EUVIMEDEUVIMED entdecken
Crossref

Graphene-based dual-mode electrochemical-FET aptasensor for rapid zeptomolar detection of CEA

Anastasiia Kudriavtseva, Alexey V Orlov, Maxim P Nikitin, Konstantin Tsarik, Ivan Bobrinetskiy, Petr I Nikitin

Microchimica Acta · 2025

DOI: 10.1007/s00604-025-07710-z ↗

Kurz gefasst

Fachartikel in der Zeitschrift Microchimica Acta, 2025. Verfasst von Anastasiia Kudriavtseva und 5 weiteren Autor:innen. Beteiligte Einrichtung: Prokhorov General Physics Institute of the Russian Academy of Sciences, Moscow, 119991, Russia. lekassy2000@gmail.com. Hauptthemen laut MEDLINE: Graphite, Carcinoembryonic Antigen, Aptamers, Nucleotide, Electrochemical Techniques. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 39 Literaturangaben. Gefördert von: Russian Science Foundation, MSHE of RF.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

An electrochemical (EC) sensor has been developed integrated with a field effect transistor (FET) based on reduced graphene oxide (rGO) for ultrasensitive, rapid, and selective detection of carcinoembryonic antigen (CEA) over a wide dynamic range of up to eight orders of magnitude. An aptamer receptor was immobilized on the rGO surface, which had an initial hole mobility of 7 cm2/(V·s). Cyclic voltammetry and transfer curves measurements were used to monitor the biosensor assembly and to detect the CEA biomarker. This dual-mode aptasensor provides both high sensitivity across a wide dynamic range and the capability for mutual verification. The FET biosensor has a limit of detection as low as 60 ag/mL (0.33 zM), with a response time of 2 min. The overall dynamic range of the EC-FET aptasensor covers concentrations from 0.1 fg/mL to 10 ng/mL, with the FET mode spanning from 0.1 fg/mL to 10 fg/mL and for EC mode spanning from 1 fg/mL to 10 ng/mL. Thus, we have developed the scalable, low-cost method for fabricating a highly sensitive, label-free aptasensor with an improved dynamic range for CEA detection.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

GraphiteCarcinoembryonic AntigenAptamers, NucleotideElectrochemical TechniquesBiosensing TechniquesTransistors, ElectronicLimit of DetectionHumans

Stichwörter der Autor:innen

AptamerCarcinoembryonic antigenElectrochemical sensorField effect transistorGraphene oxideInterdigitated electrodes

Substanzen

GraphiteCarcinoembryonic AntigenAptamers, Nucleotidegraphene oxide

Autor:innen und Institutionen

  • Prokhorov General Physics Institute of the Russian Academy of Sciences, Moscow, 119991, Russia. lekassy2000@gmail.com.
  • Moscow Center for Advanced Studies, Kulakova Str. 20, Moscow, Russia. lekassy2000@gmail.com.
  • Prokhorov General Physics Institute of the Russian Academy of Sciences, Moscow, 119991, Russia.
  • Moscow Center for Advanced Studies, Kulakova Str. 20, Moscow, Russia.
  • Department of Nanobiomedicine, Sirius University of Science and Technology, 1 Olympic Ave., Sochi, 354340, Russia.
  • National Research University of Electronic Technology (MIET), Zelenograd, Moscow, 124498, Russia.

Förderung

  • Russian Science Foundation
  • MSHE of RF

Zitieren

APA

Kudriavtseva, A., Orlov, A. V., Nikitin, M. P., Tsarik, K., Bobrinetskiy, I., & Nikitin, P. I. (2025). Graphene-based dual-mode electrochemical-FET aptasensor for rapid zeptomolar detection of CEA. Microchimica Acta. https://doi.org/10.1007/s00604-025-07710-z

Vancouver

Kudriavtseva A, Orlov AV, Nikitin MP, Tsarik K, Bobrinetskiy I, Nikitin PI. Graphene-based dual-mode electrochemical-FET aptasensor for rapid zeptomolar detection of CEA. Microchimica Acta. 2025. doi:10.1007/s00604-025-07710-z

BibTeX
@article{Kudriavtseva2025,
  author = {Kudriavtseva, Anastasiia and Orlov, Alexey V and Nikitin, Maxim P and Tsarik, Konstantin and Bobrinetskiy, Ivan and Nikitin, Petr I},
  title = {Graphene-based dual-mode electrochemical-FET aptasensor for rapid zeptomolar detection of CEA},
  journal = {Microchimica Acta},
  year = {2025},
  doi = {10.1007/s00604-025-07710-z}
}
Kostenlos starten →