Repeatability and reproducibility of a new fully automatic measurement optical low coherence reflectometry biometer and agreement with swept-source optical coherence tomography-based biometer
British Journal of Ophthalmology · 2024-05 · 108 (5) · 673-678
DOI: 10.1136/bjo-2023-323268 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift British Journal of Ophthalmology, online erschienen am 04.05.2023. Verfasst von Jinjin Yu und 11 weiteren Autor:innen. Beteiligte Einrichtung: Eye Institute and Department of Ophthalmology, Institute for Medical and Engineering Innovation, Eye & ENT Hospital, NHC Key Laboratory of Myopia (Fu…. Hauptthemen laut MEDLINE: Tomography, Optical Coherence, Biometry, Cornea, Axial Length, Eye. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 27 Literaturangaben. Gefördert von: Project of National Natural Science Foundation of China, EYE & ENT Hospital of Fudan University High-level Talents Program, Shanghai Science and Technology.
Abstract (Originalsprache)
Aims
To assess the repeatability and reproducibility of the ocular measurements obtained with the Suoer SW-9000 μm Plus, a new fully automatic biometer based on optical low coherence reflectometry (OLCR) biometer, and to compare them to those obtained by a swept-source optical coherence tomography (SS-OCT)-based biometer.
Methods
This prospective study consisted of 115 eyes of 115 healthy subjects. The measurements were taken by the two optical biometers in random order. The measured parameters were axial length (AL), central corneal thickness (CCT), aqueous depth (AQD), anterior chamber depth (ACD), mean keratometry (Km), lens thickness (LT) and corneal diameter (CD). To evaluate the intraobserver repeatability and interobserver reproducibility, the within-subject SD, test–retest variability, coefficient of variation (CoV) and intraclass correlation coefficient (ICC) were adopted. The Bland-Altman plot was drawn to assess the agreement.
Results
The repeatability and reproducibility of all parameters for the new device were excellent (ICC>0.960 and CoV
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Eye Institute and Department of Ophthalmology, Institute for Medical and Engineering Innovation, Eye & ENT Hospital, NHC Key Laboratory of Myopia (Fudan University), Key Laboratory of Myopia, Chinese Academy of Medical…
- Shanghai Research Center of Ophthalmology and Optometry, Shanghai, China.
- Eye Hospital and School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, China.
- G.B. Bietti Foundation I.R.C.C.S, Rome, Italy.
ORCID-Profile: Giacomo Savini, Jinhai Huang
Förderung
- Project of National Natural Science Foundation of China
- EYE & ENT Hospital of Fudan University High-level Talents Program
- Shanghai Science and Technology
- Program for Professor of Special Appointment (Eastern Scholar) at Shanghai Institutions of Higher Learning
- Fondazione Roma and the Italian Ministry of Health
Zitieren
APA
Yu, J., Zhao, G., Lei, C. S., Wan, T., Ning, R., Xing, W., Ma, X., Pan, H., Savini, G., Schiano-Lomoriello, D., Zhou, X., & Huang, J. (2024). Repeatability and reproducibility of a new fully automatic measurement optical low coherence reflectometry biometer and agreement with swept-source optical coherence tomography-based biometer. British Journal of Ophthalmology, 108(5), 673-678. https://doi.org/10.1136/bjo-2023-323268
Vancouver
Yu J, Zhao G, Lei CS, Wan T, Ning R, Xing W, et al. Repeatability and reproducibility of a new fully automatic measurement optical low coherence reflectometry biometer and agreement with swept-source optical coherence tomography-based biometer. British Journal of Ophthalmology. 2024;108(5):673-678. doi:10.1136/bjo-2023-323268
BibTeX
@article{Yu2024,
author = {Yu, Jinjin and Zhao, Guoli and Lei, Chak Seng and Wan, Ting and Ning, Rui and Xing, Wenqian and Ma, Xindi and Pan, Hongxian and Savini, Giacomo and Schiano-Lomoriello, Domenico and Zhou, Xingtao and Huang, Jinhai},
title = {Repeatability and reproducibility of a new fully automatic measurement optical low coherence reflectometry biometer and agreement with swept-source optical coherence tomography-based biometer},
journal = {British Journal of Ophthalmology},
year = {2024},
volume = {108},
number = {5},
pages = {673--678},
doi = {10.1136/bjo-2023-323268}
}