EUVIMEDEUVIMED entdecken
BeobachtungsstudieOpen Access · CC BY 4.0

Cervical pedicle screw placement with patient-specific 3D-printed guides: accuracy and safety in a clinical experience

Nicola Marengo, Stefano Colonna iD, Enrico Lo Bue, Alessandro Pesaresi, Ayoub Saaid, Mario Allevi, Marco Ajello, Geert Mahieu, Diego Garbossa, Fabio Cofano

European Spine Journal · 2025

DOI: 10.1007/s00586-025-08679-2 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift European Spine Journal, 2025. Verfasst von Nicola Marengo und 9 weiteren Autor:innen. Beteiligte Einrichtung: Neurosurgery, Department of Neuroscience, University of Turin, Turin, Italy. Hauptthemen laut MEDLINE: Pedicle Screws, Printing, Three-Dimensional, Cervical Vertebrae, Spinal Fusion. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Università degli Studi di Torino.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Cervical pedicle screw (CPS) instrumentation offers significant biomechanical advantages compared to lateral mass or transarticular fixation. Nonetheless, malpositioning complications constitute a relevant concern. Customized patient-specific 3D-printed templates have been developed to improve CPS placement accuracy and safety. The aim of this study is to present our experience with this surgical technique and its accuracy and safety in a clinical setting.

Methods

This single-center retrospective observational study of prospectively collected data included patients undergoing CPS fixation surgery using a patient-specific 3D template guide system. All patients underwent a 3D-volumetric high-resolution CT scan of the cervical spine for preoperative surgical planning. Postoperative CT scans were used to evaluate pedicle perforation, CPS trajectories, and deviations between the planned and the actual screw position.

Results

A total of 115 CPS were implanted in 25 patients, with 107 (93.1%) of the screws completely placed inside the pedicle. Cortical breach within 2 mm was observed in 8 (6.9%) cases, with no cases of more severe pedicle infractions or perioperative neurovascular complication. No differences of CPS accuracies were found between each metameric fusion level, and between monolateral or bilateral templates. Mean total deviations were 0.75 mm vertically and 0.51 mm horizontally at the screw entry point, and 0.72 mm vertically and horizontally at the narrowest pedicle point. Mean total sagittal and transverse angular deviations were 2.94° and 3.04°, respectively.

Conclusion

Cervical pedicle screw placement using patient-specific guides is safe and accurate, supporting the feasibility of this technique in posterior cervical spine fusion surgery.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pedicle ScrewsPrinting, Three-DimensionalCervical VertebraeSpinal FusionSurgery, Computer-AssistedHumansMaleFemaleMiddle AgedRetrospective StudiesAgedAdultTomography, X-Ray Computed

Stichwörter der Autor:innen

Cervical pedicle screwPatient-specific templatePedicle screw instrumentationThree-dimensional model

Autor:innen und Institutionen

  • Neurosurgery, Department of Neuroscience, University of Turin, Turin, Italy.
  • Neurosurgery, Department of Neuroscience, University of Turin, Turin, Italy. stefano.colonna@unito.it.
  • Orthopedic Surgery Department, ORTHOCA - AZ Monica Hospital, Antwerp, Belgium.

ORCID-Profile: Stefano Colonna

Förderung

  • Università degli Studi di Torino

Zitieren

APA

Marengo, N., Colonna, S., Bue, E. L., Pesaresi, A., Saaid, A., Allevi, M., Ajello, M., Mahieu, G., Garbossa, D., & Cofano, F. (2025). Cervical pedicle screw placement with patient-specific 3D-printed guides: accuracy and safety in a clinical experience. European Spine Journal. https://doi.org/10.1007/s00586-025-08679-2

Vancouver

Marengo N, Colonna S, Bue EL, Pesaresi A, Saaid A, Allevi M, et al. Cervical pedicle screw placement with patient-specific 3D-printed guides: accuracy and safety in a clinical experience. European Spine Journal. 2025. doi:10.1007/s00586-025-08679-2

BibTeX
@article{Marengo2025,
  author = {Marengo, Nicola and Colonna, Stefano and Bue, Enrico Lo and Pesaresi, Alessandro and Saaid, Ayoub and Allevi, Mario and Ajello, Marco and Mahieu, Geert and Garbossa, Diego and Cofano, Fabio},
  title = {Cervical pedicle screw placement with patient-specific 3D-printed guides: accuracy and safety in a clinical experience},
  journal = {European Spine Journal},
  year = {2025},
  doi = {10.1007/s00586-025-08679-2}
}