Endplate weakening during cage bed preparation significantly reduces endplate load capacity

Anna-Katharina Calek iD, Frédéric Cornaz, Mauro Suter, Marie-Rosa Fasser, Mazda Farshad, Jonas Widmer

European Spine Journal · 2024

DOI: 10.1007/s00586-024-08289-4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Spine Journal, 2024. Verfasst von Anna-Katharina Calek und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedics, Balgrist University Hospital, University of Zurich, Forchstrasse 340, 8008, Zurich, Switzerland. anna-katharina.calek@balg…. Hauptthemen laut MEDLINE: Spinal Fusion, Lumbar Vertebrae, Weight-Bearing. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Zurich.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To analyze the effect of endplate weakness prior to PLIF or TLIF cage implantation and compare it to the opposite intact endplate of the same vertebral body. In addition, the influence of bone quality on endplate resistance was investigated.

Methods

Twenty-two human lumbar vertebrae were tested in a ramp-to-failure test. One endplate of each vertebral body was tested intact and the other after weakening with a rasp (over an area of 200 mm 2 ). Either a TLIF or PLIF cage was then placed and the compression load was applied across the cage until failure of the endplate. Failure was defined as the first local maximum of the force measurement. Bone quality was assessed by determining the Hounsfield units (HU) on CT images.

Results

With an intact endplate and a TLIF cage, the median force to failure was 1276.3N (693.1–1980.6N). Endplate weakening reduced axial endplate resistance to failure by 15% (0–23%). With an intact endplate and a PLIF cage, the median force to failure was 1057.2N (701.2–1735.5N). Endplate weakening reduced axial endplate resistance to failure by 36.6% (7–47.9%). Bone quality correlated linearly with the force at which endplate failure occurred. Intact and weakened endplates showed a strong positive correlation: intact-TLIF: r = 0.964, slope of the regression line (slope) = 11.8, p < 0.001; intact-PLIF: r = 0.909, slope = 11.2, p = 5.5E−05; weakened-TLIF: r = 0.973, slope = 12.5, p < 0.001; weakened-PLIF: r = 0.836, slope = 6, p = 0.003.

Conclusion

Weakening of the endplate during cage bed preparation significantly reduces the resistance of the endplate to subsidence to failure: endplate load capacity is reduced by 15% with TLIF and 37% with PLIF. Bone quality correlates with the force at which endplate failure occurs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Spinal FusionLumbar VertebraeWeight-BearingHumansMiddle AgedMaleAgedFemaleBiomechanical PhenomenaAdultAged, 80 and over

Stichwörter der Autor:innen

Biomechanical studyCageDorsal fusionEndplatePLIFSubsidenceTLIF

Autor:innen und Institutionen

  • Department of Orthopedics, Balgrist University Hospital, University of Zurich, Forchstrasse 340, 8008, Zurich, Switzerland. anna-katharina.calek@balgrist.ch.
  • Department of Orthopedics, Balgrist University Hospital, University of Zurich, Forchstrasse 340, 8008, Zurich, Switzerland.
  • Spine Biomechanics, Department of Orthopedic Surgery, Balgrist University Hospital, University of Zurich, Zurich, Switzerland.
  • University Spine Center Zurich, Balgrist University Hospital, University of Zurich, Zurich, Switzerland.

ORCID-Profile: Anna-Katharina Calek

Förderung

  • University of Zurich

Zitieren

APA

Calek, A. K., Cornaz, F., Suter, M., Fasser, M. R., Farshad, M., & Widmer, J. (2024). Endplate weakening during cage bed preparation significantly reduces endplate load capacity. European Spine Journal. https://doi.org/10.1007/s00586-024-08289-4

Vancouver

Calek AK, Cornaz F, Suter M, Fasser MR, Farshad M, Widmer J. Endplate weakening during cage bed preparation significantly reduces endplate load capacity. European Spine Journal. 2024. doi:10.1007/s00586-024-08289-4

BibTeX
@article{Calek2024,
  author = {Calek, Anna-Katharina and Cornaz, Frédéric and Suter, Mauro and Fasser, Marie-Rosa and Farshad, Mazda and Widmer, Jonas},
  title = {Endplate weakening during cage bed preparation significantly reduces endplate load capacity},
  journal = {European Spine Journal},
  year = {2024},
  doi = {10.1007/s00586-024-08289-4}
}
Kostenlos starten →