Examination of rapid adjustment system based on screen score obtained using continuous shear wave elastography
Journal of Medical Ultrasonics · 2024-07 · 51 (3) · 407-418
DOI: 10.1007/s10396-024-01439-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Medical Ultrasonics, online erschienen am 12.04.2024. Verfasst von Marie Tabaru und 4 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Innovative Research, Tokyo Institute of Technology, 4259 R2-25, Nagatsuta-cho, Midori-ku, Yokohama-shi, Kanagawa, 226-8503, Japan. Hauptthemen laut MEDLINE: Elasticity Imaging Techniques, Phantoms, Imaging. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Japan Society for the Promotion of Science, the Cooperative Research Project of Research Center for Biomedical Engineering.
Abstract (Originalsprache)
Purpose
Continuous shear wave elastography (C-SWE) can be expected to be applied to portable muscle elasticity diagnosis. To establish diagnostic technology, it will be necessary to improve measurement techniques and quantitative measurement accuracy.
Methods
In this study, we investigated two screen scores: the quality index ( Q -index), which determines whether the intensity of a power Doppler image is appropriate, and the shear wave propagation direction index (SWDI), which determines the uniformity of shear wave propagation.
Results
First, we performed numerical simulations with white noise and found that the coefficient of variation of shear wave velocity estimation was less than 5% when the normalized Q -index was greater than 0.27. Furthermore, regarding the SWDI, we clarified the relationship between the standard deviation in shear wave propagation direction and the SWDI. Next, the relationship between the Q -index and coefficient of variation of estimated shear wave velocity was evaluated through experiments using a tissue-mimicking phantom. The results showed that there was a negative correlation between the Q -index and the coefficient of variation, and the fluctuation of the propagation velocity could be inferred from the Q -index. Finally, we showed the results of applying the screen scores to muscle relaxation monitoring and confirmed its usefulness in clinical applications.
Conclusion
By applying the screen scores, we showed improved stability in speed estimation in C-SWE, and demonstrated the possibility of clinical applicability.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Innovative Research, Tokyo Institute of Technology, 4259 R2-25, Nagatsuta-cho, Midori-ku, Yokohama-shi, Kanagawa, 226-8503, Japan.
- Graduate School of Science and Technology, Gunma University, 1-5-1 Tenjin-cho, Kiryu-shi, Gunma, 376-8515, Japan.
- Department of Orthopaedic Surgery, Graduate School of Medicine, Gunma University, Maebashi-shi, Gunma, 371-8511, Japan.
- Graduate School of Science and Technology, Gunma University, 1-5-1 Tenjin-cho, Kiryu-shi, Gunma, 376-8515, Japan. yamakoshi@gunma-u.ac.jp.
Förderung
- Japan Society for the Promotion of Science
- the Cooperative Research Project of Research Center for Biomedical Engineering
Zitieren
APA
Tabaru, M., Koda, R., Shitara, H., Chikuda, H., & Yamakoshi, Y. (2024). Examination of rapid adjustment system based on screen score obtained using continuous shear wave elastography. Journal of Medical Ultrasonics, 51(3), 407-418. https://doi.org/10.1007/s10396-024-01439-7
Vancouver
Tabaru M, Koda R, Shitara H, Chikuda H, Yamakoshi Y. Examination of rapid adjustment system based on screen score obtained using continuous shear wave elastography. Journal of Medical Ultrasonics. 2024;51(3):407-418. doi:10.1007/s10396-024-01439-7
BibTeX
@article{Tabaru2024,
author = {Tabaru, Marie and Koda, Ren and Shitara, Hitoshi and Chikuda, Hirotaka and Yamakoshi, Yoshiki},
title = {Examination of rapid adjustment system based on screen score obtained using continuous shear wave elastography},
journal = {Journal of Medical Ultrasonics},
year = {2024},
volume = {51},
number = {3},
pages = {407--418},
doi = {10.1007/s10396-024-01439-7}
}