Fachartikel in der Zeitschrift CardioVascular and Interventional Radiology, 2025. Verfasst von Austin Zhang und 2 weiteren Autor:innen. Beteiligte Einrichtung: Perelman School of Medicine at the University of Pennsylvania, Philadelphia, PA, USA. Hauptthemen laut MEDLINE: Renal Dialysis, Catheters, Indwelling, Central Venous Catheters. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To describe long-term physical durability of tunneled hemodialysis catheters, highlighted in the 2019 Kidney Disease Outcomes Quality Initiative (KDOQI) guidelines as a specific area for future research. Materials and Methods Tunneled hemodialysis catheters with known outcomes and dwell times > 1 year were entered into this retrospective study. Data includes demographics, complications, catheter type, dwell time, reason for removal, access site, and placement via exchange or de novo. Catheter durability < 1 year dwell was analyzed as a secondary aim.

Results

272 catheters in 229 patients were included. Dwell times ranged from 366 to 3,802 days (median 504), totaling 162,439 catheter days. 17 (6%) catheters > 1 year dwell had broken external components. For these, dwell times until breaking ranged from 377 to 1,436 days (median 489), totaling 10,434 catheter days. 5 had a broken hub, 11 had a broken clamp, and 1 had broken hub and clamp. 12 were Ash Split Cath (n = 240) and 5 were Arrow-Clark VectorFlow (n = 32). In the durability < 1 year sub-analysis, 6,515 catheters with dwell times < 1 year in 3,693 patients were included, totaling 425,018 catheter days. 48 were damaged, with 24 broken hubs, 17 broken clamps, and 7 holes. Median time to breakage was 110 days. 38 were Ash Split Cath (n = 5,636) and 10 Arrow-Clark VectorFlow (n = 812). In both analyses, breakage was limited to hubs, clamps, and extensions.

Conclusions

Tunneled hemodialysis catheters are exceptionally durable, rarely requiring removal for hub-related issues after one year. Breakdown was not observed as a long-term durability issue. Further, broken external components can be replaced using external repair kits. Level of Evidence Level 2b, Retrospective Study. Graphical Abstract

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Renal DialysisCatheters, IndwellingCentral Venous CathetersHumansRetrospective StudiesMaleFemaleMiddle AgedEquipment FailureAgedTime FactorsEquipment DesignAdultDevice Removal

Stichwörter der Autor:innen

DurabilityRepairTunneled hemodialysis catheter

Autor:innen und Institutionen

  • Perelman School of Medicine at the University of Pennsylvania, Philadelphia, PA, USA.
  • Department of Radiology, Division of Interventional Radiology, Perelman School of Medicine at the University of Pennsylvania, Philadelphia, PA, USA.
  • Perelman School of Medicine at the University of Pennsylvania, Philadelphia, PA, USA. streroto@pennmedicine.upenn.edu.
  • Department of Radiology, Division of Interventional Radiology, Perelman School of Medicine at the University of Pennsylvania, Philadelphia, PA, USA. streroto@pennmedicine.upenn.edu.

Zitieren

APA

Zhang, A., Clark, T. W., & Trerotola, S. O. (2025). Long-Term Durability of Tunneled Hemodialysis Catheters: Outcomes from a Single Institution 22-Year Experience. CardioVascular and Interventional Radiology. https://doi.org/10.1007/s00270-024-03941-4

Vancouver

Zhang A, Clark TW, Trerotola SO. Long-Term Durability of Tunneled Hemodialysis Catheters: Outcomes from a Single Institution 22-Year Experience. CardioVascular and Interventional Radiology. 2025. doi:10.1007/s00270-024-03941-4

BibTeX
@article{Zhang2025,
  author = {Zhang, Austin and Clark, Timothy Wi and Trerotola, Scott O},
  title = {Long-Term Durability of Tunneled Hemodialysis Catheters: Outcomes from a Single Institution 22-Year Experience},
  journal = {CardioVascular and Interventional Radiology},
  year = {2025},
  doi = {10.1007/s00270-024-03941-4}
}
Kostenlos starten →