Novel force feedback technology improves suturing in robotic-assisted surgery: a pre-clinical study
Surgical Endoscopy · 2025-02 · 39 (2) · 1217-1226
DOI: 10.1007/s00464-024-11472-9 ↗Kurz gefasst
Fachartikel in der Zeitschrift Surgical Endoscopy, online erschienen am 30.12.2024. Verfasst von Elliot L Servais und 4 weiteren Autor:innen. Beteiligte Einrichtung: Lahey Hospital and Medical Center, UMass Chan Medical School, Burlington, MA, USA. Hauptthemen laut MEDLINE: Suture Techniques, Robotic Surgical Procedures, Clinical Competence. Die Arbeit wurde bisher 28-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Intuitive Surgical, Cedars-Sinai Medical Center.
Abstract (Originalsprache)
Introduction
The inability to sense force applied to tissue is suggested as a limitation to robotic-assisted surgery (RAS). This pre-clinical study evaluated the impact of a novel force feedback (FFB) technology, integrated on a next-generation robotic system that allows surgeons to sense forces exerted at the instrument tips, on suturing performance by novice surgeons during RAS.
Methods
Twenty-nine novice surgeons (< 50 RAS cases in the last 5 years) were randomized into two groups with ( n = 15) or without ( n = 14) FFB sensing. Participants performed interrupted stitches on ex vivo porcine bladder and running stitches on porcine aorta (Fig. 1A) over four runs. Average forces applied, number of errors, time for exercise completion, and Robotic Anastomosis Competence Evaluation (RACE) technical skill ratings were compared using a three-way mixed-model ANOVA and applicable post hoc tests.
Results
FFB sensing significantly lowered the mean force applied (bladder, 1.71 N vs 2.40 N, p < 0.006; aorta, 1.80 N vs 2.53 N, p < 0.006), average number of errors (bladder, 0.59 vs 1.76, p < 0.001; aorta, 0.38 vs 1.14, p < 0.001), and the time to completion (bladder, 659 s vs 781 s, p = 0.002; aorta, 460 s vs 570 s, p = 0.001) (Fig. 1C). The FFB group applied less tissue trauma with a higher RACE skill score (3.75 vs 3.03, p = 0.012).
Conclusion
This study showed that novice surgeons using FFB-enabled instruments completed suturing tasks using less force, with fewer errors, taking less time, and less tissue trauma during RAS. Future studies are required to better understand the impact of FFB technology on surgical performance and potential patient benefits.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Lahey Hospital and Medical Center, UMass Chan Medical School, Burlington, MA, USA.
- MultiCare Health System, Seattle, WA, USA.
- Intuitive Surgical, Sunnyvale, CA, USA.
- Cedars-Sinai Medical Center, Los Angeles, CA, USA. andrew.hung@cshs.org.
ORCID-Profile: Andrew J Hung
Förderung
- Intuitive Surgical
- Cedars-Sinai Medical Center
Zitieren
APA
Servais, E. L., Rashidi, L., Porwal, P., Garibaldi, M., & Hung, A. J. (2025). Novel force feedback technology improves suturing in robotic-assisted surgery: a pre-clinical study. Surgical Endoscopy, 39(2), 1217-1226. https://doi.org/10.1007/s00464-024-11472-9
Vancouver
Servais EL, Rashidi L, Porwal P, Garibaldi M, Hung AJ. Novel force feedback technology improves suturing in robotic-assisted surgery: a pre-clinical study. Surgical Endoscopy. 2025;39(2):1217-1226. doi:10.1007/s00464-024-11472-9
BibTeX
@article{Servais2025,
author = {Servais, Elliot L and Rashidi, Laila and Porwal, Priyanshi and Garibaldi, Mark and Hung, Andrew J},
title = {Novel force feedback technology improves suturing in robotic-assisted surgery: a pre-clinical study},
journal = {Surgical Endoscopy},
year = {2025},
volume = {39},
number = {2},
pages = {1217--1226},
doi = {10.1007/s00464-024-11472-9}
}