In‐house development of CT‐based 3D‐printed head & neck phantom for Radiotherapy applications
Journal of Applied Clinical Medical Physics · 2026-07 · 27 (7)
DOI: 10.1002/acm2.70706 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 15.07.2026. Verfasst von Subhalaxmi Mishra und 6 weiteren Autor:innen. Beteiligte Einrichtung: Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, Mumbai, Maharashtra, India. Hauptthemen laut MEDLINE: Phantoms, Imaging, Printing, Three-Dimensional, Tomography, X-Ray Computed, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Anthropomorphic phantoms are widely used in Radiotherapy for quality assurance (QA), treatment verification and dosimetric validation purposes. However, commercially available anthropomorphic phantoms are expensive and offers limited customization, restricting their broader clinical use. The advancement three‐dimensional (3D) printing technology has provided a versatile and cost‐effective approach for fabrication of such patient‐specific anthropomorphic phantoms with accurate anatomical and radiological properties.
Purpose
This study aims to design, fabricate and evaluate a CT‐based Head & Neck anthropomorphic phantom using 3D printing technique for Radiotherapy applications.
Methods
The design of the phantom is developed from the computed tomography (CT) images of a male patient. Relevant organs such as brain, eyes, trachea, thyroid, skull, mandible, spine, soft tissue etc were segmented and reconstructed. Suitable tissue‐equivalent materials compatible with 3D printing were selected and optimized to replicate the radiological properties of the human tissues. CT scan images of the fabricated phantom were acquired using a CT simulator. Hounsfield Unit (HU) values of each organ was compared with the patient CT data to assess tissue equivalence and uniformity. For dosimetric evaluation of the phantom, the acquired CT images of the fabricated phantom were imported into the treatment planning system (TPS). A treatment plan was generated in TPS to deliver 2 Gy dose to planning target volume (PTV) and executed in a 6MV Linear accelerator. Dose verification was performed by placing thermoluminescent dosimeters (TLDs) at predefined locations within the PTV and organs at risk (OARs). Measured doses were compared with TPS‐calculated dose values to evaluate dosimetric accuracy.
Results
The developed phantom demonstrated good agreement between the patient CT data and the phantom CT images. The agreement of HU values across analyzed slices was within 5%, indicating accurate replication of tissue‐equivalent attenuation properties. Intra‐organ HU uniformity was also within 5%, confirming consistent material composition and printing quality. Dosimetric comparison revealed that deviations between TPS‐calculated and TLD‐measured doses were less than 5% within the PTV, 1%–5% for OARs, and 7%–9% at distal exit dose measurement points. The PTV received approximately 190cGy, showing good agreement between planned and delivered doses. Larger percentage deviations observed at distal exit points, where doses were below 10 cGy, were attributed primarily to detector uncertainties and increased measurement uncertainty at low dose levels rather than inaccuracies in phantom design or treatment delivery.
Conclusion
The developed 3D‐printed CT‐based indigenous Head & Neck phantom accurately replicates both the anatomical and tissue equivalence characteristics of the patient CT data. The phantom demonstrated reliable dosimetric parameters supporting its suitability for Radiotherapy applications. The phantom is customizable, simple to use, cost effective and could be a promising alternative to commercial anthropomorphic phantoms.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, Mumbai, Maharashtra, India.
- Homi Bhabha National Institute, Anushaktinagar, Mumbai, Maharashtra, India.
- Department of Radiation Oncology, P. D. Hinduja National Hospital and MRC, Mumbai, Maharashtra, India.
Zitieren
APA
Mishra, S., Shrivastava, V., Kumar, R., Deshpande, S., Selvam, T. P., Sharma, S. D., & Sapra, B. K. (2026). In‐house development of CT‐based 3D‐printed head & neck phantom for Radiotherapy applications. Journal of Applied Clinical Medical Physics, 27(7). https://doi.org/10.1002/acm2.70706
Vancouver
Mishra S, Shrivastava V, Kumar R, Deshpande S, Selvam TP, Sharma SD, et al. In‐house development of CT‐based 3D‐printed head & neck phantom for Radiotherapy applications. Journal of Applied Clinical Medical Physics. 2026;27(7). doi:10.1002/acm2.70706
BibTeX
@article{Mishra2026,
author = {Mishra, Subhalaxmi and Shrivastava, Vandana and Kumar, Rajesh and Deshpande, Sudesh and Selvam, T Palani and Sharma, S D and Sapra, B K},
title = {In‐house development of CT‐based 3D‐printed head & neck phantom for Radiotherapy applications},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {7},
doi = {10.1002/acm2.70706}
}
EUVIMED