Fachartikel in der Zeitschrift European Archives of Oto-Rhino-Laryngology, online erschienen am 08.11.2026. Verfasst von Annett Franke-Trieger und 3 weiteren Autor:innen. Beteiligte Einrichtung: Saxonian Cochlear Implant Centre, Department of Otorhinolaryngology, Head and Neck Surgery, Faculty of Medicine (and University Hospital), Technische…. Die Arbeit nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Technische Universität Dresden.
Abstract (Originalsprache)
Objective
Accurate prediction of the angular insertion depth (AID) is essential for optimising cochlear implant (CI) outcomes. However, discrepancies between preoperative predictions and postoperative results persist. In this study, we evaluated systematically the accuracy of equation-based AID prediction to identify the main contributors to prediction errors.
Methods
In this retrospective study, preoperative AID predictions for various MED-EL electrode arrays were generated by using an equation-based method, assuming full insertion with the electrode stopper at the round window. These predictions were compared with postoperatively measured AIDs obtained by digital volume tomography. The primary variable investigated was the linear insertion depth of the electrode closest to the basal position (LID E12 ), defined as its distance from the round window.
Results
The LID E12 ranged from − 13.8 mm to 11.0 mm; negative values indicate incomplete insertions, which were observed in 9% of cases. Between predicted and measured AID a median difference of − 11° (interquartile range 74°) was observed. After correcting the prediction by incorporating the measured LID E12 instead of assuming insertion to the stopper, the median difference shifted to + 35° with a markedly reduced interquartile range (37°).
Conclusion
While current AID prediction models generally show clinically acceptable accuracy, considerable discrepancies can still occur. One major contributing factor is the actual surgical placement of the electrode array. This variability, captured by the LID E12 measurement, demonstrates that the real-life result of the electrode placement often differs from the assumed placing, and that this factor should be considered when planning electrode positioning.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Saxonian Cochlear Implant Centre, Department of Otorhinolaryngology, Head and Neck Surgery, Faculty of Medicine (and University Hospital), Technische Universität Dresden, Fetscherstraße 74, 01307, Dresden, Germany. anne…
- Saxonian Cochlear Implant Centre, Department of Otorhinolaryngology, Head and Neck Surgery, Faculty of Medicine (and University Hospital), Technische Universität Dresden, Fetscherstraße 74, 01307, Dresden, Germany.
Förderung
- Technische Universität Dresden
Zitieren
APA
Franke-Trieger, A., Zahnert, T., Murrmann, K., & Lailach, S. (2026). Influence of basal-most electrode positioning on the accuracy of angular insertion depth predictions in cochlear implantation. European Archives of Oto-Rhino-Laryngology. https://doi.org/10.1007/s00405-026-10744-5
Vancouver
Franke-Trieger A, Zahnert T, Murrmann K, Lailach S. Influence of basal-most electrode positioning on the accuracy of angular insertion depth predictions in cochlear implantation. European Archives of Oto-Rhino-Laryngology. 2026. doi:10.1007/s00405-026-10744-5
BibTeX
@article{FrankeTrieger2026,
author = {Franke-Trieger, Annett and Zahnert, Thomas and Murrmann, Katrin and Lailach, Susen},
title = {Influence of basal-most electrode positioning on the accuracy of angular insertion depth predictions in cochlear implantation},
journal = {European Archives of Oto-Rhino-Laryngology},
year = {2026},
doi = {10.1007/s00405-026-10744-5}
}
EUVIMED