EUVIMEDEUVIMED entdecken
Crossref

Selectivity optimization of real-time and continuous sensing of endogenous H2S in biological fluids

Na Zou, Xin Li, Meiling Xu, Zhaoxia Wang, Junhua Zhang, Xueliang Wang

Microchimica Acta · 2025

DOI: 10.1007/s00604-025-07298-4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Microchimica Acta, 2025. Verfasst von Na Zou und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Chemistry and Chemical Engineering, Heze University, Heze, 274015, PR China. Hauptthemen laut MEDLINE: Hydrogen Sulfide, Electrochemical Techniques. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 67 Literaturangaben. Gefördert von: doctor's Foundation of Heze University, the Natural Science Foundation of Shandong Province, China, doctor’s Foundation of Heze University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In the direct electrochemical sensing of endogenous hydrogen sulfide, the utilization of triple-pulse amperometry (TPA) enables the delivery of distinct cleaning and measurement pulses, effectively mitigating electrode surface passivation due to sulfur deposition. In order to further improve the sensor's sensitivity and selectivity, gold nanoparticles(nano-Au), platinum nanoparticles (nano-Pt), and poly(3,4-ethylenedioxythiophene) (PEDOT) were separately employed to modify electrodes for constructing electrochemical sensors. Their performance in detecting hydrogen sulfide was evaluated using constant potential amperometry (CPA) alongside TPA. Selectivity coefficients were determined based on current responses to hydrogen sulfide as well as four major interfering substances: ascorbic acid (AA), dopamine (DA), uric acid (UA), and epinephrine (EP). After quantitative comparison, the optimal solution for direct electrochemical sensing of hydrogen sulfide involved employing a PEDOT/nano-Au composite film in conjunction with TPA technology. The sensor responds to hydrogen sulfide in the concentration range 3.0-24.0 µM with a detection limit of 0.035 µM. Furthermore, the sensor demonstrates excellent repeatability and stability, rendering it suitable for continuous electrochemical monitoring of hydrogen sulfide in simulated real biological environments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hydrogen SulfideElectrochemical TechniquesGoldMetal NanoparticlesHumansPlatinumPolymersBridged Bicyclo Compounds, HeterocyclicLimit of DetectionElectrodesAscorbic AcidDopamine

Stichwörter der Autor:innen

Electrochemical sensingEndogenous hydrogen sulfideSelectivity coefficientsTriple-pulse amperometry

Substanzen

Hydrogen SulfideGoldpoly(3,4-ethylene dioxythiophene)PlatinumPolymersBridged Bicyclo Compounds, HeterocyclicAscorbic AcidDopamine

Autor:innen und Institutionen

  • School of Chemistry and Chemical Engineering, Heze University, Heze, 274015, PR China.
  • State Key Laboratory of Polymer Materials Engineering, Polymer Research Institute of Sichuan University, Chengdu, 610065, PR China.
  • School of Chemistry and Chemical Engineering, Heze University, Heze, 274015, PR China. qust-1977wxl@163.com.

Förderung

  • doctor's Foundation of Heze University
  • the Natural Science Foundation of Shandong Province, China
  • doctor’s Foundation of Heze University

Zitieren

APA

Zou, N., Li, X., Xu, M., Wang, Z., Zhang, J., & Wang, X. (2025). Selectivity optimization of real-time and continuous sensing of endogenous H2S in biological fluids. Microchimica Acta. https://doi.org/10.1007/s00604-025-07298-4

Vancouver

Zou N, Li X, Xu M, Wang Z, Zhang J, Wang X. Selectivity optimization of real-time and continuous sensing of endogenous H2S in biological fluids. Microchimica Acta. 2025. doi:10.1007/s00604-025-07298-4

BibTeX
@article{Zou2025,
  author = {Zou, Na and Li, Xin and Xu, Meiling and Wang, Zhaoxia and Zhang, Junhua and Wang, Xueliang},
  title = {Selectivity optimization of real-time and continuous sensing of endogenous H2S in biological fluids},
  journal = {Microchimica Acta},
  year = {2025},
  doi = {10.1007/s00604-025-07298-4}
}
Kostenlos starten →