Fachartikel in der Zeitschrift Journal of Artificial Organs, 2026. Verfasst von Mizuki Ichikawa und 4 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Biology-Oriented Science and Technology, Kindai University, 930 Nishimitani, Kinokawa, Wakayama, 649-6493, Japan. Hauptthemen laut MEDLINE: Membranes, Artificial, Hemodiafiltration, Polymers. Die Arbeit nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
The novel lab-scale hollow fiber membrane module was fabricated to determine the apparent sieving coefficient ( $${SC}_{obs}$$ ) of commercial hemodiafiltration membranes, MFX-SW eco (polyethersulfone, Sample A) and NVF-P (polysulfone, Sample B), using the aqueous solution of dextran with a broader molecular weight distribution. All the molecular weight cut-off curves plotted as $${SC}_{obs}$$ versus filtration flow rate ( Q F ) exhibited smooth profiles without intersecting each other, showing distinct variations depending on Q F across a wide range of molecular weights. Field emission scanning electron microscopy (FE-SEM) observations revealed that the equivalent pore diameters of samples A and B were nearly identical at 23.0 ± 15.5 nm and 23.3 ± 15.5 nm, respectively. Although the surface porosity of sample A (14.5 ± 3.9%) was higher than that of sample B (10.2 ± 3.2%), the difference was not statistically significant. These numerical data were in excellent agreement with the qualitative morphological findings obtained from the FE-SEM images. Furthermore, it is suggested that the subtle structural differences between the two membranes are also reflected in the relationship between the real sieving coefficients ( $${SC}_{real}$$ ) calculated by using $${SC}_{obs}$$ and the equivalent pore diameters. Since the lab-scale modules are fabricated relatively easily by hand, this methodology is extremely valuable to rigorously define membrane performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Graduate School of Biology-Oriented Science and Technology, Kindai University, 930 Nishimitani, Kinokawa, Wakayama, 649-6493, Japan.
- Graduate School of Biology-Oriented Science and Technology, Kindai University, 930 Nishimitani, Kinokawa, Wakayama, 649-6493, Japan. fukuda@waka.kindai.ac.jp.
- Department of Biomedical Engineering, Kindai University, 930 Nishimitani, Kinokawa, Wakayama, 649-6493, Japan. fukuda@waka.kindai.ac.jp.
- Clinical Engineering, Suita tokushukai Hospital, 21-1 Senrioka-Nishi, Suita, Osaka, 565-0814, Japan.
- Professor Emeritus of Chemical Engineering, Waseda University, 3-4-1 Okubo, Shinjuku-ku, Tokyo, 169-8555, Japan.
Zitieren
APA
Ichikawa, M., Fukuda, M., Morikawa, T., Takeuchi, K., & Sakai, K. (2026). Correlation between real sieving coefficient determined using novel lab-scale module and pore structure of hemodiafiltration membranes. Journal of Artificial Organs. https://doi.org/10.1007/s10047-026-01577-4
Vancouver
Ichikawa M, Fukuda M, Morikawa T, Takeuchi K, Sakai K. Correlation between real sieving coefficient determined using novel lab-scale module and pore structure of hemodiafiltration membranes. Journal of Artificial Organs. 2026. doi:10.1007/s10047-026-01577-4
BibTeX
@article{Ichikawa2026,
author = {Ichikawa, Mizuki and Fukuda, Makoto and Morikawa, Takuma and Takeuchi, Kousei and Sakai, Kiyotaka},
title = {Correlation between real sieving coefficient determined using novel lab-scale module and pore structure of hemodiafiltration membranes},
journal = {Journal of Artificial Organs},
year = {2026},
doi = {10.1007/s10047-026-01577-4}
}
EUVIMED