EUVIMEDEUVIMED entdecken
Randomisierte kontrollierte Studie

Comparison of Clinical Performance Between Two Types of Symmetric-Tip Hemodialysis Catheters: A Single-Centre, Randomized Trial

Pauline Braet, Andries Van Holsbeeck, Pieter-Jan Buyck, Annouschka Laenen, Kathleen Claes, Katrien De Vusser, Geert Maleux

CardioVascular and Interventional Radiology · 2023

DOI: 10.1007/s00270-023-03476-0 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift CardioVascular and Interventional Radiology, 2023. Verfasst von Pauline Braet und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nephrology, University Hospitals KU Leuven, Leuven, Belgium. Hauptthemen laut MEDLINE: Kidney Failure, Chronic, Central Venous Catheters, Catheterization, Central Venous. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 20 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To compare the clinical performance of a newly designed, symmetric-tip Arrow-Clark™ VectorFlow® tunnelled haemodialysis catheter, with a Glidepath™, symmetric-tip tunnelled haemodialysis catheter. MATERIAL AND METHODS: From November 2018 to October 2020, patients with End-Stage Renal Disease requiring a de novo tunnelled catheter for hemodialysis, were randomized to Vectorflow® (n = 50) or to Glidepath™ catheter (n = 48). The primary outcome was catheter patency at one year following catheter insertion. Catheter failure was defined as the removal of the catheter due to infectious complications, or low blood flow rate by intraluminal thrombosis or fibrin sheath occlusion. Secondary outcomes were blood flow rate, fractional urea clearance and urea reduction ratio during dialysis. RESULTS: Demographic characteristics were not different between the two groups. At three months and on the one-year endpoint the patency rates with the Vectorflow® catheter were 95.83% and 83.33% respectively, compared to 93.02% at both endpoints with the Glidepath™ catheter (P = 0.27). Catheter failure to infectious complications or low blood flow rate was similar in both groups. Catheter blood flow rate reached the threshold of 300 ml/min at all time points for both catheters. All patients had a high mean fractional urea clearance (1.6-1.7). CONCLUSIONS: The catheter patency rate was not significantly different in patients with a VectorFlow® or a Glidepath™ catheter. Both catheters presented satisfactory dialysis adequacy over one year.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Kidney Failure, ChronicCentral Venous CathetersCatheterization, Central VenousHumansCatheters, IndwellingEquipment DesignRenal DialysisUrea

Stichwörter der Autor:innen

ComplicationDialysisRandomized trialTunnelled catheter

Substanzen

Urea

Autor:innen und Institutionen

  • Department of Nephrology, University Hospitals KU Leuven, Leuven, Belgium.
  • Department of Radiology, AZ Sint-Jan Hospital Brugge, Brugge, Belgium.
  • Department of Radiology, AZ Sint-Lucas Hospital Brugge, Brugge, Belgium.
  • Department of Radiology, University Hospitals KU Leuven, Herestraat 49, 3000, Leuven, Belgium.
  • Department of Public Health and Primary Care, Leuven Biostatistics and Statistical Bioinformatics Centre, Leuven, Belgium.
  • Department of Radiology, University Hospitals KU Leuven, Herestraat 49, 3000, Leuven, Belgium. geert.maleux@uzleuven.be.

Zitieren

APA

Braet, P., Van Holsbeeck, A., Buyck, P. J., Laenen, A., Claes, K., De Vusser, K., & Maleux, G. (2023). Comparison of Clinical Performance Between Two Types of Symmetric-Tip Hemodialysis Catheters: A Single-Centre, Randomized Trial. CardioVascular and Interventional Radiology. https://doi.org/10.1007/s00270-023-03476-0

Vancouver

Braet P, Van Holsbeeck A, Buyck PJ, Laenen A, Claes K, De Vusser K, et al. Comparison of Clinical Performance Between Two Types of Symmetric-Tip Hemodialysis Catheters: A Single-Centre, Randomized Trial. CardioVascular and Interventional Radiology. 2023. doi:10.1007/s00270-023-03476-0

BibTeX
@article{Braet2023,
  author = {Braet, Pauline and Van Holsbeeck, Andries and Buyck, Pieter-Jan and Laenen, Annouschka and Claes, Kathleen and De Vusser, Katrien and Maleux, Geert},
  title = {Comparison of Clinical Performance Between Two Types of Symmetric-Tip Hemodialysis Catheters: A Single-Centre, Randomized Trial},
  journal = {CardioVascular and Interventional Radiology},
  year = {2023},
  doi = {10.1007/s00270-023-03476-0}
}
Kostenlos starten →