Clinically relevant differences in stress shielding between two short-stemmed femoral prostheses
Archives of Orthopaedic and Trauma Surgery · 09.07.2025 · 145 (1)
DOI: 10.1007/s00402-025-05975-w ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Archives of Orthopaedic and Trauma Surgery, online erschienen am 09.07.2025. Verfasst von Felix Werneburg und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic and Trauma Surgery, University Hospital Halle (Saale), Martin Luther University Halle-Wittenberg, Ernst-Grube Str. 40, 06120…. Hauptthemen laut MEDLINE: Hip Prosthesis, Prosthesis Design, Osteoarthritis, Hip, Arthroplasty, Replacement, Hip. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Martin-Luther-Universität Halle-Wittenberg.
Abstract (Originalsprache)
Background
Short-stemmed endoprostheses were developed to implement proximal load transmission and thus avoid stress-shielding in the proximal femur. Various prosthesis systems have been developed, which are discussed in the literature regarding stress shielding, clinical outcome, and long-term implant stability.
Methods
In this prospective randomized study, 52 patients (27 male, 25 female; average age 60.8 years) with conservatively unsuccessfully treated coxarthrosis were implanted with either a Nanos™ or Optimys™ short-stem prosthesis. Assessment included Gruen-zone based DEXA examinations immediately postoperatively and at one year to evaluate bone mineral density (BMD) and stress shielding, along with clinical outcomes using the Harris Hip Score (HHS). Radiographic measurements included offset (OFF), caput-collum-diaphyseal angle (CCD), leg length (LL), stem migration and inclination, and the occurrence of radiolucent lines (RL), assessed preoperatively, postoperatively, and at 12 months.
Results
DEXA showed differing stress-shielding profiles between stem types, favoring Optimys™ for BMD preservation. The Nanos™ group exhibited significantly greater BMD reduction in Gruen zones 1 (− 10.1%; p = 0.001), 4 (− 3.2%; p = 0.02), and 7 (− 21.3%; p = 0.001), whereas Optimys™ showed a significant decrease only in zone 7 (− 16.2%; p = 0.001). Although OFF, CCD, and LL changed significantly within groups postoperatively ( p < 0.05), no statistically significant differences were found between the two stem designs in the final postoperative measurements (all p > 0.05). Stem migration remained clinically irrelevant in both groups. A statistically significant intra-group change was observed only in the Optimys™ group (Nanos™: 1.7 mm, p = 0.13; Optimys™: 2.5 mm, p = 0.01). Similarly, a small but statistically significant change in stem inclination was observed within both groups (Nanos™: 2.2°, p = 0.002; Optimys™: 1.5°, p = 0.01). Clinical improvement as measured by the Harris Hip Score (HHS) was excellent in both groups, with no significant differences between systems (Nanos™ pre/post: 52.0 / 98.0; Optimys™ pre/post: 51.6 / 97.0; both p < 0.001).
Conclusions
When compared, the Optimys stem demonstrated reduced stress shielding through improved proximal load transmission, resulting in significantly better preservation of bone mineral density in the proximal femur. Level of evidence Ib.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopedic and Trauma Surgery, University Hospital Halle (Saale), Martin Luther University Halle-Wittenberg, Ernst-Grube Str. 40, 06120, Halle, Germany. felix.werneburg@uk-halle.de.
- Faculty of Medicine, Martin Luther University Halle-Wittenberg, Halle (Saale), Germany.
- Department of Orthopedics and Trauma Surgery, Dölau Hospital, Halle (Saale), Germany.
- Department of Orthopedic and Trauma Surgery, University Hospital Halle (Saale), Martin Luther University Halle-Wittenberg, Ernst-Grube Str. 40, 06120, Halle, Germany.
- Department of Orthopedic and Trauma Surgery, University Hospital Halle (Saale), Martin Luther University Halle-Wittenberg, Ernst-Grube Str. 40, 06120, Halle, Germany. alexander.zeh@uk-halle.de.
Förderung
- Martin-Luther-Universität Halle-Wittenberg
Zitieren
APA
Werneburg, F., Herntrich, A., Dietz, J., Wohlrab, D., Gutteck, N., Karl-Stefan, D., & Zeh, A. (2025). Clinically relevant differences in stress shielding between two short-stemmed femoral prostheses. Archives of Orthopaedic and Trauma Surgery, 145(1). https://doi.org/10.1007/s00402-025-05975-w
Vancouver
Werneburg F, Herntrich A, Dietz J, Wohlrab D, Gutteck N, Karl-Stefan D, et al. Clinically relevant differences in stress shielding between two short-stemmed femoral prostheses. Archives of Orthopaedic and Trauma Surgery. 2025;145(1). doi:10.1007/s00402-025-05975-w
BibTeX
@article{Werneburg2025,
author = {Werneburg, Felix and Herntrich, Annabell and Dietz, Julia and Wohlrab, David and Gutteck, Natalia and Karl-Stefan, Delank and Zeh, Alexander},
title = {Clinically relevant differences in stress shielding between two short-stemmed femoral prostheses},
journal = {Archives of Orthopaedic and Trauma Surgery},
year = {2025},
volume = {145},
number = {1},
doi = {10.1007/s00402-025-05975-w}
}
EUVIMED