT System. Tomography. 2025. doi:10.3390/tomography11090094

BibTeX
@article{Schuppert2025,
  author = {Schuppert, Christopher and Rahn, Stefanie and Schnellbächer, Nikolas D and Bergner, Frank and Grass, Michael and Kauczor, Hans-Ulrich and Skornitzke, Stephan and Weber, Tim F and Do, Thuy D},
  title = {Performance of a Deep Learning Reconstruction Method on Clinical Chest–Abdomen–Pelvis Scans from a Dual-Layer Detector CT System},
  journal = {Tomography},
  year = {2025},
  doi = {10.3390/tomography11090094}
}
Kostenlos starten →