EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Impact of different post-polymerization protocols on fracture load and two-body wear of 3D-printed fixed dental prostheses

Sophie Gekle, Bogna Stawarczyk, Felicitas Mayinger, Ursula Kreitmair, Moritz Hoffmann, John Meinen, Marcel Reymus

Clinical Oral Investigations · 2026

DOI: 10.1007/s00784-026-07232-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Oral Investigations, 2026. Verfasst von Sophie Gekle und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthetic Dentistry, University Hospital, LMU Munich, Goethestrasse 70, Munich, 80336, Germany. sophie.gekle@med.uni-muenchen.de. Hauptthemen laut MEDLINE: Printing, Three-Dimensional, Denture, Partial, Fixed, Dental Restoration Wear. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To evaluate five post-polymerization protocols (385, 400 and 460 nm LEDs; one- and two-capacitor flashlights) and aging on fracture load (FL) and two-body wear (TBW) of three 3D-printed resins for fixed dental prostheses (FDPs). Materials and methods 360 single-unit FDPs for mandibular molars were DLP-printed from two permanent resins (Varseo Smile CrownPlus [VAR], Crowntec [CRO]) and one temporary resin (Freeprint temp [FRE]). Half was artificially aged (480,000 mastication cycles, thermocycling 5/55°C). FL was measured in a universal testing machine, TBW by laser scanning. Data were analysed with three-way ANOVA, Kruskal–Wallis and post-hoc tests (α = 0.05).

Results

The protocol ( p < 0.001) and the material ( p < 0.001) significantly affected FL, whereas aging showed no significant main effect ( p = 0.100); the three-way interaction was significant ( p = 0.009). Initially, 460 nm yielded the highest FL for VAR (1505 ± 251 N) and FRE (1910 ± 237 N), the two-capacitor flashlight for CRO (1783 ± 333 N). Aging increased FL significantly in two and decreased it in three of 15 combinations. TBW (− 0.16 to − 0.35 mm) did not differ between protocols or materials.

Conclusions

The protocol effect on FL depended on the material and the aging level. Aging had no overall effect but increased FL in two combinations, possibly through continued polymerization. Protocol selection should consider the material. Clinical relevance Mean fracture loads exceeded reported maximum molar bite forces, although individual specimens failed below that range. Two-capacitor flashlight post-curing gave the most consistent performance for the materials tested.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Printing, Three-DimensionalDenture, Partial, FixedDental Restoration WearMaterials TestingDental Stress AnalysisHumansPolymerizationComposite ResinsDental MaterialsDental Restoration FailureSurface Properties

Stichwörter der Autor:innen

3D printingDental prosthesesDigital light processingFracture loadPost-polymerizationTwo-body wear

Substanzen

Composite ResinsDental Materials

Autor:innen und Institutionen

  • Department of Prosthetic Dentistry, University Hospital, LMU Munich, Goethestrasse 70, Munich, 80336, Germany. sophie.gekle@med.uni-muenchen.de.
  • Department of Prosthetic Dentistry, University Hospital, LMU Munich, Goethestrasse 70, Munich, 80336, Germany.
  • Biomaterials and Technology, Department Research, University Center for Dental Medicine Basel UZB, University of Basel, Basel, Switzerland.
  • Department of Reconstructive Dentistry, University Center for Dental Medicine Basel UZB, University of Basel, Basel, Switzerland.
  • NK Optik GmbH, Baierbrunn, Germany.
  • Department of Conservative Dentistry, Periodontology and Digital Dentistry, University Hospital, LMU Munich, Munich, Germany.

Zitieren

APA

Gekle, S., Stawarczyk, B., Mayinger, F., Kreitmair, U., Hoffmann, M., Meinen, J., & Reymus, M. (2026). Impact of different post-polymerization protocols on fracture load and two-body wear of 3D-printed fixed dental prostheses. Clinical Oral Investigations. https://doi.org/10.1007/s00784-026-07232-9

Vancouver

Gekle S, Stawarczyk B, Mayinger F, Kreitmair U, Hoffmann M, Meinen J, et al. Impact of different post-polymerization protocols on fracture load and two-body wear of 3D-printed fixed dental prostheses. Clinical Oral Investigations. 2026. doi:10.1007/s00784-026-07232-9

BibTeX
@article{Gekle2026,
  author = {Gekle, Sophie and Stawarczyk, Bogna and Mayinger, Felicitas and Kreitmair, Ursula and Hoffmann, Moritz and Meinen, John and Reymus, Marcel},
  title = {Impact of different post-polymerization protocols on fracture load and two-body wear of 3D-printed fixed dental prostheses},
  journal = {Clinical Oral Investigations},
  year = {2026},
  doi = {10.1007/s00784-026-07232-9}
}
Kostenlos starten →