Impact of harmonization on the reproducibility of MRI radiomic features when using different scanners, acquisition parameters, and image pre-processing techniques: a phantom study

Ghasem Hajianfar, Seyyed Ali Hosseini, Sara Bagherieh, Mehrdad Oveisi, Isaac Shiri, Habib Zaidi

Medical & Biological Engineering & Computing · 2024-08 · 62 (8) · 2319-2332

DOI: 10.1007/s11517-024-03071-6 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical & Biological Engineering & Computing, online erschienen am 27.03.2024. Verfasst von Ghasem Hajianfar und 5 weiteren Autor:innen. Beteiligte Einrichtung: Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, CH-1211, Geneva, Switzerland. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 29-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung, University of Geneva.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study investigated the impact of ComBat harmonization on the reproducibility of radiomic features extracted from magnetic resonance images (MRI) acquired on different scanners, using various data acquisition parameters and multiple image pre-processing techniques using a dedicated MRI phantom. Four scanners were used to acquire an MRI of a nonanatomic phantom as part of the TCIA RIDER database. In fast spin-echo inversion recovery (IR) sequences, several inversion durations were employed, including 50, 100, 250, 500, 750, 1000, 1500, 2000, 2500, and 3000 ms. In addition, a 3D fast spoiled gradient recalled echo (FSPGR) sequence was used to investigate several flip angles (FA): 2, 5, 10, 15, 20, 25, and 30 degrees. Nineteen phantom compartments were manually segmented. Different approaches were used to pre-process each image: Bin discretization, Wavelet filter, Laplacian of Gaussian, logarithm, square, square root, and gradient. Overall, 92 first-, second-, and higher-order statistical radiomic features were extracted. ComBat harmonization was also applied to the extracted radiomic features. Finally, the Intraclass Correlation Coefficient (ICC) and Kruskal-Wallis’s (KW) tests were implemented to assess the robustness of radiomic features. The number of non-significant features in the KW test ranged between 0–5 and 29–74 for various scanners, 31–91 and 37–92 for three times tests, 0–33 to 34–90 for FAs, and 3–68 to 65–89 for IRs before and after ComBat harmonization, with different image pre-processing techniques, respectively. The number of features with ICC over 90% ranged between 0–8 and 6–60 for various scanners, 11–75 and 17–80 for three times tests, 3–83 to 9–84 for FAs, and 3–49 to 3–63 for IRs before and after ComBat harmonization, with different image pre-processing techniques, respectively. The use of various scanners, IRs, and FAs has a great impact on radiomic features. However, the majority of scanner-robust features is also robust to IR and FA. Among the effective parameters in MR images, several tests in one scanner have a negligible impact on radiomic features. Different scanners and acquisition parameters using various image pre-processing might affect radiomic features to a large extent. ComBat harmonization might significantly impact the reproducibility of MRI radiomic features. Graphical Abstract

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingMagnetic Resonance ImagingImage Processing, Computer-AssistedReproducibility of ResultsHumansAlgorithmsRadiomics

Stichwörter der Autor:innen

HarmonizationMRIPre-processingRadiomicsRobustnessScanner effect

Autor:innen und Institutionen

  • Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, CH-1211, Geneva, Switzerland.
  • Translational Neuroimaging Laboratory, McGill University Research Centre for Studies in Aging, Douglas Hospital, McGill University, Montréal, Québec, Canada.
  • Department of Neurology and Neurosurgery, Faculty of Medicine, McGill University, Montréal, Québec, Canada.
  • School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran.
  • Department of Computer Science, University of British Columbia, Vancouver, BC, Canada.
  • Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.

Förderung

  • Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung
  • University of Geneva

Zitieren

APA

Hajianfar, G., Hosseini, S. A., Bagherieh, S., Oveisi, M., Shiri, I., & Zaidi, H. (2024). Impact of harmonization on the reproducibility of MRI radiomic features when using different scanners, acquisition parameters, and image pre-processing techniques: a phantom study. Medical & Biological Engineering & Computing, 62(8), 2319-2332. https://doi.org/10.1007/s11517-024-03071-6

Vancouver

Hajianfar G, Hosseini SA, Bagherieh S, Oveisi M, Shiri I, Zaidi H. Impact of harmonization on the reproducibility of MRI radiomic features when using different scanners, acquisition parameters, and image pre-processing techniques: a phantom study. Medical & Biological Engineering & Computing. 2024;62(8):2319-2332. doi:10.1007/s11517-024-03071-6

BibTeX
@article{Hajianfar2024,
  author = {Hajianfar, Ghasem and Hosseini, Seyyed Ali and Bagherieh, Sara and Oveisi, Mehrdad and Shiri, Isaac and Zaidi, Habib},
  title = {Impact of harmonization on the reproducibility of MRI radiomic features when using different scanners, acquisition parameters, and image pre-processing techniques: a phantom study},
  journal = {Medical & Biological Engineering & Computing},
  year = {2024},
  volume = {62},
  number = {8},
  pages = {2319--2332},
  doi = {10.1007/s11517-024-03071-6}
}
Kostenlos starten →