An in-house fabricated patient-specific pediatric head CT phantom for trauma-oriented emergency imaging: two-scanner validation and expert reader assessment

Hamza Sekkat, Oussama El Mouden, Youssef Madkouri, Yasmina Berrada, Abdellah Khallouqi, Abdellah Halimi, Omar El Rhazouani

Emergency Radiology · 30.05.2026

DOI: 10.1007/s10140-026-02497-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Emergency Radiology, online erschienen am 30.05.2026. Verfasst von Hamza Sekkat und 6 weiteren Autor:innen. Beteiligte Einrichtung: Sciences and Engineering of Biomedicals, Biophysics and Health Laboratory, Higher Institute of Health Sciences, Hassan First University, Settat, 2600…. Die Arbeit nennt 20 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To evaluate an in-house newly fabricated patient-specific pediatric head CT phantom developed for trauma-oriented emergency imaging, with emphasis on multiplanar visual realism, cross-platform interpretability, and expert-reader assessment of diagnostic acceptability. METHODS: A heterogeneous phantom was generated from an anonymized retrospective head CT dataset of a 5-year-old child and designed to preserve emergency-relevant osseous, aerated, orbital, and intracranial anatomy. A complex orbito-nasal midface trauma was incorporated as a trauma-mimicking feature. CT was performed on two clinical multidetector systems from different vendors: a 64-row GE Discovery CT750 HD and an 80-row Canon Aquilion RXL, each using routine pediatric head CT protocols. Ten radiologists independently reviewed axial, coronal, and sagittal reformations in bone and soft-tissue windows using a structured 20-item, 4-domain scoring framework. RESULTS: Representative multiplanar images showed preserved calvarial contour, skull-base and orbital anatomy, aerated sinonasal spaces, intracranial contour, falcine/midline anatomy, ventricular depiction, posterior fossa appearance, and visualization of the complex orbito-nasal midface trauma. Expert ratings were favorable on both scanners, with higher scores on Canon across all domains. Domain means on Canon versus GE were 4.38 versus 4.02 for osseous realism, 4.32 versus 4.00 for soft-tissue realism, 4.36 versus 4.18 for multiplanar interpretability, and 4.57 versus 4.10 for clinical utility. Overall composite scores were 4.38 on Canon and 4.07 on GE, paired domain-level differences showed higher mean scores for the Canon dataset in all domains, with the largest difference in clinical utility (+ 0.47). CONCLUSION: The phantom provided clinically recognizable, trauma-relevant pediatric emergency head CT appearances across multiplanar bone and soft-tissue review and was rated as realistic and diagnostically acceptable by expert readers, supporting its use in protocol evaluation, training, and observer-based emergency radiology research.

Schlagwörter

Stichwörter der Autor:innen

Emergency radiologyExpert reader assessmentMultiplanar CTPediatric head CT phantomTrauma-oriented imaging

Autor:innen und Institutionen

  • Sciences and Engineering of Biomedicals, Biophysics and Health Laboratory, Higher Institute of Health Sciences, Hassan First University, Settat, 26000, Morocco. sekkat.isss@uhp.ac.ma.
  • Higher Institute of Nursing Professions and Health Techniques, Rabat, Morocco. sekkat.isss@uhp.ac.ma.
  • Sciences and Engineering of Biomedicals, Biophysics and Health Laboratory, Higher Institute of Health Sciences, Hassan First University, Settat, 26000, Morocco.
  • Department of Radiotherapy, International Clinic of Settat, Settat, Morocco.
  • Laboratory of Electronic Systems, Information Processing, Mechanics and Energetics, Faculty of Sciences, Ibn Tofail University, Kenitra, Morocco.
  • Department of Radiology, 3GCOM Company, Rabat, Morocco.

Zitieren

APA

Sekkat, H., Mouden, O. E., Madkouri, Y., Berrada, Y., Khallouqi, A., Halimi, A., & El Rhazouani, O. (2026). An in-house fabricated patient-specific pediatric head CT phantom for trauma-oriented emergency imaging: two-scanner validation and expert reader assessment. Emergency Radiology. https://doi.org/10.1007/s10140-026-02497-9

Vancouver

Sekkat H, Mouden OE, Madkouri Y, Berrada Y, Khallouqi A, Halimi A, et al. An in-house fabricated patient-specific pediatric head CT phantom for trauma-oriented emergency imaging: two-scanner validation and expert reader assessment. Emergency Radiology. 2026. doi:10.1007/s10140-026-02497-9

BibTeX
@article{Sekkat2026,
  author = {Sekkat, Hamza and Mouden, Oussama El and Madkouri, Youssef and Berrada, Yasmina and Khallouqi, Abdellah and Halimi, Abdellah and El Rhazouani, Omar},
  title = {An in-house fabricated patient-specific pediatric head CT phantom for trauma-oriented emergency imaging: two-scanner validation and expert reader assessment},
  journal = {Emergency Radiology},
  year = {2026},
  doi = {10.1007/s10140-026-02497-9}
}
Kostenlos starten →