IDEAL-D Phase 0 Evaluation of the Avatera System in Robot-Assisted Prostate, Bladder and Renal Surgery
Journal of Laparoendoscopic & Advanced Surgical Techniques · 01.03.2024 · 34 (3) · 239-245
DOI: 10.1089/lap.2023.0454 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Laparoendoscopic & Advanced Surgical Techniques, online erschienen am 01.03.2024. Verfasst von Caelan-Max Haney und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, University Hospital Leipzig, Leipzig, Germany. Hauptthemen laut MEDLINE: Robotics, Robotic Surgical Procedures. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben.
Abstract (Originalsprache)
Purpose
To evaluate the utilization of novel Avatera system in urological operations according to the IDEAL-D framework recommendations for high-risk invasive surgical devices. Materials and Methods: Three surgeons attempted to perform 23 upper and lower urinary tract operations on human cadavers and in live porcine models using the Avatera system. Total operative time and the duration of the substeps were evaluated. Surgical performance was assessed with the Global Evaluative Assessment of Robotic Skills (GEARS) score. Suturing was rated using the technical checklist for the assessment of suturing in robotic surgery. Attending surgeons rated their satisfaction with the Avatera system on a scale of 1–5. Results and Limitation: Seventeen out of 18 operations performed on cadavers were completed, while one pyeloplasty was discontinued. All five operations performed in porcine models were completed. Although 1 pig was euthanized on the fifth postoperative day, its symptoms were unrelated to surgery. Mean GEARS and Suturing scores in the upper urinary tract were 29 ± 0.7 and 29.5 ± 0.95, respectively, and in the lower urinary 28.5 ± 1.2 and 29.5 ± 0.5, respectively. Surgeons' satisfaction was high or very high for all procedures.
Conclusions
The Avatera system was associated with good surgical performance and high surgeons' satisfaction rates. All urological procedures performed were shown to be feasible, with comparable risks to other robot-assisted surgery systems.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Urology, University Hospital Leipzig, Leipzig, Germany.
- Department of Urology, University of Patras, Patras, Greece.
- Department of Urology, Clinic for Urology and Urologic Oncology, Hannover Medical School, Hannover, Germany.
ORCID-Profile: Evagelos Liatsikos
Zitieren
APA
Haney, C. M., Holze, S., Liatsikos, E., Dietel, A., Kallidonis, P., Tatanis, V., Katsakiori, P., Spinos, T., Imkamp, F., & Stolzenburg, J. U. (2024). IDEAL-D Phase 0 Evaluation of the Avatera System in Robot-Assisted Prostate, Bladder and Renal Surgery. Journal of Laparoendoscopic & Advanced Surgical Techniques, 34(3), 239-245. https://doi.org/10.1089/lap.2023.0454
Vancouver
Haney CM, Holze S, Liatsikos E, Dietel A, Kallidonis P, Tatanis V, et al. IDEAL-D Phase 0 Evaluation of the Avatera System in Robot-Assisted Prostate, Bladder and Renal Surgery. Journal of Laparoendoscopic & Advanced Surgical Techniques. 2024;34(3):239-245. doi:10.1089/lap.2023.0454
BibTeX
@article{Haney2024,
author = {Haney, Caelan-Max and Holze, Sigrun and Liatsikos, Evagelos and Dietel, Anja and Kallidonis, Panagiotis and Tatanis, Vasileios and Katsakiori, Paraskevi and Spinos, Theodoros and Imkamp, Florian and Stolzenburg, Jens-Uwe},
title = {IDEAL-D Phase 0 Evaluation of the Avatera System in Robot-Assisted Prostate, Bladder and Renal Surgery},
journal = {Journal of Laparoendoscopic & Advanced Surgical Techniques},
year = {2024},
volume = {34},
number = {3},
pages = {239--245},
doi = {10.1089/lap.2023.0454}
}
EUVIMED