Novel force feedback technology improves suturing in robotic-assisted surgery: a pre-clinical study

Elliot L Servais, Laila Rashidi, Priyanshi Porwal, Mark Garibaldi, Andrew J Hung iD

Surgical Endoscopy · 2025-02 · 39 (2) · 1217-1226

DOI: 10.1007/s00464-024-11472-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Surgical Endoscopy, online erschienen am 30.12.2024. Verfasst von Elliot L Servais und 4 weiteren Autor:innen. Beteiligte Einrichtung: Lahey Hospital and Medical Center, UMass Chan Medical School, Burlington, MA, USA. Hauptthemen laut MEDLINE: Suture Techniques, Robotic Surgical Procedures, Clinical Competence. Die Arbeit wurde bisher 28-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Intuitive Surgical, Cedars-Sinai Medical Center.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

The inability to sense force applied to tissue is suggested as a limitation to robotic-assisted surgery (RAS). This pre-clinical study evaluated the impact of a novel force feedback (FFB) technology, integrated on a next-generation robotic system that allows surgeons to sense forces exerted at the instrument tips, on suturing performance by novice surgeons during RAS.

Methods

Twenty-nine novice surgeons (< 50 RAS cases in the last 5 years) were randomized into two groups with ( n = 15) or without ( n = 14) FFB sensing. Participants performed interrupted stitches on ex vivo porcine bladder and running stitches on porcine aorta (Fig. 1A) over four runs. Average forces applied, number of errors, time for exercise completion, and Robotic Anastomosis Competence Evaluation (RACE) technical skill ratings were compared using a three-way mixed-model ANOVA and applicable post hoc tests.

Results

FFB sensing significantly lowered the mean force applied (bladder, 1.71 N vs 2.40 N, p < 0.006; aorta, 1.80 N vs 2.53 N, p < 0.006), average number of errors (bladder, 0.59 vs 1.76, p < 0.001; aorta, 0.38 vs 1.14, p < 0.001), and the time to completion (bladder, 659 s vs 781 s, p = 0.002; aorta, 460 s vs 570 s, p = 0.001) (Fig. 1C). The FFB group applied less tissue trauma with a higher RACE skill score (3.75 vs 3.03, p = 0.012).

Conclusion

This study showed that novice surgeons using FFB-enabled instruments completed suturing tasks using less force, with fewer errors, taking less time, and less tissue trauma during RAS. Future studies are required to better understand the impact of FFB technology on surgical performance and potential patient benefits.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Suture TechniquesRobotic Surgical ProceduresClinical CompetenceAnimalsSwineHumansUrinary BladderAortaAnastomosis, Surgical

Stichwörter der Autor:innen

Force feedbackSurgical trainingSuturing

Autor:innen und Institutionen

  • Lahey Hospital and Medical Center, UMass Chan Medical School, Burlington, MA, USA.
  • MultiCare Health System, Seattle, WA, USA.
  • Intuitive Surgical, Sunnyvale, CA, USA.
  • Cedars-Sinai Medical Center, Los Angeles, CA, USA. andrew.hung@cshs.org.

ORCID-Profile: Andrew J Hung

Förderung

  • Intuitive Surgical
  • Cedars-Sinai Medical Center

Zitieren

APA

Servais, E. L., Rashidi, L., Porwal, P., Garibaldi, M., & Hung, A. J. (2025). Novel force feedback technology improves suturing in robotic-assisted surgery: a pre-clinical study. Surgical Endoscopy, 39(2), 1217-1226. https://doi.org/10.1007/s00464-024-11472-9

Vancouver

Servais EL, Rashidi L, Porwal P, Garibaldi M, Hung AJ. Novel force feedback technology improves suturing in robotic-assisted surgery: a pre-clinical study. Surgical Endoscopy. 2025;39(2):1217-1226. doi:10.1007/s00464-024-11472-9

BibTeX
@article{Servais2025,
  author = {Servais, Elliot L and Rashidi, Laila and Porwal, Priyanshi and Garibaldi, Mark and Hung, Andrew J},
  title = {Novel force feedback technology improves suturing in robotic-assisted surgery: a pre-clinical study},
  journal = {Surgical Endoscopy},
  year = {2025},
  volume = {39},
  number = {2},
  pages = {1217--1226},
  doi = {10.1007/s00464-024-11472-9}
}