Fachartikel in der Zeitschrift Pharmaceutical Research, 2026. Verfasst von Abinaya Nallathambi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Drug Sciences, University of Pavia, 27100, Pavia, Italy. Hauptthemen laut MEDLINE: Sterilization, Hydrogen Peroxide, Biocompatible Materials, Prostheses and Implants. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: HORIZON EUROPE Marie Sklodowska-Curie Actions, Università degli Studi di Pavia.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To develop a deep vacuum vaporized hydrogen peroxide (VH₂O₂) sterilization method capable of preserving the structural and functional integrity of biocompatible polymeric electrospun mats usable as advanced biocompatible medical implants and devices, with particular focus on the influence of vacuum level during treatment. Methodology Electrospun mats made of poly-L-lactide-co-glycolide (PLGA), poly-L-lactide-co-ε-caprolactone (PLC), and thermoplastic polyurethane (TPU) were prepared and sterilized with VH₂O₂ under low-vacuum (LV, 10 mbar) and high-vacuum (HV, 2 mbar) conditions at 20–50 °C. Scanning electron microscopy (SEM) assessed fiber morphology and pore area, contact angle measurements evaluated surface wettability, Fourier-transform infrared spectroscopy (FTIR) detected chemical modifications, and gel permeation chromatography (GPC) analyzed weight-average molecular weight (Mw) and polydispersity index (PI).

Results

HV sterilization caused fiber compression across all polymers, increasing fiber diameter and reducing pore area. LV sterilization conditions prevented morphological changes in TPU and PLC, while PLGA remained sensitive under both vacuum levels. PLGA also exhibited increased hydrophilicity due to surface reorganization and transition from Cassie–Baxter to Wenzel-type wetting. FTIR showed no chemical changes, and GPC confirmed stable Mw and PI for all tested materials (PLGA: 25–27 kDa; PLC: 16–18 kDa; TPU: 23–25 kDa).

Conclusion

Deep vacuum VH₂O₂ sterilization at room temperature is feasible, with vacuum level critical for maintaining scaffold integrity. LV sterilization effectively preserves morphology in PLC and TPU, while HV is also reliable for less sensitive polymers. Chemical and molecular properties remain unaffected, supporting VH₂O₂ sterilization as a gentle and effective promising method for implantable textile-based biomaterials.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SterilizationHydrogen PeroxideBiocompatible MaterialsProstheses and ImplantsPolymersVacuumPolyestersPolylactic Acid-Polyglycolic Acid CopolymerPolyurethanesVolatilizationPolyglycolic AcidLactic AcidWettability

Stichwörter der Autor:innen

electrospun scaffoldslow-temperature sterilizationvaporized hydrogen peroxide sterilization

Substanzen

Hydrogen PeroxidePolyestersPolylactic Acid-Polyglycolic Acid CopolymerPolyurethanesBiocompatible MaterialsPolyglycolic AcidLactic AcidPolymers

Autor:innen und Institutionen

  • Department of Drug Sciences, University of Pavia, 27100, Pavia, Italy.
  • Department of Physical Chemistry, University of Pavia, 27100, Pavia, Italy.
  • De Lama S.p.A., 27028, S. Martino Siccomario (Pavia), Italy.
  • Department of Drug Sciences, University of Pavia, 27100, Pavia, Italy. ida.genta@unipv.it.

Förderung

  • HORIZON EUROPE Marie Sklodowska-Curie Actions
  • Università degli Studi di Pavia

Zitieren

APA

Nallathambi, A., Chiesa, E., Rosalia, M., Bruni, G., Tamborini, A., Crotti, S., & Genta, I. (2026). Advancing Sterilization of Medical Implant Polymers: Novel Low-Temperature Deep-vacuum Vaporized H2O2 Technology Surpasses Current Methods. Pharmaceutical Research. https://doi.org/10.1007/s11095-026-04063-x

Vancouver

Nallathambi A, Chiesa E, Rosalia M, Bruni G, Tamborini A, Crotti S, et al. Advancing Sterilization of Medical Implant Polymers: Novel Low-Temperature Deep-vacuum Vaporized H2O2 Technology Surpasses Current Methods. Pharmaceutical Research. 2026. doi:10.1007/s11095-026-04063-x

BibTeX
@article{Nallathambi2026,
  author = {Nallathambi, Abinaya and Chiesa, Enrica and Rosalia, Mariella and Bruni, Giovanna and Tamborini, Aurora and Crotti, Sergio and Genta, Ida},
  title = {Advancing Sterilization of Medical Implant Polymers: Novel Low-Temperature Deep-vacuum Vaporized H2O2 Technology Surpasses Current Methods},
  journal = {Pharmaceutical Research},
  year = {2026},
  doi = {10.1007/s11095-026-04063-x}
}
Kostenlos starten →