EUVIMEDEUVIMED entdecken
Crossref

PVADet: fast patient-ventilator asynchrony detection on waveforms

Longxiang Su, Yan Li, Yunping Lan, Qiang Sun, Fuhong Cai, Hongli He, Siyi Yuan, Song Zhang, Xianlong Liu, Elias Baedorf-Kassis, Xiaobo Huang, Yun Long

Journal of Clinical Monitoring and Computing · 2026

DOI: 10.1007/s10877-025-01405-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Monitoring and Computing, 2026. Verfasst von Longxiang Su und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Critical Care Medicine, State Key Laboratory of Complex Severe and Rare Diseases, Peking Union Medical College Hospital, Chinese Academ…. Hauptthemen laut MEDLINE: Respiration, Artificial, Ventilators, Mechanical. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 57 Literaturangaben. Gefördert von: National High-Level Hospital Clinical Research Funding, CAMS Innovation Fund for Medical Sciences, 5G Network Construction of Remote Intensive care and Application of Critical Infection Standard Diagnosis and Treatment Cloud Platform.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Patient-ventilator asynchrony (PVA) is a common and critically import clinical problem in patients receiving mechanical ventilation. However, PVAs are often underrecognized, underestimated and delayed, and there has been minimal success in automating their detection. In this study, we develop an efficient and fast end-to-end model to recognize PVAs on ventilator waveforms: running the model costs 106.5ms on CPUs and 7.8ms on GPUs. We propose label striping and stripe-mask encoding for efficient multi-class multi-target detecting. The model innovatively integrates causal convolutional, depth-wise separable convolutional, and recurrent neural networks to memorize long short-term causal features. With 60s waveform segments, our model performs a cross-validation mean average precision (mAP) of 88.1% and a testing mAP of 65.7% for comprehensive PVA detection. Our approach might be implemented as a monitoring tool to automatically identify PVAs for improving bedside and remote care and prioritizing patient comfort.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Respiration, ArtificialVentilators, MechanicalHumansNeural Networks, ComputerAlgorithmsMonitoring, PhysiologicSignal Processing, Computer-AssistedReproducibility of ResultsPatient-Ventilator Asynchrony

Stichwörter der Autor:innen

Critical CareDeep LearningMechenical VentilationPatient-ventilator asynchronyTime-series AnalysisWaveform Monitoring

Autor:innen und Institutionen

  • Department of Critical Care Medicine, State Key Laboratory of Complex Severe and Rare Diseases, Peking Union Medical College Hospital, Chinese Academy of Medical Science and Peking Union Medical College, Beijing, 100730…
  • Department of Pulmonary, Critical Care and Sleep Medicine, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA, 02215, United States of America.
  • Shanghai SVM (Shumu) Medical Technology Co., Ltd, Shanghai, 200230, China.
  • Intensive Care Unit, Sichuan Provincial People's Hospital, School of Medicine of University of Electronic Science and Technology, Sichuan Academy of Medical Sciences, Chengdu, 610000, China.
  • Intensive Care Unit, Affiliated Hospital of Jining Medical University, Jining, 272029, China.
  • Department of Automation, Tsinghua University, Beijing, 100084, China.

Förderung

  • National High-Level Hospital Clinical Research Funding
  • CAMS Innovation Fund for Medical Sciences
  • 5G Network Construction of Remote Intensive care and Application of Critical Infection Standard Diagnosis and Treatment Cloud Platform

Zitieren

APA

Su, L., Li, Y., Lan, Y., Sun, Q., Cai, F., He, H., Yuan, S., Zhang, S., Liu, X., Baedorf-Kassis, E., Huang, X., & Long, Y. (2026). PVADet: fast patient-ventilator asynchrony detection on waveforms. Journal of Clinical Monitoring and Computing. https://doi.org/10.1007/s10877-025-01405-9

Vancouver

Su L, Li Y, Lan Y, Sun Q, Cai F, He H, et al. PVADet: fast patient-ventilator asynchrony detection on waveforms. Journal of Clinical Monitoring and Computing. 2026. doi:10.1007/s10877-025-01405-9

BibTeX
@article{Su2026,
  author = {Su, Longxiang and Li, Yan and Lan, Yunping and Sun, Qiang and Cai, Fuhong and He, Hongli and Yuan, Siyi and Zhang, Song and Liu, Xianlong and Baedorf-Kassis, Elias and Huang, Xiaobo and Long, Yun},
  title = {PVADet: fast patient-ventilator asynchrony detection on waveforms},
  journal = {Journal of Clinical Monitoring and Computing},
  year = {2026},
  doi = {10.1007/s10877-025-01405-9}
}
Kostenlos starten →