Impact of pupil dilation on IOL power calculation based on biometric measurements with Heidelberg Anterion
Documenta Ophthalmologica · 12.06.2026
DOI: 10.1007/s10633-026-10117-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift Documenta Ophthalmologica, online erschienen am 12.06.2026. Verfasst von Sara Marquina-Martín und 5 weiteren Autor:innen. Beteiligte Einrichtung: GIMSO, Institute for Health Research Aragón, University Hospital Miguel Servet, Po de Isabel La Católica, 1-3, 50009, Zaragoza, Spain. saramarqu@hotm…. Die Arbeit nennt 24 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: To evaluate the impact of pharmacologic pupil dilation (PD) on biometric measurements and intraocular lens (IOL) power calculation using different generation formulas in patients undergoing cataract surgery. METHODS: This prospective study included 126 eyes from 75 patients. Biometric data were obtained before and after pharmacologic dilation using the Anterion Swept-Source OCT biometer (Heidelberg Engineering). Parameters recorded included axial length (AL), mean keratometry (Km), anterior chamber depth (ACD), lens thickness (LT), and white-to-white (WTW) distance. IOL power was calculated using six formulas: SRK/T, Haigis, Barrett Universal II (BUII), Kane, Hoffer QST, and Okulix. Statistical comparisons of biometric values and predicted IOL power before and after dilation were conducted using mixed-effects models. RESULTS: No significant changes were observed in AL, CCT, LT or Km after dilation. However, ACD, WTW distance and PD showed statistically significant increases (both p < 0.01). IOL power estimations based on SRK/T, Haigis, Hoffer QST, BUII and Kane were not significantly affected after dilation. In contrast, Okulix formulas demonstrated statistically significant variations in calculated IOL power post-dilation (p < 0.0001). CONCLUSIONS: Pharmacologic pupil dilation induces changes in ACD and WTW measurements obtained using the Anterion Swept-Source biometer, thereby influencing IOL power estimation in formulas incorporating these parameters. Ray-tracing Okulix formula is more sensitive to dilation, whereas third-generation formulas remain more stable. Consistency in pupil status during biometry acquisition is essential to ensure refractive predictability.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- GIMSO, Institute for Health Research Aragón, University Hospital Miguel Servet, Po de Isabel La Católica, 1-3, 50009, Zaragoza, Spain. saramarqu@hotmail.com.
- UFR, Department of Ophthalmology, University Hospital Miguel Servet, Zaragoza, Spain. saramarqu@hotmail.com.
- UFR, Department of Ophthalmology, University Hospital Miguel Servet, Zaragoza, Spain.
- Surgery Department, Ophthalmology Area, University of Zaragoza, Zaragoza, Spain.
- GIMSO, Institute for Health Research Aragón, University Hospital Miguel Servet, Po de Isabel La Católica, 1-3, 50009, Zaragoza, Spain.
- Applied Physics Department, Optic Area, University of Zaragoza, Zaragoza, Spain.
Zitieren
APA
Marquina-Martín, S., Segura-Calvo, F. J., Puzo-Bayod, M., Jiménez-García, M., Castro-Alonso, F. J., & UFR-HUMS Group Zaragoza (2026). Impact of pupil dilation on IOL power calculation based on biometric measurements with Heidelberg Anterion. Documenta Ophthalmologica. https://doi.org/10.1007/s10633-026-10117-7
Vancouver
Marquina-Martín S, Segura-Calvo FJ, Puzo-Bayod M, Jiménez-García M, Castro-Alonso FJ, UFR-HUMS Group Zaragoza. Impact of pupil dilation on IOL power calculation based on biometric measurements with Heidelberg Anterion. Documenta Ophthalmologica. 2026. doi:10.1007/s10633-026-10117-7
BibTeX
@article{MarquinaMartn2026,
author = {Marquina-Martín, Sara and Segura-Calvo, Francisco J and Puzo-Bayod, Martín and Jiménez-García, Marta and Castro-Alonso, Francisco J and UFR-HUMS Group Zaragoza, UFR-HUMS Group Zaragoza},
title = {Impact of pupil dilation on IOL power calculation based on biometric measurements with Heidelberg Anterion},
journal = {Documenta Ophthalmologica},
year = {2026},
doi = {10.1007/s10633-026-10117-7}
}
EUVIMED