EUVIMEDEUVIMED entdecken
EvaluationsstudieOpen Access · CC BY 4.0

Implementation of the first practical automatic micropulse blood pressure measurement device using the published revolutionary Newcastle technique

Alan Murray, Dingchang Zheng, Clive J Griffiths, Chengyu Liu, David J Graham, Jeffrey Neasham, Adrian J Cossor

Medical & Biological Engineering & Computing · 2026

DOI: 10.1007/s11517-026-03687-w ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Medical & Biological Engineering & Computing, 2026. Verfasst von Alan Murray und 6 weiteren Autor:innen. Beteiligte Einrichtung: School of Engineering & Faculty of Medical Sciences, Newcastle University, Newcastle upon Tyne, UK. alan.murray@ncl.ac.uk. Hauptthemen laut MEDLINE: Blood Pressure Determination, Auscultation. Die Arbeit nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Engineering and Physical Sciences Research Council.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

True accuracy for blood pressure measurement devices is required, but current techniques only estimate, deriving measurements from various private sets of clinical measurements. We previously reported a new technique that detects micropulse arterial opening, present only for beats between systole and diastole, enabling direct determination of systolic and diastolic blood pressures (SBP & DBP). We have now developed a first automated device using our technique and have compared its accuracy with current gold standard auscultatory measurements in 32 subjects. The differences between our novel device and simultaneous auscultatory measurements were for SBP -2.5 ± 3.1 mmHg, and for DBP -1.0 ± 1.2 mmHg (mean ± standard deviation SD). Results were well within the clinical international standard’s limits of 5 ± 8 mmHg. Graphical Abstract

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood Pressure DeterminationAuscultationHumansReproducibility of ResultsBlood PressureSystoleDiastoleReference ValuesHealthy VolunteersMaleFemaleYoung AdultAdultMiddle AgedAgedFeasibility Studies

Stichwörter der Autor:innen

Blood pressure measurementsClinical diagnostic measurementMedical deviceMicropulse technique

Autor:innen und Institutionen

  • School of Engineering & Faculty of Medical Sciences, Newcastle University, Newcastle upon Tyne, UK. alan.murray@ncl.ac.uk.
  • Centre for Healthcare and Community Transformation, Coventry University, Coventry, UK.
  • School of Engineering & Faculty of Medical Sciences, Newcastle University, Newcastle upon Tyne, UK.
  • School of Instrument Science and Engineering, Southeast University, Nanjing, China.
  • School of Engineering, Newcastle University, Newcastle upon Tyne, UK.
  • AC Cossor & Son (Technology) Ltd, Anstey Bury, Anstey, Buntingford, UK.

Förderung

  • Engineering and Physical Sciences Research Council

Zitieren

APA

Murray, A., Zheng, D., Griffiths, C. J., Liu, C., Graham, D. J., Neasham, J., & Cossor, A. J. (2026). Implementation of the first practical automatic micropulse blood pressure measurement device using the published revolutionary Newcastle technique. Medical & Biological Engineering & Computing. https://doi.org/10.1007/s11517-026-03687-w

Vancouver

Murray A, Zheng D, Griffiths CJ, Liu C, Graham DJ, Neasham J, et al. Implementation of the first practical automatic micropulse blood pressure measurement device using the published revolutionary Newcastle technique. Medical & Biological Engineering & Computing. 2026. doi:10.1007/s11517-026-03687-w

BibTeX
@article{Murray2026,
  author = {Murray, Alan and Zheng, Dingchang and Griffiths, Clive J and Liu, Chengyu and Graham, David J and Neasham, Jeffrey and Cossor, Adrian J},
  title = {Implementation of the first practical automatic micropulse blood pressure measurement device using the published revolutionary Newcastle technique},
  journal = {Medical & Biological Engineering & Computing},
  year = {2026},
  doi = {10.1007/s11517-026-03687-w}
}
Kostenlos starten →