Assessing post-TAVR cardiac conduction abnormalities risk using an electromechanically coupled beating heart
Biomechanics and Modeling in Mechanobiology · 2025
DOI: 10.1007/s10237-024-01893-9 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomechanics and Modeling in Mechanobiology, 2025. Verfasst von Symon Reza und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Stony Brook University, Stony Brook, NY, 11794-8084, USA. Hauptthemen laut MEDLINE: Transcatheter Aortic Valve Replacement, Heart Conduction System. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 78 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, National Institute of Biomedical Imaging and Bioengineering.
Abstract (Originalsprache)
Transcatheter aortic valve replacement (TAVR) has rapidly displaced surgical aortic valve replacement (SAVR). However, certain post-TAVR complications persist, with cardiac conduction abnormalities (CCA) being one of the major ones. The elevated pressure exerted by the TAVR stent onto the conduction fibers situated between the aortic annulus and the His bundle, in proximity to the atrioventricular (AV) node, may disrupt the cardiac conduction leading to the emergence of CCA. In this study, an in silico framework was developed to assess the CCA risk, incorporating the effect of a dynamic beating heart and preprocedural parameters such as implantation depth and preexisting cardiac asynchrony in the new onset of post-TAVR CCA. A self-expandable TAVR device deployment was simulated inside an electromechanically coupled beating heart model in five patient scenarios, including three implantation depths and two preexisting cardiac asynchronies: (i) a right bundle branch block (RBBB) and (ii) a left bundle branch block (LBBB). Subsequently, several biomechanical parameters were analyzed to assess the post-TAVR CCA risk. The results manifested a lower cumulative contact pressure on the conduction fibers following TAVR for aortic deployment (0.018 MPa) compared to nominal condition (0.29 MPa) and ventricular deployment (0.52 MPa). Notably, the preexisting RBBB demonstrated a higher cumulative contact pressure (0.34 MPa) compared to the nominal condition and preexisting LBBB (0.25 MPa). Deeper implantation and preexisting RBBB cause higher stresses and contact pressure on the conduction fibers leading to an increased risk of post-TAVR CCA. Conversely, implantation above the MS landmark and preexisting LBBB reduces the risk.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Stony Brook University, Stony Brook, NY, 11794-8084, USA.
- Department of Biomedical Engineering, Stony Brook University, Stony Brook, NY, 11794-8084, USA. danny.bluestein@stonybrook.edu.
Förderung
- NIBIB NIH HHS
- National Institute of Biomedical Imaging and Bioengineering
Zitieren
APA
Reza, S., Kovarovic, B., & Bluestein, D. (2025). Assessing post-TAVR cardiac conduction abnormalities risk using an electromechanically coupled beating heart. Biomechanics and Modeling in Mechanobiology. https://doi.org/10.1007/s10237-024-01893-9
Vancouver
Reza S, Kovarovic B, Bluestein D. Assessing post-TAVR cardiac conduction abnormalities risk using an electromechanically coupled beating heart. Biomechanics and Modeling in Mechanobiology. 2025. doi:10.1007/s10237-024-01893-9
BibTeX
@article{Reza2025,
author = {Reza, Symon and Kovarovic, Brandon and Bluestein, Danny},
title = {Assessing post-TAVR cardiac conduction abnormalities risk using an electromechanically coupled beating heart},
journal = {Biomechanics and Modeling in Mechanobiology},
year = {2025},
doi = {10.1007/s10237-024-01893-9}
}
EUVIMED