Role of dedicated port cleaning devices in laparoscopic surgery

Shinnosuke Nagano, Shota Fujii, Kota Momose, Kotaro Yamashita, Takuro Saito, Koji Tanaka, Kazuyoshi Yamamoto, Tomoki Makino, Tsuyoshi Takahashi, Yukinori Kurokawa, Hidetoshi Eguchi, Yuichiro Doki und 1 weitere

Surgical Endoscopy · 2024-12 · 38 (12) · 7613-7620

DOI: 10.1007/s00464-024-11366-w ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Surgical Endoscopy, online erschienen am 11.11.2024. Verfasst von Shinnosuke Nagano und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Next Generation Endoscopic Intervention (Project ENGINE), Osaka University Graduate School of Medicine, 1-3 Yamadaoka, Suita, Osaka, 56…. Hauptthemen laut MEDLINE: Laparoscopy, Laparoscopes. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Osaka University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Various techniques have been used to prevent smudge on a laparoscope when inserting through trocars; however, there has been no standardized method. The purpose of this study was to compare the performance of different cleaning techniques with or without using dedicated devices, and to evaluate the features of cleaning devices.

Methods

The smudge was created in the standard 12-mm and 5-mm ports using pseudo-blood, and port cleaning was attempted using 5 different methods: (1) a surgical gauze + surgical forceps, (2) a surgical gauze + laparoscopic forceps, (3) a small laparoscopic gauze + laparoscopic forceps, (4) a cylinder-type cleaner (Endo Wiper; Osaki Medical), and (5) a swab-type cleaner (Port Cleaner; Hakuzo Medical). The “port cleaning rate” was calculated by measuring the absorbance of remained pseudo-blood after single cleaning procedure using UV spectrophotometry. In addition, the port cleaning rate was compared between two dedicated devices after multiple (5 times) cleaning procedures.

Results

The two dedicated devices had a statistically higher cleaning rate for 12-mm port than the methods using surgical gauze (p < 0.05). Regarding the 5-mm port, a swab-type cleaner showed the highest cleaning rate than the gauze method and a cylinder-type cleaner (p < 0.05). After multiple cleaning procedures for 12-mm port, cleaning rate of a swab-type cleaner decreased by an average of 5.4% (p = 0.044), but cleaning rate did not decrease for a cylinder-type cleaner. Regarding the 5-mm port, cleaning rate statistically decreased for both two dedicated devices (p < 0.01).

Conclusion

Higher port cleaning rates were observed in techniques using dedicated devices. A swab-type cleaner had better port cleaning rate in single use, especially for the 5-mm port. A cylinder-type cleaner showed higher durability in cleaning 12-mm port. The features of these dedicated devices should be well understood, and cleaning methods should be selected according to the environment and surgical techniques.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LaparoscopyLaparoscopesHumansEquipment ContaminationEquipment Design

Stichwörter der Autor:innen

Cleaning deviceLaparoscopic surgeryPort cleaning

Autor:innen und Institutionen

  • Department of Next Generation Endoscopic Intervention (Project ENGINE), Osaka University Graduate School of Medicine, 1-3 Yamadaoka, Suita, Osaka, 565-0871, Japan.
  • Department of Gastroenterological Surgery, Osaka University Graduate School of Medicine, Osaka, Japan.
  • SANYO Co., LTD, Osaka, Japan.
  • Department of Next Generation Endoscopic Intervention (Project ENGINE), Osaka University Graduate School of Medicine, 1-3 Yamadaoka, Suita, Osaka, 565-0871, Japan. knakajima@gesurg.med.osaka-u.ac.jp.
  • Department of Gastroenterological Surgery, Osaka University Graduate School of Medicine, Osaka, Japan. knakajima@gesurg.med.osaka-u.ac.jp.

ORCID-Profile: Kiyokazu Nakajima

Förderung

  • Osaka University

Zitieren

APA

Nagano, S., Fujii, S., Momose, K., Yamashita, K., Saito, T., Tanaka, K., Yamamoto, K., Makino, T., Takahashi, T., Kurokawa, Y., Eguchi, H., Doki, Y., & Nakajima, K. (2024). Role of dedicated port cleaning devices in laparoscopic surgery. Surgical Endoscopy, 38(12), 7613-7620. https://doi.org/10.1007/s00464-024-11366-w

Vancouver

Nagano S, Fujii S, Momose K, Yamashita K, Saito T, Tanaka K, et al. Role of dedicated port cleaning devices in laparoscopic surgery. Surgical Endoscopy. 2024;38(12):7613-7620. doi:10.1007/s00464-024-11366-w

BibTeX
@article{Nagano2024,
  author = {Nagano, Shinnosuke and Fujii, Shota and Momose, Kota and Yamashita, Kotaro and Saito, Takuro and Tanaka, Koji and Yamamoto, Kazuyoshi and Makino, Tomoki and Takahashi, Tsuyoshi and Kurokawa, Yukinori and Eguchi, Hidetoshi and Doki, Yuichiro and Nakajima, Kiyokazu},
  title = {Role of dedicated port cleaning devices in laparoscopic surgery},
  journal = {Surgical Endoscopy},
  year = {2024},
  volume = {38},
  number = {12},
  pages = {7613--7620},
  doi = {10.1007/s00464-024-11366-w}
}
Kostenlos starten →