EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Speckle Vibrometry for Contactless Instantaneous Heart Rate and Respiration Rate Monitoring on Mechanically Ventilated Patients

Shuhao Que iD, Iris Cramer iD, Lukas Dekker, Sebastiaan Overeem iD, Arthur Bouwman iD, Svitlana Zinger, Sander Stuijk iD, Fokke van Meulen iD

Sensors · 2024

DOI: 10.3390/s24196374 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Shuhao Que und 7 weiteren Autor:innen. Beteiligte Einrichtung: Electrical Engineering Department, Eindhoven University of Technology, 5612 AZ Eindhoven, The Netherlands. Hauptthemen laut MEDLINE: Heart Rate, Respiratory Rate, Respiration, Artificial, Algorithms. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: the NWO/HTSM program in the context of the project: Unobtrusive Monitoring of Sleep Apnea, UMOSA, NWO/HTSM program in the context of the project: Unobtrusive Monitoring of Sleep Apnea, UMOSA.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

Contactless monitoring of instantaneous heart rate and respiration rate has a significant clinical relevance. This work aims to use Speckle Vibrometry (i.e., based on the secondary laser speckle effect) to contactlessly measure these two vital signs in an intensive care unit.

Methods

In this work, we propose an algorithm for the estimation of instantaneous heart rate and respiration rate from mechanically ventilated patients. The algorithm uses multiple regions, principal component analysis, and dominant angle analysis. A semi-automated peak detection method is implemented to precisely label the aortic valve opening peak within the cardiac waveform.

Results

Compared with electrocardiography, the present work achieves limits of agreement of [−2.19, 1.73] beats per minute of instantaneous heart rate. The measurement spot is on the chest covered with two to three layers of duvet blankets. Compared with the airway flow signal measured by the mechanical ventilator, the present work achieves limits of agreement of [−0.68, 0.46] respirations per minute of instantaneous respiration rate.

Conclusions

These results showcased Speckle Vibrometry’s potential in vital sign monitoring in a clinical setting. Significance: This is the first human clinical study for Speckle Vibrometry.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart RateRespiratory RateRespiration, ArtificialAlgorithmsHumansMonitoring, PhysiologicElectrocardiographyMaleVibrationFemaleSignal Processing, Computer-AssistedMiddle Aged

Stichwörter der Autor:innen

ICUcameracontactlessinstantaneous heart rateinstantaneous respiration ratelaser specklespeckle vibrometryvital sign monitoring

Autor:innen und Institutionen

  • Electrical Engineering Department, Eindhoven University of Technology, 5612 AZ Eindhoven, The Netherlands.
  • Catharina Ziekenhuis, 5623 EJ Eindhoven, The Netherlands.
  • Kempenhaeghe, 5591 VE Heeze, The Netherlands.

ORCID-Profile: Shuhao Que, Iris Cramer, Sebastiaan Overeem, Arthur Bouwman, Sander Stuijk, Fokke van Meulen

Förderung

  • the NWO/HTSM program in the context of the project: Unobtrusive Monitoring of Sleep Apnea, UMOSA
  • NWO/HTSM program in the context of the project: Unobtrusive Monitoring of Sleep Apnea, UMOSA

Zitieren

APA

Que, S., Cramer, I., Dekker, L., Overeem, S., Bouwman, A., Zinger, S., Stuijk, S., & van Meulen, F. (2024). Speckle Vibrometry for Contactless Instantaneous Heart Rate and Respiration Rate Monitoring on Mechanically Ventilated Patients. Sensors. https://doi.org/10.3390/s24196374

Vancouver

Que S, Cramer I, Dekker L, Overeem S, Bouwman A, Zinger S, et al. Speckle Vibrometry for Contactless Instantaneous Heart Rate and Respiration Rate Monitoring on Mechanically Ventilated Patients. Sensors. 2024. doi:10.3390/s24196374

BibTeX
@article{Que2024,
  author = {Que, Shuhao and Cramer, Iris and Dekker, Lukas and Overeem, Sebastiaan and Bouwman, Arthur and Zinger, Svitlana and Stuijk, Sander and van Meulen, Fokke},
  title = {Speckle Vibrometry for Contactless Instantaneous Heart Rate and Respiration Rate Monitoring on Mechanically Ventilated Patients},
  journal = {Sensors},
  year = {2024},
  doi = {10.3390/s24196374}
}
Kostenlos starten →