Effect of offset on the precision of 3D-printed orthognathic surgical splints
Clinical Oral Investigations · 2023
DOI: 10.1007/s00784-023-05134-8 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Oral Investigations, 2023. Verfasst von Peiqi Wang und 7 weiteren Autor:innen. Beteiligte Einrichtung: State Key Laboratory of Oral Diseases & National Clinical Research Center for Oral Diseases & Department of Orthodontics, West China Hospital of Stom…. Hauptthemen laut MEDLINE: Orthognathic Surgical Procedures, Exostoses. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 42 Literaturangaben. Gefördert von: West China Hospital of Stomatology, Sichuan University, Chinese Orthodontic Society, Chinese Stomatological Association, Natural Science Foundation of Sichuan Province.
Abstract (Originalsprache)
OBJECTIVE: This study evaluated the effect of offset on the precision of three-dimensional (3D)-printed splints, proposing to optimize the splint design to compensate for systematic errors. MATERIALS AND METHODS: 14 resin model sets were scanned and offset as a whole by given distances (0.05, 0.10, 0.15, 0.20, 0.25, 0.30, 0.35, and 0.40 mm). Intermediate splints (ISs) and final splints (FSs) were generated from the non-offset and offset models and grouped correspondingly, named as splint type-offset value, IS-0.05, for instance. Dentitions occluded with the splint were scanned. Translational and rotational deviations of the lower dentition relative to the upper dentition were 3D measured. RESULTS: Deviations of ISs and FSs were more evident in the vertical and pitch dimensions, and were mostly acceptable in other dimensions. ISs with offset ≥ 0.05 mm showed vertical deviations significantly below 1 mm (P < 0.05) while ISs with 0.10- to 0.30-mm offsets had pitch rotations significantly lower than 1° (P < 0.05). The Pitch of IS-0.35 was significantly larger than ISs with 0.15- to 0.30-mm offsets (P < 0.05). Meanwhile, FSs fit better as the offset increased and FSs with offsets ≥ 0.15 mm all had deviations significantly lower than 1 mm (for translation) or 1° (for rotation) (P < 0.05). CONCLUSIONS: Offset affects the precision of 3D-printed splints. Moderate offset values of 0.10 to 0.30 mm are recommendable for ISs. Offset values ≥ 0.15 mm are recommended for FSs in cases with stable final occlusion. CLINICAL RELEVANCE: This study found the optimal offset ranges for 3D-printed ISs and FSs via a standardized protocol.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- State Key Laboratory of Oral Diseases & National Clinical Research Center for Oral Diseases & Department of Orthodontics, West China Hospital of Stomatology, Sichuan University, No. 14, 3rd Section of Renmin Nan Road, C…
Förderung
- West China Hospital of Stomatology, Sichuan University
- Chinese Orthodontic Society, Chinese Stomatological Association
- Natural Science Foundation of Sichuan Province
Zitieren
APA
Wang, P., Wang, Y., Xu, H., Huang, Y., Shi, Y., Chen, S., Bai, D., & Xue, C. (2023). Effect of offset on the precision of 3D-printed orthognathic surgical splints. Clinical Oral Investigations. https://doi.org/10.1007/s00784-023-05134-8
Vancouver
Wang P, Wang Y, Xu H, Huang Y, Shi Y, Chen S, et al. Effect of offset on the precision of 3D-printed orthognathic surgical splints. Clinical Oral Investigations. 2023. doi:10.1007/s00784-023-05134-8
BibTeX
@article{Wang2023,
author = {Wang, Peiqi and Wang, Yipeng and Xu, Hui and Huang, Yixi and Shi, Yu and Chen, Siqi and Bai, Ding and Xue, Chaoran},
title = {Effect of offset on the precision of 3D-printed orthognathic surgical splints},
journal = {Clinical Oral Investigations},
year = {2023},
doi = {10.1007/s00784-023-05134-8}
}
EUVIMED