Validation of the Sleepiz One + as a radar-based sensor for contactless diagnosis of sleep apnea
Sleep and Breathing · 2024
DOI: 10.1007/s11325-024-03057-6 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Sleep and Breathing, 2024. Verfasst von Jonas Alexander Gross-Isselmann und 5 weiteren Autor:innen. Beteiligte Einrichtung: Devision of Sleep & Telemedicine, Ruhrlandklinik, University Medicine Essen, University of Duisburg-Essen, Essen, Germany. jonas.gross-isselmann@stud…. Hauptthemen laut MEDLINE: Polysomnography, Radar, Sleep Apnea Syndromes. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum Essen.
Abstract (Originalsprache)
Purpose
The cardiorespiratory polysomnography (PSG) is an expensive and limited resource. The Sleepiz One + is a novel radar-based contactless monitoring device that can be used e.g. for longitudinal detection of nocturnal respiratory events. The present study aimed to compare the performance of the Sleepiz One + device to the PSG regarding the accuracy of apnea–hypopnea index (AHI).
Methods
From January to December 2021, a total of 141 adult volunteers who were either suspected of having sleep apnea or who were healthy sleepers took part in a sleep study. This examination served to validate the Sleepiz One + device in the presence and absence of additional SpO2 information. The AHI determined by the Sleepiz One + monitor was estimated automatically and compared with the AHI derived from manual PSG scoring.
Results
The correlation between the Sleepiz-AHI and the PSG-AHI with and without additional SpO2 measurement was r p = 0.94 and r p = 0,87, respectively. In general, the Bland–Altman plots showed good agreement between the two methods of AHI measurement, though their deviations became larger with increasing sleep-disordered breathing. Sensitivity and specificity for recordings without additional SpO2 was 85% and 88%, respectively. Adding a SpO2 sensor increased the sensitivity to 88% and the specificity to 98%.
Conclusion
The Sleepiz One + device is a valid diagnostic tool for patients with moderate to severe OSA. It can also be easily used in the home environment and is therefore beneficial for e.g. immobile and infectious patients. Trial registration number and date of registration for prospectively registered trials This study was registered on clinicaltrials.gov (NCT04670848) on 2020–12-09.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Devision of Sleep & Telemedicine, Ruhrlandklinik, University Medicine Essen, University of Duisburg-Essen, Essen, Germany. jonas.gross-isselmann@stud.uni-due.de.
- Devision of Sleep & Telemedicine, Ruhrlandklinik, University Medicine Essen, University of Duisburg-Essen, Essen, Germany.
ORCID-Profile: Jonas Alexander Gross-Isselmann, Torsten Eggert, Alina Wildenauer, Sarah Dietz-Terjung, Christoph Schoebel
Förderung
- Universitätsklinikum Essen
Zitieren
APA
Gross-Isselmann, J. A., Eggert, T., Wildenauer, A., Dietz-Terjung, S., Grosse Sundrup, M., & Schoebel, C. (2024). Validation of the Sleepiz One + as a radar-based sensor for contactless diagnosis of sleep apnea. Sleep and Breathing. https://doi.org/10.1007/s11325-024-03057-6
Vancouver
Gross-Isselmann JA, Eggert T, Wildenauer A, Dietz-Terjung S, Grosse Sundrup M, Schoebel C. Validation of the Sleepiz One + as a radar-based sensor for contactless diagnosis of sleep apnea. Sleep and Breathing. 2024. doi:10.1007/s11325-024-03057-6
BibTeX
@article{GrossIsselmann2024,
author = {Gross-Isselmann, Jonas Alexander and Eggert, Torsten and Wildenauer, Alina and Dietz-Terjung, Sarah and Grosse Sundrup, Martina and Schoebel, Christoph},
title = {Validation of the Sleepiz One + as a radar-based sensor for contactless diagnosis of sleep apnea},
journal = {Sleep and Breathing},
year = {2024},
doi = {10.1007/s11325-024-03057-6}
}
EUVIMED