EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY-NC 4.0

Comparisons of an automated oscillometric device with a hybrid manual auscultatory device for the Korea National Health and Nutrition Examination Survey

Yu-Mi Kim iD, Sang Hyeon Park, Jinho Shin iD, Ki-Chul Sung iD, Sang-Hyun Ihm, Seong Heon Kim, Dae-Hee Kim, Sang Min Park, In Jeong Cho, Kyung Won Oh, Eun Mi Lee

The Journal of Clinical Hypertension · 2024

DOI: 10.1111/jch.14805 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The Journal of Clinical Hypertension, 2024. Verfasst von Yu-Mi Kim und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Preventive Medicine, College of Medicine, Hanyang University, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Nutrition Surveys, Blood Pressure Determination, Oscillometry, Auscultation. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: the Korea Disease Control and Prevention Agency, the Wonkwang University 2022.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study evaluated an oscillometric device (OD), Microlife WatchBP Office AFIB, and a hybrid manual auscultatory device (AD), Greenlight 300TM, to determine a suitable blood pressure (BP) measurement device for the Korea National Health and Nutrition Examination Survey in a mercury‐free context. Adhering to the 2018 Universal Standard's suggested consensus, the study involved 800 subjects (mean age 51.2 ± 17.5 years; 44.3% male), who underwent triplicate BP measurements following 5 min of rest in a randomized order (OD‐first: 398 participants; AD‐first: 402 participants). BP difference was calculated as OD value minus AD value, with results stratified by measurement sequence. The overall BP difference and tolerable error probability were ‐1.1 ± 6.5/‐2.6 ± 4.9 mmHg and 89.2%/92.5% for systolic/diastolic BP (SBP/DBP), respectively. Lin's concordance correlation coefficient was 0.907/0.844 for SBP/DBP (OD‐first/AD‐first: 0.925/0.892 for SBP, 0.842/0.845 for DBP). The overall agreement for hypertension (BP ≥ 140 and/or 90 mmHg) was 0.71 ( p < 0.0001), and the OD underestimated the overall hypertension prevalence by 5.1%. Analysis of the AD‐first data revealed a lower level of agreement compared to the OD‐first data; however, the observed blood pressure difference adhered to Criterion 1 of the 2018 Universal Standard. Microlife met the Criterion 1 of 2018 Universal Standard but underestimated the prevalence of hypertension. The BP discrepancy increased with higher BP levels, male sex, and smaller AC. With increasing age, the discrepancy decreased for SBP and increased for DBP.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Nutrition SurveysBlood Pressure DeterminationOscillometryAuscultationHumansMaleMiddle AgedFemaleRepublic of KoreaAdultAgedHypertensionBlood PressureReproducibility of Results

Stichwörter der Autor:innen

agreementblood pressure determinationhealth surveyhypertensionsphygmomanometers

Autor:innen und Institutionen

  • Department of Preventive Medicine, College of Medicine, Hanyang University, Seoul, Republic of Korea.
  • Graduate School of Public Health, Hanyang University, Seoul, Republic of Korea.
  • Division of Cardiology, Department of Internal Medicine, Wonkwang University Sanbon Hospital, Gunpo, Gyeonggi-do, Republic of Korea.
  • Division of Cardiology, Department of Internal Medicine, College of Medicine, Hanyang University, Seoul, Republic of Korea.
  • Division of Cardiology, Department of Internal Medicine, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Division of Cardiology, Department of Internal Medicine, The Catholic University of Korea, Seoul, Republic of Korea.

ORCID-Profile: Yu-Mi Kim, Jinho Shin, Ki-Chul Sung

Förderung

  • the Korea Disease Control and Prevention Agency
  • the Wonkwang University 2022

Zitieren

APA

Kim, Y. M., Park, S. H., Shin, J., Sung, K. C., Ihm, S. H., Kim, S. H., Kim, D. H., Park, S. M., Cho, I. J., Oh, K. W., & Lee, E. M. (2024). Comparisons of an automated oscillometric device with a hybrid manual auscultatory device for the Korea National Health and Nutrition Examination Survey. The Journal of Clinical Hypertension. https://doi.org/10.1111/jch.14805

Vancouver

Kim YM, Park SH, Shin J, Sung KC, Ihm SH, Kim SH, et al. Comparisons of an automated oscillometric device with a hybrid manual auscultatory device for the Korea National Health and Nutrition Examination Survey. The Journal of Clinical Hypertension. 2024. doi:10.1111/jch.14805

BibTeX
@article{Kim2024,
  author = {Kim, Yu-Mi and Park, Sang Hyeon and Shin, Jinho and Sung, Ki-Chul and Ihm, Sang-Hyun and Kim, Seong Heon and Kim, Dae-Hee and Park, Sang Min and Cho, In Jeong and Oh, Kyung Won and Lee, Eun Mi},
  title = {Comparisons of an automated oscillometric device with a hybrid manual auscultatory device for the Korea National Health and Nutrition Examination Survey},
  journal = {The Journal of Clinical Hypertension},
  year = {2024},
  doi = {10.1111/jch.14805}
}
Kostenlos starten →