Localisation by Impedance
Audiology and Neurotology · 2025
DOI: 10.1159/000548171 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Audiology and Neurotology, 2025. Verfasst von Alhassan Algazlan und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology-Head and Neck Surgery, College of Medicine, King Saud University, Riyadh, Saudi Arabia, algazlan@hotmail.com. Hauptthemen laut MEDLINE: Cochlear Implantation, Cochlear Implants. Die Arbeit nennt 23 Literaturangaben.
Abstract (Originalsprache)
Background
Accurate intracochlear positioning of electrode arrays is critical for optimizing auditory outcomes during cochlear implantation (CI). While radiographic imaging remains the standard for postoperative evaluation, transimpedance matrix (TIM) measurements have emerged as a promising intraoperative technique for real-time assessment of electrode placement, particularly with a slim modiolar electrode array.
Methods
This retrospective observational study included 15 patients (24 ears) who underwent CI using CI532® and CI632® electrode arrays. Intraoperative TIM data were obtained and compared with intraoperative and immediate postoperative radiographic imaging data to determine the accuracy and reliability of TIM in detecting electrode placement anomalies. The electrode positioning patterns were analyzed using TIM heatmaps, with a focus on the voltage gradient distribution and identification of the electrode tip fold-over (TFO).
Results
Of the 24 CIs, 23 (96%) exhibited a regular TIM voltage distribution consistent with proper modiolar placement. A single instance of TFO (4%) was identified intraoperatively using TIM and confirmed via radiographic imaging. The electrode was reinserted following intraoperative correction with TIM and imaging, subsequently confirming the optimal placement. In all remaining cases, the TIM findings were in concordance with radiological assessments.
Conclusion
TIM may be a reliable intraoperative monitoring tool for CI using slim modiolar CI532® and CI632® electrode arrays position during CI. Their ability to accurately detect deviations in electrode configuration, such as the TFO, affirms their potential to augment surgical precision, eliminate the need of intraoperative radiation exposure by providing real-time feedback, minimize postoperative imaging requirements, and enhance patient outcomes. Further prospective studies with larger cohorts are warranted to validate these preliminary findings and establish standardized protocols for clinical implementation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology-Head and Neck Surgery, College of Medicine, King Saud University, Riyadh, Saudi Arabia, algazlan@hotmail.com.
- King Abdullah Ear Specialist Center (KAESC), King Saud University Medical City, Riyadh, Saudi Arabia, algazlan@hotmail.com.
- Division of Otorhinolaryngology-Head and Neck Surgery, Department of Surgery, King Abdulaziz Medical City, Riyadh, Saudi Arabia.
- Cochlear Arabia Regional Headquarter, Riyadh, Saudi Arabia.
- King Abdullah Ear Specialist Center (KAESC), King Saud University Medical City, Riyadh, Saudi Arabia.
- Department of Otolaryngology-Head and Neck Surgery, College of Medicine, King Saud University, Riyadh, Saudi Arabia.
Zitieren
APA
Algazlan, A., Salamah, M., Alrushaydan, D., Yalcouy, H., Hagr, A., & Almuhawas, F. (2025). Localisation by Impedance. Audiology and Neurotology. https://doi.org/10.1159/000548171
Vancouver
Algazlan A, Salamah M, Alrushaydan D, Yalcouy H, Hagr A, Almuhawas F. Localisation by Impedance. Audiology and Neurotology. 2025. doi:10.1159/000548171
BibTeX
@article{Algazlan2025,
author = {Algazlan, Alhassan and Salamah, Marzouqi and Alrushaydan, Dalal and Yalcouy, Hassan and Hagr, Abdulrahman and Almuhawas, Fida},
title = {Localisation by Impedance},
journal = {Audiology and Neurotology},
year = {2025},
doi = {10.1159/000548171}
}
EUVIMED