The impact of 3D real-IR delayed post gadolinium MRI parameterisation on the diagnostic performance and optimal descriptor selection in Ménière’s disease
European Radiology · 03.12.2024 · 35 (3) · 1290-1302
DOI: 10.1007/s00330-024-11218-0 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Radiology, online erschienen am 03.12.2024. Verfasst von Steve Connor und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK. steve.connor@kcl.ac.uk. Hauptthemen laut MEDLINE: Meniere Disease, Magnetic Resonance Imaging, Contrast Media, Gadolinium. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wellcome/Engineering and Physical Sciences Research Council Centre for Medical Engineering at King's College London, Cancer Research UK National Cancer Imaging Translational Accelerator, Wellcome/Engineering and Physical Sciences Research Council Centre for Medical Engineering at King’s College London.
Abstract (Originalsprache)
Objectives
To compare the performance and optimal combination of MRI descriptors used for the diagnosis of Ménière’s disease (MD) between a real-IR sequence with “zero-point” endolymph (ZPE), and an optimised real-IR sequence with negative signal endolymph (NSE). Materials and methods This retrospective single-centre cross-sectional study evaluated delayed post-gadolinium ZPE and NSE real-IR MRI in consecutive patients with Ménièriform symptoms (8/2020–10/2023). Two observers assessed 14 MRI descriptors. “Definite MD” (2015 criteria) and “all MD” ears (wider clinical criteria) were compared to controls. Cohen’s kappa and risk ratios (RR) were evaluated for each descriptor. Forward stepwise logistic regression established which combination of descriptors best predicted MD.
Results
The study included 132 patients (57 men; mean age 57.7 ± 13.6), with 87 “all MD” (56 “definite”) and 39 control ears. The NSE sequence demonstrated increased perilymph SNR, and improved both diagnostic performance and reliability for 9/14 descriptors. However, ZPE demonstrated superior diagnostic performance for the best descriptor of “saccule absent, large as or confluent with the utricle” (RR 6.571, ZPE; 6.300, NSE) and that of “asymmetric perilymphatic enhancement” (RR 3.628, ZPE; 2.903, NSE). Both sequences combined these two descriptors in the optimal predictive model for “definite MD”, with “grade 2 cochlear hydrops” also significant for NSE. ZPE and NSE descriptor combinations both correctly classified 95.8% of ears. The ZPE descriptor combination performed better for “all MD” (ZPE, AUC-ROC 0.914; NSE, AUC-ROC 0.893).
Conclusion
Parameter optimisation with NSE Real-IR influenced the optimal selection of MRI descriptors but did not improve their diagnostic performance in definite MD. Key Points Question Delayed post-gadolinium ZPE (FLAIR) and NSE (REAL-IR) sequences are widely applied for diagnosing MD, but their relative benefits remain unclear .
Findings
Optimised NSE sequences improve perilymphatic depiction and influence the selection of the optimal MRI descriptors, but do not improve diagnostic performance . Clinical relevance Radiologists may continue to apply either ZPE or NSE sequences since they offer similar diagnostic abilities, but the choice of the sequence will influence which MRI features should be evaluated to support the diagnosis of MD .
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK. steve.connor@kcl.ac.uk.
- Department of Neuroradiology, King's College Hospital, London, UK. steve.connor@kcl.ac.uk.
- Department of Radiology, Guy's Hospital and St Thomas' Hospital, London, UK. steve.connor@kcl.ac.uk.
- School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
- Department of Ear, Nose and Throat Surgery, Guy's and St Thomas' Hospital, London, UK.
- Department of Radiology, Guy's Hospital and St Thomas' Hospital, London, UK.
Förderung
- Wellcome/Engineering and Physical Sciences Research Council Centre for Medical Engineering at King's College London
- Cancer Research UK National Cancer Imaging Translational Accelerator
- Wellcome/Engineering and Physical Sciences Research Council Centre for Medical Engineering at King’s College London
- National Institute for Health Research Biomedical Research Centre at Guy’s & St Thomas’ Hospitals and King’s College London
- UK Research & Innovation London Medical Imaging and Artificial Intelligence Centre
Zitieren
APA
Connor, S., Pai, I., Touska, P., Price, D., Ourselin, S., & Hajnal, J. V. (2024). The impact of 3D real-IR delayed post gadolinium MRI parameterisation on the diagnostic performance and optimal descriptor selection in Ménière’s disease. European Radiology, 35(3), 1290-1302. https://doi.org/10.1007/s00330-024-11218-0
Vancouver
Connor S, Pai I, Touska P, Price D, Ourselin S, Hajnal JV. The impact of 3D real-IR delayed post gadolinium MRI parameterisation on the diagnostic performance and optimal descriptor selection in Ménière’s disease. European Radiology. 2024;35(3):1290-1302. doi:10.1007/s00330-024-11218-0
BibTeX
@article{Connor2024,
author = {Connor, Steve and Pai, Irumee and Touska, Philip and Price, David and Ourselin, Sebastien and Hajnal, Joseph V},
title = {The impact of 3D real-IR delayed post gadolinium MRI parameterisation on the diagnostic performance and optimal descriptor selection in Ménière’s disease},
journal = {European Radiology},
year = {2024},
volume = {35},
number = {3},
pages = {1290--1302},
doi = {10.1007/s00330-024-11218-0}
}
EUVIMED