Contrast media and radiation dose optimization with task-based automatic keV selection: a proof-of-concept study with photon-counting detector CT
European Radiology · 2025
DOI: 10.1007/s00330-025-11738-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Radiology, 2025. Verfasst von Konstantin Klambauer und 8 weiteren Autor:innen. Beteiligte Einrichtung: Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Radiation Dosage, Contrast Media, Computed Tomography Angiography, Tomography, X-Ray Computed. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
To evaluate whether task-based automatic keV selection of photon-counting detector (PCD)-CT with optimizing radiation and contrast media (CM) dose yields consistent image quality in CT angiography (CTA). Materials and methods PCD-CTA of the aorta was performed in six healthy minipigs across two scan sessions, with virtual monoenergetic images (VMI) reconstructed. In the first session, three protocols were conducted: the reference protocol A1 simulated standard CTA (210 mg iodine/kg CM, image quality (IQ)-level 117, non-contrast task, VMI: 70 keV); protocol A2 reduced radiation while keeping CM dose constant (210 mgI/kg, IQ-level 117, vascular task, VMI: 55 keV); and protocol A3 reduced CM dose while maintaining radiation (164 mgI/kg, IQ-level 117, non-contrast task, VMI: 55 keV). In the second session, protocols A2 and A3 were repeated as B1 and B2 to assess reproducibility, and protocol B3 further reduced the radiation dose with increased CM dose (252 mgI/kg, IQ-level 81, vascular task, VMI: 55 keV). Aortic CNR was measured; subjective assessments included contrast, noise, IQ, and visibility of intrahepatic arteries using a 4-point discrete visual scale.
Results
The median CTDIvol was 3.8 mGy (A1, A3), 2.4 mGy (A2, B1), 3.9 mGy (B2), and 1.6 mGy (B3), respectively; median CM doses were 23 mL (A1, A2, B1), 18 mL (A3, B2), and 28 mL (B3), respectively. CNR was comparable across protocols ( p = 0.906–0.947). Subjective metrics indicated diagnostic image quality (scores ≥ 2) for all protocols, with A1 and A3 having higher noise ( p = 0.007–0.008) and lower vascular contrast ( p = 0.003–0.008). Subjective image quality ( p = 0.226–0.342) and visibility of intrahepatic arteries ( p = 0.604–0.873) were similar.
Conclusion
Task-based automatic keV selection enables optimization of radiation and CM dose in PCD-CTA while maintaining image quality. Protocols can be balanced to either save radiation or CM dose, depending on individual patient needs. Key Points Question Balancing radiation and contrast media doses in CT angiography is essential, yet the full potential of photon-counting detector (PCD)CT for dose optimization remains underexplored.
Findings
Task-based automatic keV selection of PCD-CT enabled a 22% contrast reduction at constant radiation or a 58% radiation dose reduction with a compensatory 20% increase in contrast media. Clinical relevance Task-based automatic keV selection of PCD-CT allows individualized dose optimization by balancing radiation exposure and contrast media volume. This approach can improve patient safety by tailoring protocols to either radiation- or contrast media reduction. Graphical Abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
- Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, Maastricht, The Netherlands.
- Siemens Healthineers AG, Forchheim, Germany.
- Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland. hatem.alkadhi@usz.ch.
- MR and CT Contrast Media Research, Bayer AG, Berlin, Germany.
Zitieren
APA
Klambauer, K., Flohr, T., Moser, L. J., Mergen, V., Eberhard, M., Prokein, A., Alkadhi, H., Jost, G., & Pietsch, H. (2025). Contrast media and radiation dose optimization with task-based automatic keV selection: a proof-of-concept study with photon-counting detector CT. European Radiology. https://doi.org/10.1007/s00330-025-11738-3
Vancouver
Klambauer K, Flohr T, Moser LJ, Mergen V, Eberhard M, Prokein A, et al. Contrast media and radiation dose optimization with task-based automatic keV selection: a proof-of-concept study with photon-counting detector CT. European Radiology. 2025. doi:10.1007/s00330-025-11738-3
BibTeX
@article{Klambauer2025,
author = {Klambauer, Konstantin and Flohr, Thomas and Moser, Lukas Jakob and Mergen, Victor and Eberhard, Matthias and Prokein, Andreas and Alkadhi, Hatem and Jost, Gregor and Pietsch, Hubertus},
title = {Contrast media and radiation dose optimization with task-based automatic keV selection: a proof-of-concept study with photon-counting detector CT},
journal = {European Radiology},
year = {2025},
doi = {10.1007/s00330-025-11738-3}
}
EUVIMED