A surgeon-controlled mode of robotic assistance with posterolateral approach helps achieve highly medialized cervical pedicle screw placement to avoid vertebral artery injury
European Spine Journal · 2025
DOI: 10.1007/s00586-025-08753-9 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Spine Journal, 2025. Verfasst von Yoichi Tani und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Kansai Medical University, 2-5-1 Shinmachi, Hirakata City, Osaka, 573-1010, Japan. taniyoic@gmail.com. Hauptthemen laut MEDLINE: Pedicle Screws, Vertebral Artery, Cervical Vertebrae, Robotic Surgical Procedures. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: The small pedicle widths, the thin lateral cortical shell next to the vertebral artery (VA), and the strongly tilted pedicle axis anteromedially narrow the safety margin for cervical pedicle screw (CPS) placement against vertebral artery (VA) injury. We have studied whether the minimally invasive surgery (MIS) of a posterolateral approach with a table-mounted, surgeon-controlled mode of robotic assistance (group R) improves the VA safety compared to C-arm fluoroscopy-guided conventional open technique (group F). METHODS: Group R consisted of 165 screws in 37 patients and group F, 199 screws in 52 patients. The two groups covered a broad range of vertebral levels from C2 to C7 with comparable distribution (p = 0.0512) for treating a similar variety of diseases (p = 0.6958). RESULTS: Group R, compared with group F, showed a greater lateral-to-medial CPS inclination (p ≤ 0.0004) that even exceeded the obliquity of corresponding pedicle axis, leading to a higher rate of acceptable CPS placement (93.4% vs. 85.4%; p = 0.0164) with a lower rate of lateral breach (1.2% vs. 10.1%; p = 0.0004). CONCLUSION: For CPS placement, robot-assisted MIS obviously eliminates morbidity-prone soft-tissue dissection, radiation exposure to the surgical team, and human manual errors. The current study revealed its additional benefit of better safety against VA injury by allowing us to place CPS with a steep lateral-to-medial angulation owing to (1) a lack of counter pressure from the paravertebral muscles and (2) minimizing a navigation pitfall of untracked pressure-induced vertebral rotation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedic Surgery, Kansai Medical University, 2-5-1 Shinmachi, Hirakata City, Osaka, 573-1010, Japan. taniyoic@gmail.com.
- Department of Orthopaedic Surgery, Kansai Medical University, 2-5-1 Shinmachi, Hirakata City, Osaka, 573-1010, Japan.
ORCID-Profile: Yoichi Tani
Zitieren
APA
Tani, Y., Naka, N., Ono, N., Kawashima, K., Paku, M., Ishihara, M., Adachi, T., Taniguchi, S., Ando, M., & Saito, T. (2025). A surgeon-controlled mode of robotic assistance with posterolateral approach helps achieve highly medialized cervical pedicle screw placement to avoid vertebral artery injury. European Spine Journal. https://doi.org/10.1007/s00586-025-08753-9
Vancouver
Tani Y, Naka N, Ono N, Kawashima K, Paku M, Ishihara M, et al. A surgeon-controlled mode of robotic assistance with posterolateral approach helps achieve highly medialized cervical pedicle screw placement to avoid vertebral artery injury. European Spine Journal. 2025. doi:10.1007/s00586-025-08753-9
BibTeX
@article{Tani2025,
author = {Tani, Yoichi and Naka, Nobuhiro and Ono, Naoto and Kawashima, Koki and Paku, Masaaki and Ishihara, Masayuki and Adachi, Takashi and Taniguchi, Shinichirou and Ando, Muneharu and Saito, Takanori},
title = {A surgeon-controlled mode of robotic assistance with posterolateral approach helps achieve highly medialized cervical pedicle screw placement to avoid vertebral artery injury},
journal = {European Spine Journal},
year = {2025},
doi = {10.1007/s00586-025-08753-9}
}
EUVIMED