Sonography versus anterior segment OCT for imaging bleb morphology after glaucoma drainage device surgery

Pascal Schipper, Benjamin Aretz, Leonie Bourauel, Carolin Deubel, Michael Petrak, Frank G Holz, Raffael Liegl, Karl Mercieca, Constance Liegl

Graefe's Archive for Clinical and Experimental Ophthalmology · 2026-11 · 264 (11) · 3593-3599

DOI: 10.1007/s00417-026-07276-3 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Graefe's Archive for Clinical and Experimental Ophthalmology, online erschienen am 20.05.2026. Verfasst von Pascal Schipper und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, University of Bonn, Bonn, Germany. Hauptthemen laut MEDLINE: Tomography, Optical Coherence, Glaucoma Drainage Implants, Anterior Eye Segment, Glaucoma. Die Arbeit nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum Bonn.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The PAUL ® Glaucoma Implant (PGI) is increasingly used in the surgical management of complex glaucoma, but detailed evaluation of its subconjunctival bleb morphology remains challenging due to its posterior location. While ultrasonography has long been used for bleb imaging, anterior segment optical coherence tomography (AS-OCT) offers higher-resolution imaging of anterior structures. This study aimed to directly compare ultrasonography and AS-OCT in assessing bleb morphology following PGI implantation.

Methods

In this prospective observational study, 20 patients (21 eyes) with PGI underwent postoperative assessment with both AS-OCT and B-scan ultrasonography. Bleb configuration (single- vs. double-layered), bleb height, and spatial extension were evaluated and compared between modalities.

Results

Mean age was 64.0 years (SD = 16.6); 80% were male. Sonography identified double-layered blebs in 95.2% of eyes, compared to 47.7% with AS-OCT, which failed to fully visualize bleb layers in 38.1% of eyes. Sonography consistently measured greater bleb heights and lengths than AS-OCT (total bleb height 3.76 ± 1.39 mm vs. 1.77 ± 0.71 mm, p < 0.001; maximum bleb length 17.24 ± 3.17 mm vs. 2.85 ± 1.03 mm, p < 0.001).

Conclusions

Sonography provided more comprehensive visualization of PGI blebs, particularly in detecting double-layered configurations, total height and measuring full posterior extension, whereas AS-OCT offered detailed anterior segment assessment including the tube entry site and patch graft. Both modalities yield complementary information, and their combined use may enhance postoperative monitoring and early detection of bleb dysfunction.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, Optical CoherenceGlaucoma Drainage ImplantsAnterior Eye SegmentGlaucomaIntraocular PressureConjunctivaHumansProspective StudiesMaleMiddle AgedFemaleUltrasonographyAgedFollow-Up Studies

Stichwörter der Autor:innen

Anterior segment OCTGlaucoma drainage deviceGlaucoma surgeryPAUL® glaucoma implantSonography

Autor:innen und Institutionen

  • Department of Ophthalmology, University of Bonn, Bonn, Germany.
  • Institute for Medical Biometry, Informatics and Epidemiology, University of Bonn, Bonn, Germany.
  • Department of Ophthalmology, Klinikum Dortmund , University Witten/Herdecke, Dortmund, Germany.
  • Department of Ophthalmology, University of Bonn, Bonn, Germany. constance.liegl@ukbonn.de.
  • Department of Ophthalmology, Klinikum Dortmund , University Witten/Herdecke, Dortmund, Germany. constance.liegl@ukbonn.de.

Förderung

  • Universitätsklinikum Bonn

Zitieren

APA

Schipper, P., Aretz, B., Bourauel, L., Deubel, C., Petrak, M., Holz, F. G., Liegl, R., Mercieca, K., & Liegl, C. (2026). Sonography versus anterior segment OCT for imaging bleb morphology after glaucoma drainage device surgery. Graefe's Archive for Clinical and Experimental Ophthalmology, 264(11), 3593-3599. https://doi.org/10.1007/s00417-026-07276-3

Vancouver

Schipper P, Aretz B, Bourauel L, Deubel C, Petrak M, Holz FG, et al. Sonography versus anterior segment OCT for imaging bleb morphology after glaucoma drainage device surgery. Graefe's Archive for Clinical and Experimental Ophthalmology. 2026;264(11):3593-3599. doi:10.1007/s00417-026-07276-3

BibTeX
@article{Schipper2026,
  author = {Schipper, Pascal and Aretz, Benjamin and Bourauel, Leonie and Deubel, Carolin and Petrak, Michael and Holz, Frank G and Liegl, Raffael and Mercieca, Karl and Liegl, Constance},
  title = {Sonography versus anterior segment OCT for imaging bleb morphology after glaucoma drainage device surgery},
  journal = {Graefe's Archive for Clinical and Experimental Ophthalmology},
  year = {2026},
  volume = {264},
  number = {11},
  pages = {3593--3599},
  doi = {10.1007/s00417-026-07276-3}
}
Kostenlos starten →