EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A fast in silico model for preoperative risk assessment of paravalvular leakage

Michelle Spanjaards, Finja Borowski, Laura Supp, René Ubachs, Valentina Lavezzo, Olaf van der Sluis

Biomechanics and Modeling in Mechanobiology · 2024

DOI: 10.1007/s10237-024-01816-8 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomechanics and Modeling in Mechanobiology, 2024. Verfasst von Michelle Spanjaards und 5 weiteren Autor:innen. Beteiligte Einrichtung: Philips Innovation and Strategy, High Tech Campus 34, Eindhoven, The Netherlands. Hauptthemen laut MEDLINE: Computer Simulation. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Commission.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In silico simulations can be used to evaluate and optimize the safety, quality, efficacy and applicability of medical devices. Furthermore, in silico modeling is a powerful tool in therapy planning to optimally tailor treatment for each patient. For this purpose, a workflow to perform fast preoperative risk assessment of paravalvular leakage (PVL) after transcatheter aortic valve replacement (TAVR) is presented in this paper. To this end, a novel, efficient method is introduced to calculate the regurgitant volume in a simplified, but sufficiently accurate manner. A proof of concept of the method is obtained by comparison of the calculated results with results obtained from in vitro experiments. Furthermore, computational fluid dynamics (CFD) simulations are used to validate more complex stenosis scenarios. Comparing the simplified leakage model to CFD simulations reveals its potential for procedure planning and qualitative preoperative risk assessment of PVL. Finally, a 3D device deployment model and the efficient leakage model are combined to showcase the application of the presented leakage model, by studying the effect of stent size and the degree of stenosis on the regurgitant volume. The presented leakage model is also used to visualize the leakage path. To generalize the leakage model to a wide range of clinical applications, further validation on a large cohort of patients is needed to validate the accuracy of the model’s prediction under various patient-specific conditions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Computer SimulationHumansRisk AssessmentTranscatheter Aortic Valve ReplacementHydrodynamicsAortic ValveModels, CardiovascularStents

Stichwörter der Autor:innen

Aortic stenosisComputational modelingIn silico modelingPreoperative risk assessmentProcedure planningTAVRThin-film approximationparavalvular leakage

Autor:innen und Institutionen

  • Philips Innovation and Strategy, High Tech Campus 34, Eindhoven, The Netherlands.
  • Institute for Implant Technology and Biomaterials e.V., Friedrich-Barnewitz-Str. 4, Rostock-Warnemünde, Germany.
  • Philips Innovation and Strategy, High Tech Campus 34, Eindhoven, The Netherlands. o.v.d.sluis@tue.nl.
  • Eindhoven University of Technology, Groene Loper 15, Eindhoven, The Netherlands. o.v.d.sluis@tue.nl.

Förderung

  • European Commission

Zitieren

APA

Spanjaards, M., Borowski, F., Supp, L., Ubachs, R., Lavezzo, V., & van der Sluis, O. (2024). A fast in silico model for preoperative risk assessment of paravalvular leakage. Biomechanics and Modeling in Mechanobiology. https://doi.org/10.1007/s10237-024-01816-8

Vancouver

Spanjaards M, Borowski F, Supp L, Ubachs R, Lavezzo V, van der Sluis O. A fast in silico model for preoperative risk assessment of paravalvular leakage. Biomechanics and Modeling in Mechanobiology. 2024. doi:10.1007/s10237-024-01816-8

BibTeX
@article{Spanjaards2024,
  author = {Spanjaards, Michelle and Borowski, Finja and Supp, Laura and Ubachs, René and Lavezzo, Valentina and van der Sluis, Olaf},
  title = {A fast in silico model for preoperative risk assessment of paravalvular leakage},
  journal = {Biomechanics and Modeling in Mechanobiology},
  year = {2024},
  doi = {10.1007/s10237-024-01816-8}
}
Kostenlos starten →