Simultaneous sealing and bisection of porcine renal blood vessels, ex vivo, using a continuous-wave, infrared diode laser at 1470 nm
Lasers in Medical Science · 2024
DOI: 10.1007/s10103-024-04093-0 ↗Kurz gefasst
Fachartikel in der Zeitschrift Lasers in Medical Science, 2024. Verfasst von Woheeb M Saeed und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Optical Science, University of North Carolina at Charlotte, 9201 University City Boulevard, Charlotte, NC, 28223, United St…. Hauptthemen laut MEDLINE: Lasers, Semiconductor, Renal Artery. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of North Carolina at Charlotte.
Abstract (Originalsprache)
Electrosurgical and ultrasonic devices are used in surgical procedures for hemostatic sealing and bisection of vascular tissues. Previous benchtop studies alternatively demonstrated successful infrared laser sealing and cutting of blood vessels, in a sequential, two-step approach. This study describes a smaller, laparoscopic device compatible design, and simultaneous approach to sealing and bisection of vessels, with potential optical feedback. A 1470-nm infrared diode laser sealed and bisected 40 porcine renal arteries, ex vivo. A reciprocating, side-firing, optical fiber, housed in a transparent square quartz optical chamber (2.7 × 2.7 × 25 mm outer dimensions), delivered laser energy over an 11 mm scan length, with a range of incident powers (41–59 W) and treatment times (5–21 s). Vessel diameters ranged from 2.5 to 4.8 mm. Vessel burst pressure measurements were performed on each cut end ( n = 80) with success indicated by pressures exceeding 360 mmHg. All vessel ends were successfully sealed and bisected (80/80). The highest incident power, 59 W, yielded short treatment times of 5–6 s. Peak temperatures on the external chamber surface reached 103 o C. Time to cool down to body temperature measured 37 s. Infrared lasers simultaneously seal and bisect blood vessels, with treatment times comparable to, and temperatures and cooling times lower than reported for conventional devices. Future work will focus on integrating the fiber and chamber into a standard 5-mm-outer-diameter laparoscopic device. Customization of fiber scan length to match vessel size may also reduce laser energy deposition, enabling lower peak temperatures, treatment times, and cooling times.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Physics and Optical Science, University of North Carolina at Charlotte, 9201 University City Boulevard, Charlotte, NC, 28223, United States.
- Department of Mechanical Engineering, University of North Carolina at Charlotte, Charlotte, United States.
- Department of Physics and Optical Science, University of North Carolina at Charlotte, 9201 University City Boulevard, Charlotte, NC, 28223, United States. nmfried@charlotte.edu.
Förderung
- University of North Carolina at Charlotte
Zitieren
APA
Saeed, W. M., Yoshino, J. K., Traynham, A. J., & Fried, N. M. (2024). Simultaneous sealing and bisection of porcine renal blood vessels, ex vivo, using a continuous-wave, infrared diode laser at 1470 nm. Lasers in Medical Science. https://doi.org/10.1007/s10103-024-04093-0
Vancouver
Saeed WM, Yoshino JK, Traynham AJ, Fried NM. Simultaneous sealing and bisection of porcine renal blood vessels, ex vivo, using a continuous-wave, infrared diode laser at 1470 nm. Lasers in Medical Science. 2024. doi:10.1007/s10103-024-04093-0
BibTeX
@article{Saeed2024,
author = {Saeed, Woheeb M and Yoshino, Jude K and Traynham, Alexandria J and Fried, Nathaniel M},
title = {Simultaneous sealing and bisection of porcine renal blood vessels, ex vivo, using a continuous-wave, infrared diode laser at 1470 nm},
journal = {Lasers in Medical Science},
year = {2024},
doi = {10.1007/s10103-024-04093-0}
}
EUVIMED